Acquire images on Falcon II

Preparation in TUI/TIA

Acquire Gain/Dark References

Follow Falcon gain correction section

Recommended dose rate

The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera.
If you have to use lower dose rate, check the validity of the gain reference.

Preset Recommendation

An example can be found in Pre-MSI_Set-up