Picoammeter setup:
  • plug in
  • wait 1 hour
  • activate ZCHK (maybe already when turned on) (should say ZC on screen)
  • plug into faraday holder
  • set to lowest range (2nA)
  • activate ZCOR to set zero (should say ZZ on screen)
  • set auto range if desired
  • deactivate ZCHK to begin measuring
  • focus beam to a spot
  • pull out holder until spot hits the faraday cup (max current) (This is a negative value)
    Examples
    • 200keV
      • 1.21 nA
    • 120 keV
      • 2.53 nA

Exposure Time to Screen Current measurment

  • refer to Tecnai Help for "Exposure Time, Plate Camera CP"
  • formula: beam current (nA) = 2.15 * emulsion / measured_exposure_time
    • exp time is between 3.1 and 3.5 depending on size of beam
    • emulsion = 1
    • beam current is between 0.61 and 0.69 nA

Leginon screen scale factor measurement

  • adjust scale factor until beam current is same as faraday cup
    • 200 keV
      • 0.725
    • 120 keV

Camera sensitivity

  • 200 keV
    • 45-50 count/e
  • 120 keV
    • 107 count/e

2011-02-02 calibration on Tecnai 2:

  • faraday measurement: 1.52 nA
  • Tecnai main screen measurement (emulsion = 1): 2.15 / 2.6 = 0.83 nA
  • Leginon scale factor to match faraday current: 0.793
  • Legion measures Tietz sensitivity: 68.7 counts/e-

2013-02-28 Calibration on T12 Spirit

  • 120kV, Filament Heat= 29, step 1, Spot 3, mag 46K, Beam Current = 5.5-7.5 µA, LaB6 installed 2013-02-22
  • TVIPS F416 camera
  • Technai main screen measurement (emulsion = 1): 2.15 / 0.11 = 19.54 nA
  • Leginon scale factor to match Faraday current: 0.991 (was 0.80)
  • Leginon measure Tietz sensitivity: 38.8294 counts/e- (was 32.5211)

Finding the Faraday Cup

  • See picture below
  • Center to cup is 4mm
  • Use a 6mm or 7mm Allen (hex) wrench as a wedge (spacer) between the holder end cap and stage. Use the wrench as a lever to fine tune the position