Fitting the tilt axis model » History » Version 4

Anchi Cheng, 07/15/2010 10:26 PM

1 1 Anchi Cheng
h1. Fitting the tilt axis model
2
3 2 Anchi Cheng
h2. Find the Rough Tilt-Axis Location Manually
4 1 Anchi Cheng
5 2 Anchi Cheng
The easiest way to find the tilt axis uses the fact that the projected feature moves in a cosine function when it is off from CCD or optical axis.
6 1 Anchi Cheng
7 2 Anchi Cheng
Do this at the microscope:
8 1 Anchi Cheng
9 2 Anchi Cheng
# Use stage alpha wobbler (symmetrical tilts) to put the specimen to eucentric height.
10
# Send to scope alternatively between 0 and 30 degree. Track the position of an object near the center of the viewing screen at the magnification you want use the model.  If the tilt axis does not pass the center of the screen, the object will move upon this asymmetric movement.
11
# Move image shift x,y until the asymmetric tilt movement is minimized.
12
# Use a larger tilt to do the same if desired.
13
# Check the asymmetric movement between 0 and -30 degree to make sure it works reasonably in both way.
14
# Save the image shift as tomo preset's image shift.
15
# Estimate the combined image shift sqrt(x^2+y^2) and enter that as the custom, tilt axis-offset.
16
# From what is observed during the tilts, estimate the direction of the tilt axis and entered as the custom phi value.
17
# Perform test run with the custom *Fixed Tilt Axis Model"
18
19
h2. Checking and Improving Tilt-Axis Model
20
21
Test runs should be done at the magnification, defocus, and tilts up to the highest values planned for the real ones with a specimen with isotropic features in view and
22 1 Anchi Cheng
adjusted to eucentric height. Tilt steps should be made to allow at least 15 images taken
23
per tilt direction.
24
25
26
27
h3. First Test Run: Fixed Model
28
29
30
31
The first time the tomography application is used on a microscope, all model
32
parameters are default to zero. A test run should be done with the specimen at eucentric
33
height and all default settings including "keep the tilt axis parameters fixed" activated.
34
If the feature tracking is good, and the defocii of the images in the tilt series do not
35
significantly changed, the goniometer behavior and camera alignment are close to ideal,
36
and no improvement of the model is required.
37
38
39
If the xy tracking deteriorates quickly in the first few tilts, the magnification
40
should be dropped so that tracking is possible through out the tilt series. Once a rough
41
model is established, refinement can then be done at higher magnifications.
42
43
44
45
46
47
h3. Runs to Improve the Model-Fitting
48
49
50
51
A complete tilt model fitting is difficult with small number of data points or at only
52
low tilts. Therefore, two tilt series need to be taken in order to get a good averaged
53
model that is later fixed.
54
55
56
# tomography/settings/model>deactivate "keep the tilt axis parameters
57
fixed".
58
# tomography/settings/model>initialize the model with "this preset only" unless
59
this is a refinement of <link linkend="Tomo_low_mag_fitting">a rough model obtained at
60
a lower magnification</link>, in which case, initialize the model with "this preset
61
and lower mags".
62
# tomography/toolbar>click on "reset learning" !http://emg.nysbc.org/software/leginon/images/icons/refresh.png! to have
63
a fresh start of tilt series data included in the fitting.
64
# Presets Manager>send the "tomo" preset to scope.
65
# tomography>Start the tilt series image collection by clicking on "Simulated
66
Target" tool.
67
# repeat 4 and 5 to dynamically fit the model the second time if the first one goes
68
to completion.
69
 If the target shifts away during the run, reduce the magnification, repeat 1
70
through 9 in this procedure and then return to the higher, intended mag and refined
71
the model using the same procedure and proper setting for refinement as mentioned in
72
2.
73
# Check on web image viewer to see if the result of the second run in step 6
74
produces images of constant defocus through out each half of the tilt seriese.
75
# If the model is good, activate "keep the tilt axis parameters fixed" in the
76
tomography settings.
77
# Repeat step 2. Other than reset the learning, the tool produces an output of image
78
shift you should copy down and apply to the "tomo" preset in the future runs.
79
80
81
82
83
84
85
86
87
88
h2. Understand the results of the model fitting
89
90
91
92
If the above procedure does not yield a good model, it is necessary to study the graphs
93
output by the run to determine possible causes and derive the best fixed model that at least
94
work for most angles.
95
96
See [[Fitting the tilt axis model]] for definition of the displayed values
97
98
h3. First fitting tilt series
99
100
Figure 1 shows a typical result from the first dynamically fitted tilt series run. x and
101
102
*Figure 1* 
103
!http://emg.nysbc.org/software/leginon/images/images/tomoxyz1.png!
104
105
According to the Zheng et. al. model, the tilt-axis can be characterized by three
106
parameters: phi, the angle between the tilt axis and the detector x axis (column), offset,
107
the distance between the center of the detector and the tilt axis on the xy plane, and z0,
108
the distance between the specimen from the xy plane that contains the tilt axis.
109
110
As a first tilt series used in the model fitting, the data is not fitted until it
111
accumulate enough data points and spread over more tilts. We arbitrarily select 30 degrees as
112
the starting angle where the fit is started.
113
114
As can be seen in Figure 1, the model is significantly different from the initial as z-axis "Prediction" jumps once the fitting is started. In particular, the model z0 which corresponds to the offset of the specimen from the tilt axis in z direction dominates deviation. Such problem is evident by the near-straight line feature shift in x-axis over more than 1 um. There is also some offset in
115
x direction since the "Prediction" and "Position" flattens somewhat as it approaches zero tilt. The tilt axis has minimal tilt from x-axis since in the y-direction, despite the bump at about 40 degrees, the range of variation in the whole tilt series is about 0.06 um.
116
117
By activate "show model parameters" check box in the viewer, more plots are shown
118
(Figure 2). phi and "optical axis" offset remain fixed below 30 deg while z0 is recalculated at each tilt. As you can see, the fitting of the former two parameters are not very stable in this first trial. Therefore, we need to do the second run.
119
120
*Figure 2* 
121
!http://emg.nysbc.org/software/leginon/images/images/tomomodel1.png!
122
123
h3. Second fitting tilt series
124
125
Figure 3 shows the z-axis plot of the second run since the x and y does not change
126
significantly. Figure 4 shows its corresponding model parameters. The zero and first tilt
127
uses the initial model and the rest are fitted dynamically. In this run, because more than
128
one tilt series is found in the memory since last "reset learning", the model fitting starts
129
after the first tilt image is acquired. model parameters are rather stable over the whole
130
tilt series and suggests that the tilt axis is tilted from x-axis by about -1.3 degree, and
131
offset from the center of the detector by -0.6 um, and the specimen is off from eucentric
132
height by about -0.7 um
133
134
*Figure 3* 
135
!http://emg.nysbc.org/software/leginon/images/images/tomoz2.png!
136
137
*Figure 4*
138
!http://emg.nysbc.org/software/leginon/images/images/tomomodel2.png!
139
140
141
Although it is not possible to maintain perfect defocus prediction in the full range of
142
the tilt by using a fixed model, we found that the overall performance is better if an
143
average tilt axis model that works well in the mid-range tilt is used as a fixed model. For
144
example, for the behavior in the above figure, we choose:
145
146
147
h3. Fixed axis tilt series
148
149
Use these custom values as initial model and turn on "Keep the tilt axis parameters
150
fixed". If the result is good, judging by consistent defocus and target tracking through out
151
the tilt series, this model will be saved in the database as best model automatically, and
152
you can revert back to initialize with the model of "only this preset".
153 3 Anchi Cheng
154
h2. Low Magnification Model Fitting
155
156
The fitting of optical axis offset does not always works if the offset is so large that
157
the feature moves out of view with even a small tilt. In such a case, it is worth first
158
collect a tomography series at a lower magnification to define roughly the model.
159
160
161
At beginning of each session, or forced by the user, the model is initialized. By
162
default, at the initialization, Tomography node uses past fitting results that show good
163
agreement with the experimental data at the magnification of the preset used. If a good
164
model is not found, that from lower magnifications will be used. It is possible to force the
165
node to use a model fitted at a particular magnification by selecting it in
166
Tomography/Settings/Model.
167
168
169
Therefore, we recommend that, in case of fitting failure on good contrast images, the
170
followings should be done:
171
172
173
#  Tomography/Settings/Image Acquisition> change the preset to "hl".
174
#  Tomography/Settings> adjust Tilt and Exposure parameters to match.
175
#  Acquire the tiltseries images.
176
#  If the tracking is good, change the preset back.
177
#  Tomography/Settings/Model> Initialize with the model of (the mag of "hl"
178
preset).
179
#  Acquire the tomo-series.
180
#  If tracking is good, change back to Initialize with the model of "this preset and
181
lower" mag.
182
183 4 Anchi Cheng
See [[Troubles with Tomography]] section for various modeling problem solving examples