h1. Gatan K2 unique installation and setup 

 h2. Gatan K2 is controlled by a computer separated from the microscope 

 Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first. 

 h2. Extra Package and Installation 

 * Use all amd64 version of Windows installer 

 * SerialEM DigitalMicrograph Plug-in 

 {{include(Install_SerialEM_DigitalMicrograph_plugin)}} 

 h2. instruments.cfg 

 {{include(Gatan_K2_Instruments_cfg)}} 

 h2. Setup 

 * Set camera configuration to give the [[Leginon Image Orientation|standard Leginon orientation]].    For example, we need 270 degree rotation and flip around the vertical axis on FEI F20.    Note that the rotation required is different if it is installed post-GIF and/or on Krios. 
 * Set the [[Low_dose_shutter_configuration_for_Gatan_camera_in_Digital_Micrograph_program|shutter configuration in DM]] to protect the specimen when camera is not taking images. 
 * K2 interface in DM restricts the exposure time to multiple of some set number, depending on the mode.    When you enter in DM an invalid number, it automatically change it to the valid one when you click the acquire button.    pyscope/dmsem.py has hard-coded values for these precision. Compare them to the precision of the DM version you have since it may change with DM version. 
 * K2 frame exposure time is adjustable in Leginon gui.    However, you need to check if the number is acceptable by DM using the above method. 
 * Create the folder to store the dose fractionation raw frames.    Frames are saved as unsigned 16-bit mrc image stack in K2 computer under D:\frames by default as set in the code inside pyscope/dmsem.py in the function calculateFileSavingParams.    You must create the frames directory first.    Leginon will not do that for you.    You may change where the frames are saved here but will need to make corresponding changes when setting up [[DDD_raw_frame_file_transfer|raw frame file transfer]] Default location is D:\\frames. 
 * Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server. 
 * Unless you want to develop your own frame alignment program.    We recommend that you [[appion:Direct_Detector_Frame_Processing|use Appion to do frame gain correction and alignment]].    These are parallelized by images so it cam almost keep up with the acquisition. 

 h2. Change Flag in pyscope/dmsem.py if you have DM 2.30 version and requires flip/rotation to give the Leginon orientation. 

 In standard python site-package installation, this file is at C:\Python27\Lib\site-packages\pyscope\demsem.py 

 Find this line near the top: 
 <pre> 
 isDM230 = False 
 </pre> 
 Change it to 
 <pre> 
 isDM230 = True 
 </pre> 

 h2. Testing with pyscope 

 {{include(Gatan_K2_Installation_Pyscope_Test)}} 

 h2. Programs to open before Leginon Client: Digital Micrograph 

 h2. See more about K2 implementation in [[Using_Gatan_K2_Summit_in_Leginon]]