Project

General

Profile

Gatan K2 installation and setup » History » Version 17

Anchi Cheng, 08/08/2014 03:11 PM

1 1 Anchi Cheng
h1. Gatan K2 unique installation and setup
2
3 6 Anchi Cheng
h2. Gatan K2 is controlled by a computer separated from the microscope
4
5
Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first.
6
7 4 Anchi Cheng
h2. Extra Package and Installation
8 1 Anchi Cheng
9
* Use all amd64 version of Windows installer
10
11 12 Anchi Cheng
* SerialEM DigitalMicrograph Plug-in
12 13 Anchi Cheng
13 1 Anchi Cheng
{{include(Install_SerialEM_DigitalMicrograph_plugin)}}
14
15
h2. instruments.cfg
16
17
{{include(Gatan_K2_Instruments_cfg)}}
18
19 2 Anchi Cheng
h2. Setup
20
21 14 Anchi Cheng
* Set camera configuration to give the [[Leginon Image Orientation|standard Leginon orientation]].  For example, we need 270 degree rotation and flip around the vertical axis on FEI F20.  Note that the rotation required is different if it is installed post-GIF and/or on Krios.
22 8 Anchi Cheng
* Set the [[Low_dose_shutter_configuration_for_Gatan_camera_in_Digital_Micrograph_program|shutter configuration in DM]] to protect the specimen when camera is not taking images.
23 15 Anchi Cheng
* K2 interface in DM restricts the exposure time to multiple of some set number, depending on the mode.  When you enter in DM an invalid number, it automatically change it to the valid one when you click the acquire button.  pyscope/dmsem.py has hard-coded values for these precision. Compare them to the precision of the DM version you have since it may change with DM version.
24
* K2 frame exposure time is adjustable in Leginon gui.  However, you need to check if the number is acceptable by DM using the above method.
25
* Create the folder to store the dose fractionation raw frames.  Frames are saved as unsigned 16-bit mrc image stack in K2 computer under D:\frames by default as set in the code inside pyscope/dmsem.py in the function calculateFileSavingParams.  You must create the frames directory first.  Leginon will not do that for you.  You may change where the frames are saved here but will need to make corresponding changes when setting up [[DDD_raw_frame_file_transfer|raw frame file transfer]]
26 1 Anchi Cheng
* Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server.
27 15 Anchi Cheng
* Unless you want to develop your own frame alignment program.  We recommend that you [[appion:Direct_Detector_Frame_Processing|use Appion to do frame gain correction and alignment]].  These are parallelized by images so it cam almost keep up with the acquisition.
28 2 Anchi Cheng
29 16 Anchi Cheng
h2. (Version 3.0) Change Flag in pyscope/dmsem.py if you have DM 2.30 version and requires flip/rotation to give the Leginon orientation.
30 10 Anchi Cheng
31 16 Anchi Cheng
In standard python site-package installation, this file is at C:\Python27\Lib\site-packages\pyscope\dmsem.py
32 10 Anchi Cheng
33
Find this line near the top:
34
<pre>
35
isDM230 = False
36
</pre>
37
Change it to
38
<pre>
39 1 Anchi Cheng
isDM230 = True
40
</pre>
41 16 Anchi Cheng
42 17 Anchi Cheng
h2. (Version 3.1) Define local configuration in C:\Python27\Lib\site-packages\pyscope/dmsem.py
43 16 Anchi Cheng
44
<pre>
45
# DM Version
46
DM_VERSION = '2.30.542.0'
47
# the value in DM camera config
48
K2_CONFIG_FLIP = True
49
# multiple of 90 degrees (i.e. put 1 if 90 degrees, 3 if 270 degrees)
50
K2_CONFIG_ROTATE = 3
51
# raw frame base directory. Use '\\' as path separator
52
RAW_FRAME_DIR = 'D:\\frames\\'
53
</pre>
54
55 17 Anchi Cheng
1. Find DM_Version in DM gui under its menu bar : /Help/About DigitalMicrograph
56
2. Fill in K2_CONFIG_FLIP and K2_CONFIG_ROTATE according to DM gui under its menu bar: /Camera/Configure Camera
57 16 Anchi Cheng
3. Enter the drive the directory where raw frames will be saved.
58
59 10 Anchi Cheng
60 1 Anchi Cheng
h2. Testing with pyscope
61
62 5 Anchi Cheng
{{include(Gatan_K2_Installation_Pyscope_Test)}}
63 1 Anchi Cheng
64 6 Anchi Cheng
h2. Programs to open before Leginon Client: Digital Micrograph
65
66
h2. See more about K2 implementation in [[Using_Gatan_K2_Summit_in_Leginon]]