Gatan K2 installation and setup » History » Revision 34
Revision 33 (Anchi Cheng, 02/02/2016 02:32 PM) → Revision 34/47 (Anchi Cheng, 05/05/2017 02:12 PM)
h1. Gatan K2 unique installation and setup h2. Gatan K2 is controlled by a computer separated from the microscope Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first. h2. Extra Package and Installation * Use all amd64 version of Windows installer * SerialEM DigitalMicrograph Plug-in {{include(Install_SerialEM_DigitalMicrograph_plugin)}} h2. instruments.cfg * A template for instruments.cfg is in the installed pyscope directory as "instruments.cfg.template". Copy it to <pre>C:\Program Files\myami\instruments.cfg</pre> * Remove SimCam modules in the configuration. {{include(Gatan_K2_Instruments_cfg)}} h2. Setup {{include(dm_camera_configuration)}} * K2 interface in DM restricts the exposure time to multiple of some set number, depending on the mode. When you enter in DM an invalid number, it automatically change it to the valid one when you click the acquire button. pyscope/dmsem.py has hard-coded values for these precision. Compare them to the precision of the DM version you have since it may change with DM version. * K2 frame exposure time is adjustable in Leginon gui. However, you need to check if the number is acceptable by DM using the above method. * Create the folder to store the dose fractionation raw frames. Frames are saved as unsigned 16-bit mrc image stack in K2 computer under D:\frames by default as set in the code inside pyscope/dmsem.py in the function calculateFileSavingParams. You must create the frames directory first. Leginon will not do that for you. You may change where the frames are saved here but will need to make corresponding changes when setting up [[DDD_raw_frame_file_transfer|raw frame file transfer]] * Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server. * Unless you want to develop your own frame alignment program. We recommend that you [[appion:Direct_Detector_Frame_Processing|use Appion to do frame gain correction and alignment]]. These are parallelized by images so it cam almost keep up with the acquisition. h2. (Version 3.0) Change Flag in pyscope/dmsem.py if you have DM 2.30 version and requires flip/rotation to give the Leginon orientation. In standard python site-package installation, this file is at C:\Python27\Lib\site-packages\pyscope\dmsem.py Find this line near the top: <pre> isDM230 = False </pre> Change it to <pre> isDM230 = True </pre> h2. (Version 3.1) Define local configuration h2. In C:\Python27\Lib\site-packages\pyscope/dmsem.py <pre> # DM Version DM_VERSION = '2.30.542.0' # the value in DM camera config K2_CONFIG_FLIP = True # multiple of 90 degrees (i.e. put 1 if 90 degrees, 3 if 270 degrees) K2_CONFIG_ROTATE = 3 # raw frame base directory. Use '\\' as path separator RAW_FRAME_DIR = 'D:\\frames\\' </pre> 1. Find DM_Version in DM gui under its menu bar : /Help/About DigitalMicrograph 2. Fill in K2_CONFIG_FLIP and K2_CONFIG_ROTATE according to DM gui under its menu bar: /Camera/Configure Camera. The comment line above the value setting explains the meaning of the value. 3. Enter the drive the directory where raw frames will be saved. h2. (Version 3.2) Define local configuration in dmsem.cfg # Copy from your myami checkout pyscope/dmsem.cfg.template to C:\Program Files\myami\dmsem.cfg # Make changes to the configured value in dmsem.cfg Most parameters are explained above. * *The config parser takes "'" and "\" literally*. Therefore, do not include "'" in DM_VERION and RAW_FRAME_DIR. In addition, the path separator should be "\" * *CAMERA_ID* is usually 0. In the older version, it is defined in the same name below <pre> class GatanK2Base(DMSEM): </pre> * *EARLY_RETURN_FRAME_COUNT* specifies the number of frames you want to integrate and return to you before all frames are saved. This reduces idle time for the scope, but it may cause memory leak in super-resolution mode See Issue #3750 for update on this feature. h2. Testing with pyscope {{include(Gatan_K2_Installation_Pyscope_Test)}} h2. Trouble shooting SerialEM-CCD plugin can output the command it issues and results to DM's result panel if you set the environment variable SERIALEMCCD_DEBUG to 2. Please activate this debug log and include the portion that shows problem when reporting problems to Leginon team. h2. Programs to open before Leginon Client: Digital Micrograph h2. See more about K2 implementation in [[Using_Gatan_K2_Summit_in_Leginon]]