1 | 1 | Anchi Cheng | h1. Gatan K2 unique installation and setup |
---|---|---|---|
2 | |||
3 | 4 | Anchi Cheng | h2. Extra Package and Installation |
4 | 1 | Anchi Cheng | |
5 | * Use all amd64 version of Windows installer |
||
6 | |||
7 | 4 | Anchi Cheng | * "SerialEM DigitalMicrograph Plug-in":http://emg.nysbc.org/attachments/download/1817/SEMCCD-GMS2-64.dll |
8 | 1 | Anchi Cheng | {{include(Install_SerialEM_DigitalMicrograph_plugin)}} |
9 | |||
10 | h2. instruments.cfg |
||
11 | |||
12 | {{include(Gatan_K2_Instruments_cfg)}} |
||
13 | |||
14 | 2 | Anchi Cheng | h2. Setup |
15 | |||
16 | * Set camera configuration to 270 degree rotation and horizontal flip to give the [[Leginon Image Orientation|standard Leginon orientation]]. |
||
17 | * Create the folder to store the dose fractionation frames. Default location is D:\\frames. |
||
18 | * Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server. |
||
19 | |||
20 | 1 | Anchi Cheng | h2. Testing with pyscope |
21 | |||
22 | {{include(Gatan_K2_Installation_Pyscope_Test)}} |
||