h1. Gatan K2 unique installation and setup
h2. Extra Package and Installation
* Use all amd64 version of Windows installer
* "SerialEM DigitalMicrograph Plug-in":http://emg.nysbc.org/attachments/download/1817/SEMCCD-GMS2-64.dll
{{include(Install_SerialEM_DigitalMicrograph_plugin)}}
h2. instruments.cfg
{{include(Gatan_K2_Instruments_cfg)}}
h2. Setup
* Set camera configuration to 270 degree rotation and horizontal flip to give the [[Leginon Image Orientation|standard Leginon orientation]].
* Create the folder to store the dose fractionation frames. Default location is D:\\frames.
* Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server.
h2. Testing with pyscope
{{include(Gatan_K2_Installation_Pyscope_Test)}}
h2. See more about K2 implementation at [[Using_Gatan_K2_Summit_in_Leginon]]