Project

General

Profile

Gatan K2 installation and setup » History » Version 6

Anchi Cheng, 10/31/2013 07:13 PM

1 1 Anchi Cheng
h1. Gatan K2 unique installation and setup
2
3 6 Anchi Cheng
h2. Gatan K2 is controlled by a computer separated from the microscope
4
5
Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first.
6
7 4 Anchi Cheng
h2. Extra Package and Installation
8 1 Anchi Cheng
9
* Use all amd64 version of Windows installer
10
11 4 Anchi Cheng
* "SerialEM DigitalMicrograph Plug-in":http://emg.nysbc.org/attachments/download/1817/SEMCCD-GMS2-64.dll
12 1 Anchi Cheng
{{include(Install_SerialEM_DigitalMicrograph_plugin)}}
13
14
h2. instruments.cfg
15
16
{{include(Gatan_K2_Instruments_cfg)}}
17
18 2 Anchi Cheng
h2. Setup
19
20
* Set camera configuration to 270 degree rotation and horizontal flip to give the [[Leginon Image Orientation|standard Leginon orientation]].
21
* Create the folder to store the dose fractionation frames.  Default location is D:\\frames.
22
* Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server.
23
24 1 Anchi Cheng
h2. Testing with pyscope
25
26 5 Anchi Cheng
{{include(Gatan_K2_Installation_Pyscope_Test)}}
27 1 Anchi Cheng
28 6 Anchi Cheng
h2. Programs to open before Leginon Client: Digital Micrograph
29
30
h2. See more about K2 implementation in [[Using_Gatan_K2_Summit_in_Leginon]]