Gatan K2 installation and setup » History » Revision 6
Revision 5 (Anchi Cheng, 10/31/2013 06:10 PM) → Revision 6/47 (Anchi Cheng, 10/31/2013 07:13 PM)
h1. Gatan K2 unique installation and setup h2. Gatan K2 is controlled by a computer separated from the microscope Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first. h2. Extra Package and Installation * Use all amd64 version of Windows installer * "SerialEM DigitalMicrograph Plug-in":http://emg.nysbc.org/attachments/download/1817/SEMCCD-GMS2-64.dll {{include(Install_SerialEM_DigitalMicrograph_plugin)}} h2. instruments.cfg {{include(Gatan_K2_Instruments_cfg)}} h2. Setup * Set camera configuration to 270 degree rotation and horizontal flip to give the [[Leginon Image Orientation|standard Leginon orientation]]. * Create the folder to store the dose fractionation frames. Default location is D:\\frames. * Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server. h2. Testing with pyscope {{include(Gatan_K2_Installation_Pyscope_Test)}} h2. Programs to open before Leginon Client: Digital Micrograph h2. See more about K2 implementation in at [[Using_Gatan_K2_Summit_in_Leginon]]