Project

General

Profile

Gatan on Windows-32 » History » Version 8

Anchi Cheng, 11/03/2015 07:06 PM

1 1 Anchi Cheng
h1. Gatan on Windows-32
2
3
* This is only known to work for Gatan camera associate with FEI microscopes.  For JEOL microscope, dmsem is used.
4
5
h2. Extra Package and Installation
6
7
|_.program/package|_.notes|
8
|TecnaiCCD.dll |(Should come with the microscope).|
9
10
{{include(comtypes installer)}}
11
12
h3. Register TecnaiCCD.dll  (It may have been registered already)
13
14
*  From the command prompt, run the following commands.  You may need to enter the full path of the dll:
15
<pre>REGSVR32 TecnaiCCD.dll</pre>
16
17
h2. instruments.cfg
18
19
<pre>
20
[camera]
21
class: gatan.Gatan
22
zplane: 50
23
height: 4096
24
width: 4096
25
</pre>
26
27 3 Anchi Cheng
* height and width depend on your specific camera
28 1 Anchi Cheng
see [[instrumentscfg_for_22|instruments.cfg z plane configuration]] for more explanation.
29
30 4 Anchi Cheng
If you have more than one camera controlled through DM, you need to find its id and use class gatan.Gatan0 and gatan.Gatan1 etc in the configuration to distinguish them and assign different zplanes so the higher ones can retract when the lower one is asked to acquire image.  see #2201 for history of this issue
31
32 1 Anchi Cheng
h2. Setup
33
34
* Set camera configuration to give the [[Leginon Image Orientation|standard Leginon orientation]].
35
* Set the [[Low_dose_shutter_configuration_for_Gatan_camera_in_Digital_Micrograph_program|shutter configuration in DM]] to protect the specimen when camera is not taking images
36
37
h2. Testing with pyscope
38
39
In python command
40
<pre>
41 7 Anchi Cheng
from pyscope import gatan
42 5 Anchi Cheng
g = gatan.Gatan()
43 1 Anchi Cheng
g.setExposureTime(200)
44
g.getImage()
45
</pre>
46
47
You should get a bunch of numbers in a numpy array.
48
49
h2. Programs to open before Leginon Client: Digital Micrograph