Project

General

Profile

Gatan on Windows with semccddll » History » Version 10

Anchi Cheng, 02/02/2016 03:52 PM

1 2 Anchi Cheng
h1. Gatan on Windows-32 with SEMCCD.dll
2 1 Anchi Cheng
3 2 Anchi Cheng
dmsem module uses SEMCCD.dll written for SerialEM to communicate with Gatan software rather than tecnaiccd.dll
4 1 Anchi Cheng
5
If using this on a computer different from the scope computer, please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers|this]] first.
6
7
h2. Extra Package and Installation
8
9
* Use all win32 version of Windows installer
10
11
* SerialEM DigitalMicrograph Plug-in
12
13
{{include(Install_SerialEM_DigitalMicrograph_plugin)}}
14
15
h2. instruments.cfg
16
17
Currently UltraScan and Orius camera are supported.
18
19
For UltraScan camera (US4000), set instruments.cfg to use GatanUltraScan class. For example:
20
<pre>
21
[us4000]
22
class: dmsem.GatanUltraScan
23
zplane: 50
24
height: 4096
25
width: 4096
26
</pre>
27
28 5 Anchi Cheng
For Orius camera, set instruments.cfg to use GatanOrius class.
29
30 1 Anchi Cheng
h2. Setup
31
32 4 Anchi Cheng
{{include(dm camera configuration)}}
33 1 Anchi Cheng
34 10 Anchi Cheng
h2. Define local configuration in dmsem.cfg
35 1 Anchi Cheng
36
# Copy from your myami checkout pyscope/dmsem.cfg.template to C:\Program Files\myami\dmsem.cfg
37
# Make changes to the configured value in dmsem.cfg
38
39
Most parameters are explained above.
40
41
* *The config parser takes "'" and "\" literally*.  Therefore, do not include "'" in DM_VERION.  In addition, the path separator should be "\"
42
* *CAMERA_ID* is usually 0 if it is the only Gatan camera.
43
* *Ultrascan camera should be configured under utrascan module, and Orius camera should be configured under orius module.
44
45
For example, only UltraScan camera exists on this computer that has Digital Micrograph version 2.32.788.0, and DM camera configuration
46
has vertical flip but no rotation, then set the dmsem.cfg ultrascan module like this:
47
<pre>
48
[dm]
49
DM_VERSION = 2.32.788.0
50
51
[ultrascan]
52
CAMERA_ID = 0
53
FLIP = True
54
ROTATE = 0
55
</pre>
56 2 Anchi Cheng
57
h2. Testing with pyscope
58 3 Anchi Cheng
59 2 Anchi Cheng
{{include(Gatan_UltraScan_dmsem_Installation_Pyscope_Test)}}
60
61
h2. Trouble shooting
62
63
h3. Refused socket connection
64
65 9 Anchi Cheng
{{include(trouble shooting semccd socket connection)}}
66 2 Anchi Cheng
67
h3. SerialEM-CCD plugin can output the command it issues and results to DM's result panel if you set the environment variable SERIALEMCCD_DEBUG to larger than 0.  Set to 1 to show general log. Set to 2 to also include socket communication log.  Please activate this debug log and include the portion that shows problem when reporting problems to Leginon team.
68
69
h2. Programs to open before Leginon Client: Digital Micrograph