Gon Modeler¶
The GonModeler node models the goniometer/stage movement to give a more accurate stage
position/movement calibration. The stage position is modeled in both x and y directions. For a
more accurate calibration, many points or images need to be acquired to give a more accurate
mathematical fit to the function. This calibration works best on a grid that will always give
good cross correlations. Slot grids give large areas that can cross correlate well, but this
type of grid tends to drift. Negatively stained grids or grids with a carbon in the background
will work well.
<blockquote>
How does modeled stage position
work?
There are two types of results from doing a modeled stage calibration: 1) a function (in
the form of a harmonic series) that models the mechanical behavior of the stage 2) a
magnification adjustment (scaling and rotation) that allows the model function to be used at
different magnifications. Part 1 needs to be done at only one magnification, because the
result will be normalized in the database so that it can be used at any other magnification.
Part 2 needs to be done at any other magnifications that you wish to use this calibration.
The user interface of GonModeler node gives you two methods: "Fit Model" and "Mag Only"
These two methods are really identical except for the final result they store in the
database. "Fit Model" will store both part 1 and 2 above. "Mag Only" will onlystore part 2
(and assumes that you already have part 1 done). Since "Fit Model" is responsible for part
1, you generally need to measure a lot of points to get a good fit. You will normally select
between 2 and 5 terms for the harmonic series to get a good fit. The "Mag Only" method will
also fit a function to your measured points using the number of terms you specify. But the
resulting best fit function is not stored in the database. Only the constant term of the
resulting function is stored, because this can be used to scale the existing normalized
model function to the current magnification.
See ( http://emg.nysbc.org/publications/techreports/99-001/ ).
Right now it is not possible to use the modeled stage calibration for building a mosaic
of images. The reason is that we have not yet implemented the inverse transform of this
calibration. All of the matrix calibrations are easy to invert (for instance, you can
convert from a pixel shift to a stage shift, or invert that and convert from a stage shift
to a pixel shift). The modeled stage calibration is more complicated to invert, so right now
we can only convert from a pixel shift to a stage shift. The MosaicClickTargetFinder node is
responsible for assembling the images into a complete mosaic image. It does this by looking
at the stage position of each of the component images, and doing the reverse transform from
stage position to pixel shift. The result is the pixel offset of the component image in the
overall mosaic.
</blockquote>
Required bindings to use preset instrument configuration set by presets manager
<simplelist type="vert">
<member> PresetsManagerNode - (PresetChangedEvent) -> GonioMedelerNode</member>
</simplelist>
Toolbar¶
- Settings = set calibration, measurement, and modeling parameters
- Acquire Image = acquire an image within this node
- Measure = measure the distance of the latest movement
- Play = calibrate the goniometer model
- Abort = abort the goniometer modeling routine
Settings¶
- Use "cross | phase" correlation = to calculate the goniometer model
- Override Preset = use this node's <link linkend="camconfig">Instrument and Camera
Configuration</link> to acquire the images in this node
- Measurement
- Axis: x or y = the selected axis the stage model is being measured
for
- Points: 200 (default) = number of data points (stage movements) used
- Tolerance: 25 % = percentrage of error allowed in calculations
- Interval: 5e-06 m = interval used to incrementally move the stage to model
the stage movement
- Label = name of the goniometer measurement (without axis suffix)
- Modeling
- Label = name of the goniometer model (without axis suffix)
- Axis: x or y = axis the goniometer model is calculated for
- Magnification = the magnification the goniometer model is being created
for
- Terms = 5 (default)
- Model magnification only = disabled (default). When enabled, the node still
fit the full modeling but only the scale and the basic matrix that defines the
image and stage axis relationship are stored in the database.
Updated by Amber Herold over 14 years ago · 1 revisions