Good Alignments Save Time » History » Version 5
Anchi Cheng, 05/18/2012 03:22 PM
1 | 1 | Amber Herold | h1. Good Alignments Save Time |
---|---|---|---|
2 | |||
3 | Leginon relies on a stable microscope. The most likely reason for leginon to fail performing is bad microscopy alignment. Various bad alignments often compound the difficulty in problem solving. The requirement of leginon for good alignment is higher than most users who set up low dose image acquisition directly at the scope are used to, especially regarding alignment in LM mode since most low dose user has limited usage for LM mode images. |
||
4 | |||
5 | |||
6 | Apart from alignments that every user does without Leginon, Here are some of alignment critical to Leginon MSI applications but not a common low-dose kit users. |
||
7 | |||
8 | |||
9 | |||
10 | h2. HM beam-tilt pivot point and current rotation center set for eucentric focus |
||
11 | |||
12 | |||
13 | |||
14 | These are needed for beam-tilt based auto-focusing |
||
15 | |||
16 | |||
17 | |||
18 | |||
19 | |||
20 | h2. Defocus reset to eucentric at BOTH HM and LM modes |
||
21 | |||
22 | |||
23 | |||
24 | These are important starting point to allow consistent performance of calibrations and saved presets. |
||
25 | |||
26 | 5 | Anchi Cheng | What is [[eucentric focus]]? |
27 | |||
28 | 1 | Amber Herold | |
29 | |||
30 | |||
31 | |||
32 | h2. Image/Beam Calibration at HM mode |
||
33 | |||
34 | |||
35 | |||
36 | 4 | Anchi Cheng | Beam has to remain centered at CCD camera when Leginon move to targets with image shift. For FEI scopes, the [[HM_Beam_Shift_Image_Shift_Calibration|calibration procedure]] is in Tecnai/TEM User Interface. For JEOL scopes, a calibration is performed in Leginon to achieve the same effect. |
37 | 2 | Amber Herold | |
38 | ______ |
||
39 | |||
40 | 3 | Anchi Cheng | [[Calibrations required on FEI microscopes|< Calibrations required on FEI microscopes]] | [[Microscope Set-up Checklist|Microscope Set-up Checklist >]] |