1 | 1 | Amber Herold | h1. Grids for Calibration |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | The grid used for calibration should have reasonable contrast and features at multiple |
||
7 | scale. A frozen grid is generally not convenient for the purpose since the contrast is low and |
||
8 | can drift easily. Large protein complex negatively stained on a 200-400 mesh em grid with |
||
9 | Quantifoil, C-falt, or home-made holey carbon support is ideal. You can even use one that is |
||
10 | abandoned after a cryo run, just negatively stain it to improve the contrast. From our |
||
11 | experience, small protein negatively stained on continuous carbon support oftern gives low |
||
12 | contrast at intermediate magnification and therefore not ideal. If your Quantifoil or C-flat |
||