HM Beam Shift Image Shift Calibration » History » Version 12
Anchi Cheng, 10/29/2010 07:08 PM
1 | 4 | Anchi Cheng | h1. Calibration required on FEI microscopes |
---|---|---|---|
2 | 1 | Anchi Cheng | |
3 | Two calibrations should be done on FEI Tecnai, Polara, and Titan series of microscope for using Leginon MSI applications. |
||
4 | |||
5 | 12 | Anchi Cheng | h2. [[HM Beam Shift/Image Shift Calibration]] |
6 | 1 | Anchi Cheng | |
7 | 4 | Anchi Cheng | h3. Purpose |
8 | 1 | Anchi Cheng | |
9 | allow image shift without moving the beam and _vice versa_. |
||
10 | |||
11 | 4 | Anchi Cheng | h3. Procedure |
12 | 1 | Anchi Cheng | |
13 | 4 | Anchi Cheng | # From TUI Workset look for Alignment tools and start Image/Beam calibration located inside Image HM-TEM. Follow its instruction for the calibration. |
14 | 1 | Anchi Cheng | |
15 | !TUIAlignmnet.png! |
||
16 | 4 | Anchi Cheng | |
17 | h2. [[TEM Scripting Beam Tilt Calibration]] |