h1. Calibration required on FEI microscopes using TUI Two calibrations should be done on FEI Tecnai, Polara, and Titan series of microscope for using Leginon MSI applications. h2. HM Beam Shift/Image Shift Calibration h3. purpose allow image shift without moving the beam and _vice versa_. h3. procedure # From TUI Look for Image HM-TEM or equivalent folder in Alignment. !TUIAlignmnet.png! Alignment