HM Beam Shift Image Shift Calibration » History » Version 3
Anchi Cheng, 10/29/2010 05:41 PM
1 | 1 | Anchi Cheng | h1. Calibration required on FEI microscopes using TUI |
---|---|---|---|
2 | |||
3 | Two calibrations should be done on FEI Tecnai, Polara, and Titan series of microscope for using Leginon MSI applications. |
||
4 | |||
5 | h2. HM Beam Shift/Image Shift Calibration |
||
6 | |||
7 | h3. purpose |
||
8 | |||
9 | allow image shift without moving the beam and _vice versa_. |
||
10 | |||
11 | h3. procedure |
||
12 | |||
13 | 3 | Anchi Cheng | # From TUI Workset look for Alignment tools and start Image/Beam calibration located inside Image HM-TEM. On Titan, this might be of a similar name. |
14 | |||
15 | !TUIAlignmnet.png! |