HM Beam Shift Image Shift Calibration » History » Version 3
Anchi Cheng, 10/29/2010 05:41 PM
| 1 | 1 | Anchi Cheng | h1. Calibration required on FEI microscopes using TUI |
|---|---|---|---|
| 2 | |||
| 3 | Two calibrations should be done on FEI Tecnai, Polara, and Titan series of microscope for using Leginon MSI applications. |
||
| 4 | |||
| 5 | h2. HM Beam Shift/Image Shift Calibration |
||
| 6 | |||
| 7 | h3. purpose |
||
| 8 | |||
| 9 | allow image shift without moving the beam and _vice versa_. |
||
| 10 | |||
| 11 | h3. procedure |
||
| 12 | |||
| 13 | 3 | Anchi Cheng | # From TUI Workset look for Alignment tools and start Image/Beam calibration located inside Image HM-TEM. On Titan, this might be of a similar name. |
| 14 | |||
| 15 | !TUIAlignmnet.png! |