h1. Calibration required on FEI microscopes using TUI 

 Two calibrations should be done on FEI Tecnai, Polara, and Titan series of microscope for using Leginon MSI applications. 

 h2. HM Beam Shift/Image Shift Calibration 

 h3. purpose 

 allow image shift without moving the beam and _vice versa_. 

 h3. procedure 

 #    From TUI Workset look Look for Alignment tools and start Image/Beam calibration located inside Image HM-TEM. On Titan, this might be of a similar name. 

 HM-TEM or equivalent folder in Alignment. !TUIAlignmnet.png!