Hardware Troubles¶
- <link linkend="PresetSendError">Microscope temporarily loses communication during
preset change</link>
- <link linkend="GIFoff">Leginon complains that it can not activate GIF
slit</link>
- <link linkend="stagebad">Compustage pole touch error causes the image acquired at
the wrong place</link>
Preset manager fails to send a preset to the
microscope¶
Why: If Tecnai microscope is busy for reasons such as correcting a pole touch error,
Leginon can not comunicate with the scope until the latter is free.
Solution: Make sure the scope is not busy any more and send the preset to scope manually
to resume data collection.
Note: The immediate imaing condition may be wrong since manual preset sending does not
include extra image shift from the target, but it should recover in later process.
Leginon can not activate GIF slit¶
Why: Unknown
Try these:
- Toggle the filtered button in the EFTEM menu a few times.
- Send tomography preset and then the square preset to scope.
- Make sure Gatan's GIF tracking program is off.
There is no fix for this bug.
Image acquired at the wrong stage position¶
Commonly why: dirty or worn compustage parts
Solution: Call for scope service
- If this happens during grid atlas data collection, it has no impact on targets
selected on it. However, there is no way to patch to the intended location
easily.
- If this happens while a target is reacquired for drift correction, the drift
correction will be wrong. There is no fix except to "declare drift" in Drift Manager to
force another drift corection as soon as possible.
< General operation problems | Troubles with Imaging >