Hitachi 7800 specific installation¶
Install myami package on the associated camera computer¶
- Instead of installing myami package on HHT scope-controlling PC, the access can be through the computer controlling the digital camera.
- No special supporting package needed.
hht.cfg¶
hht.cfg records some instrument specific parameters that are required to convert the instrument output to SI units that Leginon depends on.
The values in pyscope/jeol.cfg.template is based on NYSBC HT7800 equipped with TVIPS XF416. You may start with these and copy it
to pyscope/hht.cfg or other valid myami configuration file locations.
instruments.cfg¶
[scope]
class: hitachi.HT7800
cs: 3e-3
- Cs for HR and HC modes are 3.0 mm and 5.2 mm, respectively. You will most likely use HR mode on the final image.
- You will also be adding the camera in the same instruments.cfg since this file is located on the camera computer in this instruction.
_____
h2. Next Step: Setup and calibrations
h2. Testing with pyscope
In python command
from pyscope import hitachi t =hitachi.HT7800() t.getHighTension()
You should get the current gun high tension value in Volts.