Hitachi install » History » Revision 7

« Previous | Revision 7/10 (diff) | Next »
Anchi Cheng, 07/29/2021 03:35 PM


Hitachi 7800 specific installation

Install myami package on the associated camera computer

  • Instead of installing myami package on HHT scope-controlling PC, the access can be through the computer controlling the digital camera.
  • No special supporting package needed.

hht.cfg

hht.cfg records some instrument specific parameters that are required to convert the instrument output to SI units that Leginon depends on.

The values in pyscope/jeol.cfg.template is based on NYSBC HT7800 equipped with TVIPS XF416. You may start with these and copy it
to pyscope/hht.cfg or other valid myami configuration file locations.

instruments.cfg

[scope]
class: hitachi.HT7800
cs: 3e-3

  • Cs for HR and HC modes are 3.0 mm and 5.2 mm, respectively. You will most likely use HR mode on the final image.
  • You will also be adding the camera in the same instruments.cfg since this file is located on the camera computer in this instruction.

_____
h2. Next Step: Setup and calibrations
h2. Testing with pyscope

In python command

from pyscope import hitachi
t =hitachi.HT7800()
t.getHighTension()

You should get the current gun high tension value in Volts.

Updated by Anchi Cheng over 3 years ago · 7 revisions