Project

General

Profile

Actions

Hole Depth » History » Revision 1

Revision 1/5 | Next »
Amber Herold, 04/23/2010 09:59 AM


Hole Depth

The hole depth node, is a set of tools to measure parameters needed to correlate the
intensity values of hole with (I) and without ice (I0) with the physical thickness of the ice.
The two intensity values gives the image contrast , c=ln(I0/I), that is used in Hole Finder
Node Class as the relative measurement of thickness. Since the actual calibration of contrast
against physical thickness is not yet written, we do not advise using this node unless you
plan to perform your own query and fit the values with the proper function outside
Leginon.

Required bindings for recieving images:

<highlights>
<simplelist type="vert">
<member> This node at the moment stands alone and is used in post session
processing.</member>
</simplelist>
</highlights>

Image Selection

Each selected image type are for different measurement required for the correlation.
  • I0= Image of the area after the ice is burned recorded at the magnification and
    defocus at which the contrast-thickness calibration is intended.

Relevent Image Control Panel-

  • PickHoles: Pick location and calculates mean intensity within given radius of
    the the Picked position.
  • I= Image of the area before the ice is burned recorded at the magnification and
    defocus at which the contrast-thickness calibration is intended.

Relevent Image Control Panel-

  • PickHoles: Pick location and calculates mean intensity within given radius of
    the the Picked position.
  • Hole Untilt= Image of the a small patch of ice burned away from carbob support at
    high magnification that is used to determine the radius of the burned cylindrical
    hole.

Relevent Image Control Panel-

  • Original: File Entry.
  • Edge: Create Edge Image of the hole.
  • Template: Create Ring Template and correlate with the edge image to obtain
    the radius of the cylindrical hole.
  • Hole Tilt= Image of the same area as "Hole Untilt" with the stage tilted to
    reveal some separation of the two cylinder end projection or partial ocluded path by
    one of the end projection.

Relevent Image Control Panel-

  • Original: File Entry.
  • Edge: Create Edge Image of the tilted hole.
  • Template: Using the tilt-projected ring template created from "Hole Untilt"
    and the microscopic stage alpha angle of the tilted image to correlate with the
    edge image to obtain the correlation peaks of the the two end of the
    cylinder.
  • Threshold: Threshold the correlation image to reveal only the two main
    peaks.
  • Blobs: Find the two thresholded peaks and calculates the depth of the hole
    from the separation.

Image Control Panel

NOTE: To see the effects of Testing the settings for a particular Hole Depth step in
the Image Control Panel, the corresponding Hole Depth step Display must be enabled before
using the Test feature in that step's Display Settings.

Original

Display

To view the original input image without any image processing done to it, enable
this button. Enabling this button will disable/override the Edge, Template, and
Threshold Displays.

Display Settings

  • Original Image: File Entry = Enter the path and name of an MRC image that
    matches the Image selection chosen in the node toolbar. The filename and path for
    each type are stored separately. Only the chosen type will be displayed.

Edge

The purpose of the Edge step is to convert the original image into an "edge" image
where the edges of the holes in the image are enhanced against a black background.

Display

To view the Edge image, enable this button. Enabling this button will
disable/override the Original, Template, and Threshold Displays.

Display Settings

  • Low Pass Filter Sigma = Enter the sigma of the low pass filter applied to the
    edge image. A value of 0 will not apply a low pass filter to the edge
    image.
  • Edge Finding Threshold = Enter a reasonable intensity value that will enhance
    the edges of the hole(s) in the edge image.
  • Test = Press this button to test/apply the LP sigma and Edge Finding
    Threshold effects on the edge image.

Template

The purpose of the Template step is to cross correlate a template of the properly
sized holes against the edge image where the edges of the holes have been enhanced in the
previous step. Unlike in HoleFinder, this template is a tilt projection of the circular
ring.

Display

To view the Template image, enable this button. Enabling this button will
disable/override the Original, Edge, and Threshold Displays.

Display Settings

  • Use "phase | cross" correlation = to do this calculation.
  • Rings (Add|Edit|Delete) = Add|Edit|Delete the inner and outer hole (ring)
    diameters in this section. Use the ruler tool while displaying the Edge image to
    measure the average inner and outer hole diameter. Enter the average inner and
    outer hole diameter in this section.
  • Low Pass Filter Sigma (for Phase Correlation only) = Enter a low pass filter
    sigma that is used only when phase correlation is used in this step.
  • Tilt Axis = Enter the angle in degree the tilt axis is measured from the
    y-axis of the image. Positive is counterclockwise.
  • Test = Press this button to test/apply the parameters that have been adjusted
    in this step.

Threshold

The purpose of the Threshold step is to apply a threshold to the previous Template
image. It is in unit of correlation. The Threshold image should have dots that represent
the center of all the found holes.

Display

To view the Threshold image, enable this button. Enabling this button will
disable/override the Original, Edge, and Template Displays.

Display Settings

  • Threshold = Enter the Threshold value of a dot that represents the center of
    a hole (relative to the value with the stdv shown for the correlation
    image).
  • Test = Press this button to test/apply the Threshold to the template
    correlation image. Only the dots hat represent the center of the holes should
    appear. It is highly likely that dots not representing the center of a hole will
    also appear.

Blobs

The purpose of the Blobs step in Hold Depth is not only to mark the center of the
thresholded image but also to calculate the depth of the hole from the two marked
locations.

Display

To view Blobs, either the Original, Edge, Template, or Threshold Displays must be
enabled. The Blobs that pass this step or manually picked will be shown with a turquoise
crosshair.

Display Settings

  • Border = The pixel value distance from the edge of the image where all
    potential hole targets will be erased.
  • Max. blobs = The maximum number of hole targets or blobs at this point that
    are allowed. If the number of blobs exceeds this number, the extra blobs will be
    eliminated.
  • Max. blob size = The maximum size of a blob that is allowed to pass. This is
    roughly the square of an average blob diameter. Use the ruler tool on the
    Threshold image to determine the average diameter of a blob.
  • Find blobs / Calc depth= Press this button to test/apply the Blobs criteria
    to the Threshold image and calculate the depth of the hole if 2 blobs are found.
    If only 1 blob is found, the correlation peaks are considered unresolved and a
    separation of 1 pixel is assumed in the depth calculation.
  • Calc w/ selections= Press this button to calculate the depth of the hole
    using the two manually picked blobs..

Pick Holes

The purpose of the step is to locate the burned position in either the images before
or after the burning and to calculate the mean intensity within given radius.

Display

To pick and calculate mean intensity of the Picked Holes, the Original Displays
must be enabled. The acquisition targets that pass this step will be shown with a green
crosshair.

Display Settings

  • Hole Statistics Radius = The radius in pixels of the holes (that pass the
    lattice criteria) that will be used to determine each hole's statistics.
  • Hole Statistics Zero Thickness = The intensity value of an empty hole, i.e.
    zero thickness intensity.
  • Shiftpicks = If a hole has been picked and tested from I0 image, pressing this
    button with a I image loaded causes the current picks shift by the offset of
    correlation peaks between I0 to I.
  • Test = Press this button to test/apply the zero thickness and radius criteria
    to calculate the mean intensity of the picked hole.

Updated by Amber Herold over 14 years ago · 1 revisions