Amber Herold, 04/27/2010 02:38 PM


Hole Finder

The hole finder node, in the general sense, is designed to pick imaging and focus targets
on an input image. More specifically, the hole finder node will find potential holes (circular
features of the proper size), determine whether these holes lie in the known lattice, and
determine which holes have good ice thickness. The values in this node vary so much from grid
to grid (and from square to square) that the best way to find these parameters is by going
through a step-by-step trial and error process. To proceed from one step of the hole targeting
process to another, simply proceed from top to bottom through the display selection buttons in
the image control panel. The display settings associated with each display selection are the
locations where the Hole Finder parameters can be adjusted. To see the final acquisition and
focus targets, enable only the Original, acquisition, and focus display selections.

Required bindings for recieving images and publishing targets:

<highlights>
<simplelist type="vert">
<member> previous Acquisition - (AcquisitionImagePublishEvent) -> Hole Finder</member>
<member> Hole Finder - (ImageTargetListPublishEvent) -> next Acquisition</member>
<member> Hole Finder - (QueuePublishEvent) -> next Acquisition</member>
</simplelist>
</highlights>
Required bindings for proper waiting among nodes:

<highlights>
<simplelist type="vert">
<member> Hole Finder- (ImageProcessDoneEvent) -> previous Acquisition</member>
<member> next Acquisition - (TargetListDoneEvent) -> Hole Finder</member>
</simplelist>
</highlights>

Settings

  • Allow for user verification of picked holes = When enabled, this feature will
    pause and wait for the user to verify whether the correct exposure and focus targets
    have been selected after the Hole Finder finished processing the input image.
  • Queue up targets = When enabled, submit targets do not publish them immediately
    but to put them in a target list queue.
  • Declare drift when queue submitted= When enabled, force target shift correction
    when the queue is submitted. Useful when accurate targeting is needed.
  • Skip automated hole finder= When enabled, the automatic Hole Finder algorithm
    will be skipped altogether.

Image Control Panel

NOTE: To see the effects of Testing the settings for a particular Hole Finder step in
the Image Control Panel, the corresponding Hole Finder step Display must be enabled before
using the Test feature in that step's Display Settings.

Original

Display

To view the original input image without any image processing done to it, enable
this button. Enabling this button will disable/override the Edge, Template, and
Threshold Displays.

Display Settings

  • Original Image: File Entry = Enter the path and name of a template MRC image
    that the Hole Finder can be tested on. This feature allows Hole Finder to be
    tested without having run a full Leginon MSI application.

Edge

The purpose of the Edge step is to convert the original image into an "edge" image
where the edges of the holes in the image are enhanced against a black background.

Display

To view the Edge image, enable this button. Enabling this button will
disable/override the Original, Template, and Threshold Displays.

Display Settings

  • Low Pass Filter Sigma = Enter the sigma of the low pass filter applied to the
    edge image. A value of 0 will not apply a low pass filter to the edge
    image.
  • Edge Finding Threshold = Enter a reasonable intensity value that will enhance
    the edges of the hole(s) in the edge image.
  • Test = Press this button to test/apply the LP sigma and Edge Finding
    Threshold effects on the edge image.

Template

The purpose of the Template step is to cross correlate a template of the properly
sized holes against the edge image where the edges of the holes have been enhanced in the
previous step.

Display

To view the Template image, enable this button. Enabling this button will
disable/override the Original, Edge, and Threshold Displays.

Display Settings

  • Use "phase | cross" correlation = to do this calculation.
  • Rings (Add|Edit|Delete) = Add|Edit|Delete the inner and outer hole (ring)
    diameters in this section. Use the ruler tool while displaying the Edge image to
    measure the average inner and outer hole diameter. Enter the average inner and
    outer hole diameter in this section.
  • Low Pass Filter Sigma (for Phase Correlation only) = Enter a low pass filter
    sigma that is used only when phase correlation is used in this step.
  • Test = Press this button to test/apply the parameters that have been adjusted
    in this step.

Threshold

The purpose of the Threshold step is to apply a threshold to the previous Template
image. It is in unit of number of standard deviation above the mean. The Threshold image
should have dots that represent the center of all the found holes.

Display

To view the Threshold image, enable this button. Enabling this button will
disable/override the Original, Edge, and Template Displays.

Display Settings

  • Threshold = Enter the Threshold value of a dot that represents the center of
    a hole (relative to the value with the stdv shown for the correlation
    image).
  • Test = Press this button to test/apply the Threshold to the template
    correlation image. Only the dots hat represent the center of the holes should
    appear. It is highly likely that dots not representing the center of a hole will
    also appear.

Blobs

The purpose of the Blobs step is to begin to narrow down the "hole center dots" that
have been found. This step serves as additional criteria for finding good holes.

Display

To view Blobs, either the Original, Edge, Template, or Threshold Displays must be
enabled. The Blobs that pass this step will be shown with a turquoise crosshair.

Display Settings

  • Border = The pixel value distance from the edge of the image where all
    potential hole targets will be erased.
  • Max. blobs = The maximum number of hole targets or blobs at this point that
    are allowed. If the number of blobs exceeds this number, the extra blobs will be
    eliminated.
  • Max. blob size = The maximum size of a blob that is allowed to pass. This is
    roughly the square of an average blob diameter. Use the ruler tool on the
    Threshold image to determine the average diameter of a blob.
  • Test = Press this button to test/apply the Blobs criteria to the Threshold
    image.

Lattice

The purpose of the Lattice step is to determine whether the blobs or "hole center
dots" fit in a lattice that describes how the hole layout.

Display

To view Lattice, either the Original, Edge, Template, or Threshold Displays must be
enabled. The Lattice blobs that pass this step will be shown with a pink
crosshair.

Display Settings

  • Lattice Fitting Spacing = The distance from the center of two holes in the
    horizontal/vertical lattice directions (not the diagonal lattice
    directions).
  • Lattice Fitting Tolerance = The tolerance allowed for holes that may not
    exactly lie in the described lattice.
  • Hole Statistics Radius = The radius in pixels of the holes (that pass the
    lattice criteria) that will be used to determine each hole's statistics.
  • Hole Statistics Zero Thickness = The intensity value of an empty hole, i.e.
    zero thickness intensity.
  • Test = Press this button to test/apply the Lattice criteria to the result of
    the Blobs criteria.

acquisition

The purpose of the acquisition step is to select holes of the proper ice thickness
and arrange the desired acquisition target layout. The "focus" and "acquisition" Hole
Finder steps Display Settings appear in the same window. For the purpose of this
documentation to facilitate the separation of steps in the user interface, acquisition and
focus will be addressed separately. Therefore, only the parameters pertaining to
acquisition targets will discussed in this section. The parameters not covered in this
section are part of the focus section.

Display

To view acquisition, either the Original, Edge, Template, or Threshold Displays
must be enabled. The acquisition targets that pass this step will be shown with a green
crosshair.

Display Settings

Ice Thickness Threshold:
  • Min. mean = Enter the minimum hole ice thickness value allowed.
  • Max. mean = Enter the maximum hole ice thickness value allowed.
  • Max. stdev = Enter the maximum hole standard deviation value allowed.

Target Template:

  • Use target template = When enabled, template targets will be used to compose
    a target list. When not enabled, the center of the hole is the default
    target.
  • Acquisition Target Template (Add|Edit|Delete) = Add|Edit|Delete acquisition
    targets that define the locations of each acquisition target. The acquisition
    target coordinates (x,y) are measured relative to the center of a hole. Use the
    ruler tool to determine this distance.
  • Test = Press this button to test/apply the ice thickness criteria and
    potentially create an acquisition target template.

focus

The purpose of the focus step is to focus targets near selected holes. The "focus"
and "acquisition" Hole Finder steps Display Settings appear in the same window. For the
purpose of this documentation to facilitate the separation of steps in the user interface,
acquisition and focus will be addressed separately. Therefore, only the parameters
pertaining to focus targets will discussed in this section. The parameters not covered in
this section are part of the acquisition section.

Display

To view focus, either the Original, Edge, Template, or Threshold Displays must be
enabled. The focus targets that pass this step will be shown with a blue
crosshair.

Display Settings

Ice Thickness Threshold:
  • Focus hole selection = Any Hole | Off | Good Hole; Select the type of Z Focus
    target selection. This hole will be used by Z Focus to correct stage z height when
    first moving to the new square.

- Good Hole will select one of the good holes that passed the ice thickness
test.

- Any Hole will select any of the holes that passed (up until before the ice
thickness test).

- Off will turn Z Focus correction off.

Target Template:

  • Use target template = When enabled, template targets will be used to compose
    a target list. When not enabled, the center of the hole is the default
    target.
  • Focus Target Template (Add|Edit|Delete) = Add|Edit|Delete focus targets that
    define the locations of each focus target. The focus target coordinates (x,y) are
    measured relative to the center of a hole. Use the ruler tool to determine this
    distance.
  • Focus template thickness = Enter values that are used to determine whether
    the area where the focusing has ice that is too thick. If the ice is to thick and
    does not pass this ice thickness criteria, then that focus target will not be
    used. Only the first focus target that passes is used. More than one focus target
    can pass the focus template thickness criteria.

- Use focus template thickness = When enabled, focus template targets will be
checked against these ice thickness criteria.

- Stats. radius = the radius surrounding the focus target that is included in
the statistics measurement

- Min. mean thickness = Enter the minimum ice thickness value allowed.

- Max. mean thickness = Enter the maximum ice thickness value allowed.

- Max. std. thickness = Enter the maximum standard deviation value
allowed.

  • Test = Press this button to test/apply the ice thicness focus criteria and
    create a focus target template.

< Gon Modeler | Hole Depth >


Updated by Amber Herold over 14 years ago · 2 revisions