1 1 Anchi Cheng
h1. How to Update from v21 (Microscope Windows Computer)
2
3
Instruction is similar to [[How to Update from v20 (Microscope Windows Computer)]]
4
5
* If you have done [[TEM_Scripting Beam Tilt Calibration]] and have changed the scale factor in tecnai.py, you should copy the value to your new installation.
6 5 Anchi Cheng
* If you want to avoid copying cfg files to new installation in the future, copy it to the global config directory
7
8
{{include(locate global config directory on Windows)}}
9 1 Anchi Cheng
10 3 Anchi Cheng
h2. Modify pyscope/instruments.cfg
11 1 Anchi Cheng
12 6 Anchi Cheng
Starting from 2.2, Cs values are saved with the scope in instruments.cfg that need to match the value saved in the database.  Also a z-plane needs to be assigned to each camera
13
14 3 Anchi Cheng
{{include(Instrumentscfg_for_22)}}