How to Update from v30 (Instrument Windows Computer) » History » Revision 3
Revision 2 (Anchi Cheng, 08/12/2014 05:22 PM) → Revision 3/8 (Anchi Cheng, 08/12/2014 05:26 PM)
h1. How to Update from v30 (Instrument Windows Computer)
The changes to v3.1 DOES require upgrade at the microscope and camera computers.
See [[Installation Troubleshooting]] and "Leginon Bulletin Board":http://emg.nysbc.org/projects/leginon/boards searching
for "install" if you run into problems.
h2. copy the old config files to the global location if you still use local files
{{include(locate global config directory on Windows)}}
|_config file_|likely location in 3.0 installation if not in your home directory|
|sinedon.cfg|C:\\python27\Lib\site-packages\sinedon\|
|leginon.cfg|C:\\python27\Lib\site-packages\leginon\config\|
|instruments.cfg|C:\\python27\Lib\site-packages\pyscope\|
h2. Enter Tietz PXL camera imaging size in instruments.cfg if needed in this format:
<pre>
[camera]
tietz.TietzPXL
zplane:10
height: 2048
width: 2048
</pre>
h2. About [[TEM Scripting Beam Tilt Calibration]]
* If you have done [[TEM_Scripting Beam Tilt Calibration]] and have changed the scale factor in tecnai.py, you should copy the value to your new installation.
h2. Gatan K2 Summit camera
Check and modify pyscope/dmsem.py for new configuration needed for the camera to indicate its DM version, orientation, and file saving location. More details in [[Gatan_K2_installation_and_setup]]
<pre>
# DM Version
DM_VERSION = '2.30.542.0'
# the value in DM camera config
K2_CONFIG_FLIP = True
# multiple of 90 degrees (i.e. put 1 if 90 degrees, 3 if 270 degrees)
K2_CONFIG_ROTATE = 3
# raw frame base directory. Use '\\' as path separator
RAW_FRAME_DIR = 'D:\\frames\\'
</pre>
______
[[How_to_Update_from_v30_(Linux)|< How to Update from v3.0 (Linux) ]]
______