Image Beam Compensation Calibration » History » Version 2
Anchi Cheng, 07/27/2015 06:51 PM
1 | 1 | Anchi Cheng | h1. Image Beam Compensation Calibration (Required by JEM microscopes) |
---|---|---|---|
2 | |||
3 | 2 | Anchi Cheng | h1. DO NOT GO THROUGH THIS CALIBRATION ON FEI SCOPES. DO [[Beam Shift matrix calibration]] instead. |
4 | 1 | Anchi Cheng | |
5 | This calibration is used to estimate the beam shift required to bring the beam back to the center when an image shift is applied to reach a target. It is saved as beam shift matrix and is only relevant to JEM microscopes. Such compensation is applied by microscope software with FEI scopes. |
||
6 | |||
7 | # Leginon/Presets Manager> Select a preset for the calibration and send its |
||
8 | parameter to the microscope. Matrix calibration depends on the instruments, i.e., the scope and camera used, magnification and |
||
9 | microscope high tension. Therefore, only one preset per combination needs to be |
||
10 | calibrated. |
||
11 | # Scope> Lower the main screen. This calibration does not actually acquire digital camera images. |
||
12 | # Scope> contract the beam sufficiently so that the center of the beam can be located easily on the main screen. |
||
13 | # Leginon/NodeSelector> select "Image Beam Comp" node. |
||
14 | # Leginon/Image Beam Comp/Toolbar> open "settings" window by clicking the icon to change the amount of image shift applied during the calibration. Click "OK" to save the settings and close the window when done. |
||
15 | # Leginon/Image Beam Comp/Toolbar> left-click "Calibrate" to open the calibration dialog. |
||
16 | # Leginon/Image Beam Comp/Calibrate> Follow the instruction for each step. Left-click Done when you have perform what it asks you to do at each step. |
||
17 | # Leginon/Image Beam Comp/Calibrate> Click OK when all steps are completed and satisfied to save the calibration. Click Cancel otherwise. |
||
18 | |||
19 | Testing: |
||
20 | |||
21 | # Leginon/Presets Manager> Select the preset sent above for the calibration and send its parameters to microscope and camera. |
||
22 | # Use Navigation node to [[Checking Matrix and Modeled Stage Position Calibration|check the result of the calibration]]. Select "Image Beam Shift" as the move type. Feature on the image should be shifted but the beam remains centered. |
||
23 | |||
24 | *Beam Shift Calibration Need for the Example MSI:* |
||
25 | |||
26 | |<filename>Preset</filename>|<filename>magnification</filename>| |
||
27 | |fc,fa,en,ef|50000| |
||
28 | |||
29 | ______ |
||