Project

General

Profile

Image Shift matrix calibration » History » Version 5

Amber Herold, 06/23/2010 12:56 PM

1 1 Amber Herold
h1. Image Shift matrix calibration
2
3
4
Image shift matrix calibrations are used at each magnification that Leginon uses. All 2x2
5
matrix calibrations covered by the "Matrix" node works in the same way. For a particular type
6
of movement, in this case an "image shift" using the electromagnetic lenses of the microscope,
7
a 2x2 transformation matrix needs to be created that relates the values sent to the microscope
8
and the amount of "image shift" movement seen on the CCD imaging area. Only one set of
9
measurement is sufficient for image shift matrix calibration.
10
11
12
13
14 3 Alex Kadokura
15 1 Amber Herold
_How does matrix calibration work?_
16
17
18 3 Alex Kadokura
bq. Matrix calibration is made by making N sets of measurements (specified by "N
19 1 Amber Herold
Average"). Each measurement set acquires three images, first at a given origin, second
20
with an x-axis movement in the specified "Parameter" by the specified "Shift Fraction" of
21
the image, and third with an y-axis movement by the same shift fraction. The resulting
22
shifts in the acquired images are obtained by cross correlation. A transformation matrix
23
is then generated for the measurement set. The origin is shifted by the "Interval"
24
specified in the node, in meters before the next set of measurements is taken. At the end,
25
the N matrixes obtained are averaged and saved in the database at the specific
26
magnification and movement type and can be applied to any camera configuration.
27
"Tolerance", expressed in fraction of image, is used as an error check. The calibration is
28
considered failed when the measured movement is much different from that calculated from
29
pixel calibration
30
31
32
*  Leginon/Presets Manager> Select a preset for the calibration and send its
33
parameter to the microscope. Matrix calibration depends only on magnification and
34
microscope high tension. Therefore, only one preset per combination needs to be
35
calibrated.
36
37
38
*  Scope> Make sure that the CCD is imaging an area with distinct feature.
39
preferable isotropic, i.e., a single line is not appropriate but lines crossing each
40
other is good.
41
42
43
*  Leginon/NodeSelector> select "Matrix" node.
44
45
46
*  Leginon/Matrix/Toolbar> left-click "acquire image" to obtain a test image with
47
current parameters.
48
49
50
*  If the camera settings are not ideal-
51
52
53
Leginon/Presets Manager> If you have presets set up, you can change the camera
54
setting of the preset and send it to scope.
55
56
57
OR
58
59
60
Leginon/Matrix/Toolbar> open "settings" window by clicking the icon to select
61
camera configuration and correlation method. The former will take into effect only if
62
"Overwrite Preset" is checked. Click "OK" to save the settings and close the window when
63
done.
64
65
66
*Tip: Use this step to set the camera configuration to 512x512 binned by 8 (and a
67
short exposure time, of course) for presets with larger dimension and lower bin can save
68
a lot of image acquisition time during the calibration.
69
70
71
*  Leginon/Matrix/Settings> select correlation method. Phase correlation is
72
especially efficient in cases where periodic pattern exist. The pattern often causes
73
cross correlation peak search to misidentify the correct peak in the multiple peak
74
correlation map.
75
76
77
*  Leginon/Matrix/Toolbar> select "image shift" as the Parameter and open the
78
"Parameter Setting" window by clicking on the icon to the right of the selector.
79
80
81
*  Leginon/Matrix/Matrix Settings> "Average # position"=1 is sufficient. The rest
82
can be left in default values. "Interval" is not a relavent parameter since " average #
83
position"=1.
84
85
86
*  Leginon/Matrix/Toolbar> left-click (Execute icon) to calibrate.
87
88
89
*  The image should be shifting 10-30% of the imaging area. If this is not the case,
90
then adjust the shift fraction so that this occurs. The images can be monitored in Image
91
Display Panel with display selection in image control panel set to "image". The beam
92
should also be covering the entire imaging area at all time. If significant beam shift
93
is produced during the calibration, then the microscope alignments need to be adjusted,
94
especially image/beam calibration through the FEI software).
95
96
97 5 Amber Herold
*  Use Navigation node to [[Checking Matrix and Modeled Stage Position Calibration|check the result of the calibration]]
98 1 Amber Herold
99
100
101
102
103 5 Amber Herold
Image Shift Matrix calibration need for the Example MSI:
104 1 Amber Herold
105
|<filename>Preset</filename>|<filename>magnification</filename>|
106
|gr|120|
107
|sq|550|
108
|hl|5000|
109
|fc,fa,en,ef|50000|
110 2 Amber Herold
111
112
113
______
114
115
[[Bright and Dark reference images|< Bright and Dark reference images]] | [[Beam Shift matrix calibration|Beam Shift matrix calibration >]]
116
117
______