Importance of making calibrations » History » Version 5
Amber Herold, 06/23/2010 12:20 PM
1 | 1 | Amber Herold | h1. Importance of making calibrations |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | Leginon will not function if calibrations have not been done. In essence, this (or a |
||
6 | similar combination of nodes) is the first application that should be used if calibrations do |
||
7 | not already exist. The term calibrations is an inclusive term that refers to different sets of |
||
8 | measurements such as image shift and stage position. |
||
9 | |||
10 | |||
11 | |||
12 | |||
13 | Unless you plan to perform calibration to all available magnification and camera |
||
14 | 4 | Amber Herold | configuration, the first step in performing calibration will be to [[Creating Presets for the first time|set up Presets]]. |
15 | 1 | Amber Herold | |
16 | |||
17 | The following lists all the calibrations that must be completed for Leginon MSI |
||
18 | application to function correctly. Some calibrations can not be performed without existing |
||
19 | calibration of the others. Therefore, it is necessary to follow the order listed below for the |
||
20 | very first calibration for an empty database: |
||
21 | |||
22 | |||
23 | |||
24 | |||
25 | 3 | Anchi Cheng | * [[Pixel Size Calibration|Pixel size calibration]] (for all the magnifications |
26 | 1 | Amber Herold | that will be used) |
27 | |||
28 | |||
29 | 3 | Anchi Cheng | * [[Bright and Dark reference images]] (for all the image dimension and binning that will be used) |
30 | 1 | Amber Herold | |
31 | |||
32 | 3 | Anchi Cheng | * (Optional) [[Beam Shift Matrix calibration]] (for all the magnifications that will be used) |
33 | 1 | Amber Herold | |
34 | |||
35 | 3 | Anchi Cheng | * [[Image Shift Matrix calibration]] (for all the magnifications that will be used) |
36 | 1 | Amber Herold | |
37 | |||
38 | 3 | Anchi Cheng | * [[Stage Position Matrix calibration]] (for the |
39 | 1 | Amber Herold | magnifications that will be used for gr preset in MSI... e.g. 120x) |
40 | |||
41 | |||
42 | 5 | Amber Herold | * [[Modeled Stage Position calibration]] (for magnifications |
43 | 1 | Amber Herold | in the range of ~550x to 1100x that will be used for sq preset in MSI...e.g. |
44 | 550x) |
||
45 | |||
46 | |||
47 | 3 | Anchi Cheng | * [[Autofocus Calibration with Beam Tilt Node|Beam Tilt Defocus calibrations]] (for the magnifications |
48 | 1 | Amber Herold | focus will be corrected at, both for z and defocus correction. These are the hl and fa |
49 | presets in MSI, respectively... e.g. 5000x and 50kx) |
||
50 | |||
51 | |||
52 | 3 | Anchi Cheng | * [[Autofocus Calibration with Beam Tilt Node|Beam Tilt Stigmator calibrations]] (for the |
53 | 1 | Amber Herold | magnifications where the focus and astigmatism will be corrected at. These are the hl |
54 | and fa presets in MSI... e.g. 5000x, 50kx) |
||
55 | |||
56 | |||
57 | 3 | Anchi Cheng | * [[Autofocus Calibration with Beam Tilt Node|Store eucentric focus value]] (for the |
58 | 1 | Amber Herold | magnifications where the stage z will be used for defocus correction. The hl presets in |
59 | MSI is typical... e.g. 5000x) |
||
60 | |||
61 | |||
62 | 3 | Anchi Cheng | * [[Autofocus Calibration with Beam Tilt Node|Store rotation center beam tilt value]] (for the |
63 | 1 | Amber Herold | autofocus magnifications where the image shift is useded as move type in targeting if |
64 | the stage is to be tilted. For example, the "fa" preset in MSI... e.g. 50000x) |
||
65 | |||
66 | |||
67 | 3 | Anchi Cheng | * [[Dose (camera sensitivity) calibration]] (for one |
68 | 1 | Amber Herold | magnification, the rest are scaled by the program automatically). |
69 | |||
70 | |||
71 | |||
72 | |||
73 | |||
74 | !http://emg.nysbc.org/software/leginon/images/images/Cal_flow.png! |
||
75 | |||
76 | |||
77 | |||
78 | 3 | Anchi Cheng | In most cases, calibration only need to be done once. Through experience, we have found the Stage Position and Calibrations involving Beam Tilt need to be refined often. The other calibrations (in our experience) do not fluctuate as frequently as these two do. Modeled stage position calibration does gradually go off over several months and a full calibration of the modeled stage position matrix is required to rectify the problem. |
79 | 1 | Amber Herold | |
80 | |||
81 | |||
82 | |||
83 | 3 | Anchi Cheng | Image Shift, Stage Position, and Modeled Stage Position matrix calibration are being completed so that targets can be accurated centered in the images at each different magnification. At the lowest magnifications, an image shift is too big without inducing significant beam shift so the stage has to be moved. As the magnification increases, a simple |
84 | calibration of stage movement (error typically of 1.5 um or more) is not adequate, so we have to model the stage movement (error typically of 1 um or less), but still move the stage. At |
||
85 | even higher magnifications (e.g. hl, en, ef, fc, and fa) image shifts (error of 0.3 um or less) can be used to go to targets. |
||
86 | 1 | Amber Herold | |
87 | |||
88 | |||
89 | |||
90 | The following are magnifications and types of positioning calibrations that typically have |
||
91 | been completed at a very minimum: |
||
92 | |||
93 | |||
94 | |||
95 | |_.Magnification:|_.Calibrations:| |
||
96 | |120 (i..e., mag for "gr" preset)|modeled stage position(mag-only), image shift| |
||
97 | |550 (i.e., mag for "sq" preset)|modeled stage position(full), image shift| |
||
98 | |5000 (i.e., mag for "hl" preset)|image shift| |
||
99 | |50000 (i,e, mag for "fa","fc","en" and "ef" presets)|image shift| |
||
100 | |||
101 | |||
102 | |||
103 | |||
104 | |||
105 | |||
106 | The Presets Manager has a "Calibration Status" section that will show which calibrations |
||
107 | have been completed for the given mag the selected preset is set to. We usually calibrate for |
||
108 | a range of magnifications so that later we will not have to complete them. |
||
109 | 2 | Amber Herold | |
110 | |||
111 | ______ |
||
112 | |||
113 | [[Grids for Calibration|Grids for Calibration >]] |
||
114 | |||
115 | ______ |