Project

General

Profile

Actions

Improving Autofocusing

Because MSI-raster is normally used for acquiring images on negatively stained particles on continuous carbon, the intermediate mag preset (hl) often lacks contrast. To reduce focusing failure, we recommend a different focusing sequence for the Z focus node.

  • Use Stage Alpha Tilt to determine defocus at the low magnification sq preset to do a rough stage-z adjustment to U-center
  • Refine stage-z position at the high magnification fa preset and Use beam-tilt to determine the required correction.
For "Stage_Wobble" Step:
  • Enable = Yes
  • Image registration= phase correlation
  • Fit limit = 1000 (not used)
  • Correct for delta Defocus/Z between 0 and 2e-4 meters
  • Correction type = Stage Z
  • Check drift greater than 3e-10 meters/s = No
  • Stigmator correction = No
  • Stigmator Defocus Max/Min = N/A
For "Z_to_Eucentric" Step, change these:
  • preset = fa
  • Correct for delta Defocus/Z between 0 and 2e-5 meters

< Queuing Option


Updated by Amber Herold over 14 years ago · 2 revisions