h1. Initial MSI application preference setup
2
3
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
4
5
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
6
7 5 Eric Hou
Table 16.1 Example MSI node setup
8
9 1 Amber Herold
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
10
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
11
|Grid|gr|modeled stage position|no|no|no|
12
|Square|sq|modeled stage position|yes|no|no|
13
|Hole|hl|modeled stage postion|yes|yes|yes|
14
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
15
|Exposure|en & ef|image shift|no|yes|yes|
16
|Focus|fc|image shift|no|no|yes|
17
18 9 Eric Hou
* *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
19 6 Eric Hou
 
20
Leave other setting options to default. These default options are:
21
 
22 9 Eric Hou
##  Wait 2.5 seconds before acquiring image
23 6 Eric Hou
 
24 9 Eric Hou
##  Save image to database = yes
25 6 Eric Hou
 
26
Remember to press the "+" button in order to assign a preset to a node._
27
 
28 9 Eric Hou
*  Presets Manager/Settings>
29 6 Eric Hou
 
30 1 Amber Herold
paus time between preset changes = 1 sec.
31 6 Eric Hou
 
32 1 Amber Herold
Cycle presets = yes
33 6 Eric Hou
 
34 1 Amber Herold
Optimize preset cycle = yes
35 6 Eric Hou
 
36 1 Amber Herold
Cycle Magnification Only = yes
37 6 Eric Hou
 
38 1 Amber Herold
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
39 6 Eric Hou
 
40 9 Eric Hou
*  Grid Targeting/Settings>
41 7 Eric Hou
 
42 9 Eric Hou
#  Select Preset = gr
43 1 Amber Herold
 
44 9 Eric Hou
#  Label =
45 1 Amber Herold
 
46
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
47
 
48 9 Eric Hou
#  Radius = 0.0009 m covers about the whole grid.
49 1 Amber Herold
 
50 9 Eric Hou
#  Mosaic center = stage center
51
52
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
53
54
*  Square Targeting/Mosaic Settings>
55 1 Amber Herold
 
56 9 Eric Hou
##  Calibration Parameter = stage model position
57 1 Amber Herold
 
58 9 Eric Hou
##  Scale Image = yes (default)
59 1 Amber Herold
 
60 9 Eric Hou
##  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
61
62
*  Hole Targeting/Settings and Exposure Targeting/Settings>
63 1 Amber Herold
 
64 9 Eric Hou
##  Allow for user verification of picked holes = yes
65 1 Amber Herold
 
66 9 Eric Hou
##  Queue up targets= no
67 1 Amber Herold
 
68 9 Eric Hou
##  Declare drift when queue is submitted = yes (default)
69 1 Amber Herold
 
70 9 Eric Hou
##  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
71
72
*  Focus/Settings> and Z Focus/Settings>
73 8 Eric Hou
 
74 9 Eric Hou
##  Acuire post-focus image = yes
75 1 Amber Herold
 
76 9 Eric Hou
##  Melt time = 0 second (should be customized for each grid)
77
78
*  Focus/Focus Sequence>
79 8 Eric Hou
 
80
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
81 1 Amber Herold
 
82 8 Eric Hou
Select individual step in the list box and set the following parameters
83 11 Eric Hou
84
Table 16.2 Example MSI Focus Setup
85
86 8 Eric Hou
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
87 1 Amber Herold
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
88
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
89
|Manual_after|fc|Manual|N/A|Yes| 
90 10 Eric Hou
91 8 Eric Hou
For both autofocus steps:
92 1 Amber Herold
 
93 9 Eric Hou
##  Tilt = 0.01 radian each direction
94 8 Eric Hou
 
95 9 Eric Hou
##  Image registration = phase correlation
96 8 Eric Hou
 
97 9 Eric Hou
##  Fit limit = 1000
98 8 Eric Hou
 
99 9 Eric Hou
##  Correct for delta Defocus/Z between 0 and 2e-5 meters
100 8 Eric Hou
 
101 9 Eric Hou
##  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
102 8 Eric Hou
 
103 9 Eric Hou
##  Stigmator correction = No (typical) Yes (if well calibrated)
104 8 Eric Hou
 
105
##  Stigmator Defocus Min = 2e-06 meters
106 1 Amber Herold
 
107 9 Eric Hou
##  Stigmator Defocus Max = 4e-06 meters
108 11 Eric Hou
109 12 Eric Hou
* Z Focus/Focus Sequence>
110 1 Amber Herold
 
111
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
112 11 Eric Hou
&nbsp;
113 1 Amber Herold
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
114 11 Eric Hou
&nbsp;
115 1 Amber Herold
Select individual step in the list box and set the following parameters
116
117 11 Eric Hou
Table 16.3 Example MSI Z Focus Steps.
118 1 Amber Herold
119
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
120
|Stage_Wobble|sq|Stage Tilt|No|
121
|Z_to_Eucentric|hl|Beam Tilt|Yes|
122
|Manual_after|fc or hl*|Manual|Yes|
123
124
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
125
126
For "Stage_Wobble" Step:
127
*  Tilt = 2 degrees each direction
128
129
*  Image registration= phase correlation
130
131
*  Fit limit = N/A
132
133
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
134
135
*  Correction type = Stage Z
136
137
*  Check drift greater than 3e-10 meters/s = No
138
139
*  Stigmator correction = No
140
141
*  Stigmator Defocus Max/Min = N/A
142
143
For "Z_to_Eucentric" Step:
144
*  Tilt = 0.01 radian each direction
145
146
*  Image registration= phase correlation
147
148
*  Fit limit = 1500
149
150
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
151
152
*  Correction type = Stage Z
153
154
*  Reset defocus = Default
155
156
*  Check drift greater than 3e-10 meters/s = No
157
158
*  Stigmator correction = No
159
160 2 Amber Herold
*  Stigmator Defocus Max/Min = N/A