Project

General

Profile

Initial MSI application preference setup » History » Version 19

Anchi Cheng, 01/19/2011 05:56 PM

1 1 Amber Herold
h1. Initial MSI application preference setup
2
3 18 Anchi Cheng
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Leginon-Appion Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you can use these straight away as long as the calibration required are made.
4 1 Amber Herold
5 19 Anchi Cheng
If you mess up your settings too badly that it does not run any more, you may [[Default Settings|revert back to your institution default]] easily.
6
7 18 Anchi Cheng
The following is the comprehensive list of what is set as default during the initial loading for basic MSI-T, MSI-Edge, and MSI-Raster applications. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users by making the changes as "Leginon-Appion Administrator" user.
8 1 Amber Herold
9 5 Eric Hou
Table 16.1 Example MSI node setup
10
11 16 Anchi Cheng
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target using ancestor:|
12 1 Amber Herold
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
13
|Grid|gr|modeled stage position|no|no|no|
14
|Square|sq|modeled stage position|yes|no|no|
15 16 Anchi Cheng
|Hole|hl|modeled stage postion|yes|yes|one|
16
|Z Focus|hl or fc*|modeled stage position|no|no|one|
17
|Exposure|en & ef|image shift|no|yes|one|
18
|Focus|fc|image shift|no|no|one|
19 1 Amber Herold
20 9 Eric Hou
* *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
21 6 Eric Hou
 
22
Leave other setting options to default. These default options are:
23
 
24 9 Eric Hou
##  Wait 2.5 seconds before acquiring image
25 6 Eric Hou
 
26 9 Eric Hou
##  Save image to database = yes
27 6 Eric Hou
 
28
Remember to press the "+" button in order to assign a preset to a node._
29
 
30 9 Eric Hou
*  Presets Manager/Settings>
31 6 Eric Hou
 
32 1 Amber Herold
paus time between preset changes = 1 sec.
33 6 Eric Hou
 
34 1 Amber Herold
Cycle presets = yes
35 6 Eric Hou
 
36 1 Amber Herold
Optimize preset cycle = yes
37 6 Eric Hou
 
38 1 Amber Herold
Cycle Magnification Only = yes
39 6 Eric Hou
 
40 1 Amber Herold
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
41 6 Eric Hou
 
42 9 Eric Hou
*  Grid Targeting/Settings>
43 7 Eric Hou
 
44 9 Eric Hou
#  Select Preset = gr
45 1 Amber Herold
 
46 9 Eric Hou
#  Label =
47 1 Amber Herold
 
48
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
49
 
50 9 Eric Hou
#  Radius = 0.0009 m covers about the whole grid.
51 1 Amber Herold
 
52 9 Eric Hou
#  Mosaic center = stage center
53
54
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
55
56
*  Square Targeting/Mosaic Settings>
57 1 Amber Herold
 
58 9 Eric Hou
##  Calibration Parameter = stage model position
59 1 Amber Herold
 
60 9 Eric Hou
##  Scale Image = yes (default)
61 1 Amber Herold
 
62 9 Eric Hou
##  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
63
64
*  Hole Targeting/Settings and Exposure Targeting/Settings>
65 1 Amber Herold
 
66 9 Eric Hou
##  Allow for user verification of picked holes = yes
67 1 Amber Herold
 
68 9 Eric Hou
##  Queue up targets= no
69 1 Amber Herold
 
70 9 Eric Hou
##  Declare drift when queue is submitted = yes (default)
71 1 Amber Herold
 
72 9 Eric Hou
##  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
73
74
*  Focus/Settings> and Z Focus/Settings>
75 8 Eric Hou
 
76 9 Eric Hou
##  Acuire post-focus image = yes
77 1 Amber Herold
 
78 9 Eric Hou
##  Melt time = 0 second (should be customized for each grid)
79
80 11 Eric Hou
Table 16.2 Example MSI Focus Setup
81 1 Amber Herold
82
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
83
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
84
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
85
|Manual_after|fc|Manual|N/A|Yes| 
86
87 13 Eric Hou
*  Focus/Focus Sequence>
88
 
89
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
90
 
91
Select individual step in the list box and set the following parameters
92
 
93 8 Eric Hou
For both autofocus steps:
94 1 Amber Herold
 
95 9 Eric Hou
##  Tilt = 0.01 radian each direction
96 8 Eric Hou
 
97 9 Eric Hou
##  Image registration = phase correlation
98 8 Eric Hou
 
99 9 Eric Hou
##  Fit limit = 1000
100 8 Eric Hou
 
101 9 Eric Hou
##  Correct for delta Defocus/Z between 0 and 2e-5 meters
102 8 Eric Hou
 
103 9 Eric Hou
##  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
104 8 Eric Hou
 
105 1 Amber Herold
##  Stigmator correction = No (typical) Yes (if well calibrated)
106
 
107
##  Stigmator Defocus Min = 2e-06 meters
108
 
109
##  Stigmator Defocus Max = 4e-06 meters
110
111
Table 16.3 Example MSI Z Focus Steps.
112
113
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
114
|Stage_Wobble|sq|Stage Tilt|No|
115
|Z_to_Eucentric|hl|Beam Tilt|Yes|
116
|Manual_after|fc or hl*|Manual|Yes|
117 11 Eric Hou
118 13 Eric Hou
* Z Focus/Focus Sequence>
119
 
120
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
121
&nbsp;
122
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
123
&nbsp;
124
Select individual step in the list box and set the following parameters
125
&nbsp;
126 1 Amber Herold
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
127 15 Eric Hou
128
**   For "Stage_Wobble" Step:
129 13 Eric Hou
&nbsp;
130
##  Tilt = 2 degrees each direction
131
&nbsp;
132
##  Image registration= phase correlation
133
&nbsp;
134
##  Fit limit = N/A
135
&nbsp;
136
##  Correct for delta Defocus/Z between 0 and 0.0002 meters
137
&nbsp;
138
##  Correction type = Stage Z
139
&nbsp;
140
##  Check drift greater than 3e-10 meters/s = No
141
&nbsp;
142
##  Stigmator correction = No
143 1 Amber Herold
&nbsp;
144
##  Stigmator Defocus Max/Min = N/A
145 15 Eric Hou
146
**   For "Z_to_Eucentric" Step:
147 14 Eric Hou
&nbsp;
148 13 Eric Hou
##  Tilt = 0.01 radian each direction
149
&nbsp;
150
##  Image registration= phase correlation
151
&nbsp;
152
##  Fit limit = 1500
153
&nbsp;
154
##  Correct for delta Defocus/Z between 0 and 0.0002 meters
155
&nbsp;
156
##  Correction type = Stage Z
157
&nbsp;
158
##  Reset defocus = Default
159
&nbsp;
160
##  Check drift greater than 3e-10 meters/s = No
161
&nbsp;
162
##  Stigmator correction = No
163
&nbsp;
164
##  Stigmator Defocus Max/Min = N/A
165 2 Amber Herold
166
167
______
168
169
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
170
171 1 Amber Herold
______