Project

General

Profile

Initial MSI application preference setup » History » Version 2

Amber Herold, 04/27/2010 02:21 PM

1 1 Amber Herold
h1. Initial MSI application preference setup
2
3
4
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
5
6
7
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
8
9
10
11
*
12
13
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
14
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
15
|Grid|gr|modeled stage position|no|no|no|
16
|Square|sq|modeled stage position|yes|no|no|
17
|Hole|hl|modeled stage postion|yes|yes|yes|
18
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
19
|Exposure|en & ef|image shift|no|yes|yes|
20
|Focus|fc|image shift|no|no|yes|
21
22
23
24
25
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
26
27
28
Leave other setting options to default. These default options are:
29
30
31
32
*  Wait 2.5 seconds before acquiring image
33
34
35
*  Save image to database = yes
36
37
38
39
40
41
_Remember to press the "+" button in order to assign a preset to a node._
42
43
44
45
46
47
*  Presets Manager/Settings>
48
49
50
paus time between preset changes = 1 sec.
51
52
Cycle presets = yes
53
54
Optimize preset cycle = yes
55
56
Cycle Magnification Only = yes
57
58
59
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
60
61
62
*  Grid Targeting/Settings>
63
64
65
66
67
68
*  Select Preset = gr
69
70
71
*  Label =
72
73
74
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
75
76
77
*  Radius = 0.0009 m covers about the whole grid.
78
79
80
*  Mosaic center = stage center
81
82
83
84
85
86
87
88
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
89
90
91
*  Square Targeting/Mosaic Settings>
92
93
94
95
96
97
*  Calibration Parameter = stage model position
98
99
100
*  Scale Image = yes (default)
101
102
103
*  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
104
105
106
107
108
*  Hole Targeting/Settings and Exposure Targeting/Settings>
109
110
111
112
*  Allow for user verification of picked holes = yes
113
114
115
*  Queue up targets= no
116
117
118
*  Declare drift when queue is submitted = yes (default)
119
120
121
*  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
122
123
124
125
126
127
*  Focus/Settings> and Z Focus/Settings>
128
129
130
131
*  Acuire post-focus image = yes
132
133
134
*  Melt time = 0 second (should be customized for each grid)
135
136
137
138
139
140
*  Focus/Focus Sequence>
141
142
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
143
144
145
Select individual step in the list box and set the following parameters
146
147
148
149
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
150
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
151
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
152
|Manual_after|fc|Manual|N/A|Yes|
153
154
155
156
157
For both autofocus steps:
158
*  Tilt = 0.01 radian each direction
159
160
161
*  Image registration = phase correlation
162
163
164
*  Fit limit = 1000
165
166
167
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
168
169
170
*  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
171
172
173
*  Stigmator correction = No (typical) Yes (if well calibrated)
174
175
176
*  Stigmator Defocus Min = 2e-06 meters
177
178
179
*  Stigmator Defocus Max = 4e-06 meters
180
181
182
183
184
185
*  Z Focus/Focus Sequence>
186
187
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
188
189
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
190
191
192
Select individual step in the list box and set the following parameters
193
194
195
196
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
197
|Stage_Wobble|sq|Stage Tilt|No|
198
|Z_to_Eucentric|hl|Beam Tilt|Yes|
199
|Manual_after|fc or hl*|Manual|Yes|
200
201
202
203
204
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
205
206
207
For "Stage_Wobble" Step:
208
*  Tilt = 2 degrees each direction
209
210
211
*  Image registration= phase correlation
212
213
214
*  Fit limit = N/A
215
216
217
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
218
219
220
*  Correction type = Stage Z
221
222
223
*  Check drift greater than 3e-10 meters/s = No
224
225
226
*  Stigmator correction = No
227
228
229
*  Stigmator Defocus Max/Min = N/A
230
231
232
233
234
235
For "Z_to_Eucentric" Step:
236
*  Tilt = 0.01 radian each direction
237
238
239
*  Image registration= phase correlation
240
241
242
*  Fit limit = 1500
243
244
245
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
246
247
248
*  Correction type = Stage Z
249
250
251
*  Reset defocus = Default
252
253
254
*  Check drift greater than 3e-10 meters/s = No
255
256
257
*  Stigmator correction = No
258
259
260
*  Stigmator Defocus Max/Min = N/A
261 2 Amber Herold
262
263
264
______
265
266
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
267
268
______