h1. Initial MSI application preference setup
2
3
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
4
5
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
6
7 3 Eric Hou
# Table 16.1 Example MSI node setup
8
 
9 1 Amber Herold
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
10
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
11
|Grid|gr|modeled stage position|no|no|no|
12
|Square|sq|modeled stage position|yes|no|no|
13
|Hole|hl|modeled stage postion|yes|yes|yes|
14
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
15
|Exposure|en & ef|image shift|no|yes|yes|
16
|Focus|fc|image shift|no|no|yes|
17
18
19
20
21
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
22
23
24
Leave other setting options to default. These default options are:
25
26
27
28
*  Wait 2.5 seconds before acquiring image
29
30
31
*  Save image to database = yes
32
33
34
35
36
37
_Remember to press the "+" button in order to assign a preset to a node._
38
39
40
41
42
43
*  Presets Manager/Settings>
44
45
46
paus time between preset changes = 1 sec.
47
48
Cycle presets = yes
49
50
Optimize preset cycle = yes
51
52
Cycle Magnification Only = yes
53
54
55
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
56
57
58
*  Grid Targeting/Settings>
59
60
61
62
63
64
*  Select Preset = gr
65
66
67
*  Label =
68
69
70
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
71
72
73
*  Radius = 0.0009 m covers about the whole grid.
74
75
76
*  Mosaic center = stage center
77
78
79
80
81
82
83
84
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
85
86
87
*  Square Targeting/Mosaic Settings>
88
89
90
91
92
93
*  Calibration Parameter = stage model position
94
95
96
*  Scale Image = yes (default)
97
98
99
*  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
100
101
102
103
104
*  Hole Targeting/Settings and Exposure Targeting/Settings>
105
106
107
108
*  Allow for user verification of picked holes = yes
109
110
111
*  Queue up targets= no
112
113
114
*  Declare drift when queue is submitted = yes (default)
115
116
117
*  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
118
119
120
121
122
123
*  Focus/Settings> and Z Focus/Settings>
124
125
126
127
*  Acuire post-focus image = yes
128
129
130
*  Melt time = 0 second (should be customized for each grid)
131
132
133
134
135
136
*  Focus/Focus Sequence>
137
138
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
139
140
141
Select individual step in the list box and set the following parameters
142
143
144
145
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
146
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
147
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
148
|Manual_after|fc|Manual|N/A|Yes|
149
150
151
152
153
For both autofocus steps:
154
*  Tilt = 0.01 radian each direction
155
156
157
*  Image registration = phase correlation
158
159
160
*  Fit limit = 1000
161
162
163
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
164
165
166
*  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
167
168
169
*  Stigmator correction = No (typical) Yes (if well calibrated)
170
171
172
*  Stigmator Defocus Min = 2e-06 meters
173
174
175
*  Stigmator Defocus Max = 4e-06 meters
176
177
178
179
180
181
*  Z Focus/Focus Sequence>
182
183
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
184
185
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
186
187
188
Select individual step in the list box and set the following parameters
189
190
191
192
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
193
|Stage_Wobble|sq|Stage Tilt|No|
194
|Z_to_Eucentric|hl|Beam Tilt|Yes|
195
|Manual_after|fc or hl*|Manual|Yes|
196
197
198
199
200
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
201
202
203
For "Stage_Wobble" Step:
204
*  Tilt = 2 degrees each direction
205
206
207
*  Image registration= phase correlation
208
209
210
*  Fit limit = N/A
211
212
213
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
214
215
216
*  Correction type = Stage Z
217
218
219
*  Check drift greater than 3e-10 meters/s = No
220
221
222
*  Stigmator correction = No
223
224
225
*  Stigmator Defocus Max/Min = N/A
226
227
228
229
230
231
For "Z_to_Eucentric" Step:
232
*  Tilt = 0.01 radian each direction
233
234
235
*  Image registration= phase correlation
236
237
238
*  Fit limit = 1500
239
240
241
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
242
243
244
*  Correction type = Stage Z
245
246
247
*  Reset defocus = Default
248
249
250
*  Check drift greater than 3e-10 meters/s = No
251
252
253
*  Stigmator correction = No
254
255
256
*  Stigmator Defocus Max/Min = N/A
257 2 Amber Herold
258
259
260
______
261
262
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
263
264
______