Project

General

Profile

Initial MSI application preference setup » History » Version 4

Eric Hou, 06/24/2010 03:13 PM

1 1 Amber Herold
h1. Initial MSI application preference setup
2
3
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
4
5
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
6
7 3 Eric Hou
# Table 16.1 Example MSI node setup
8
 
9 1 Amber Herold
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
10
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
11
|Grid|gr|modeled stage position|no|no|no|
12
|Square|sq|modeled stage position|yes|no|no|
13
|Hole|hl|modeled stage postion|yes|yes|yes|
14
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
15
|Exposure|en & ef|image shift|no|yes|yes|
16
|Focus|fc|image shift|no|no|yes|
17
18
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
19
20
21
Leave other setting options to default. These default options are:
22
23
24
25
*  Wait 2.5 seconds before acquiring image
26
27
28
*  Save image to database = yes
29
30
31
32
33
34
_Remember to press the "+" button in order to assign a preset to a node._
35
36
37
38
39
40
*  Presets Manager/Settings>
41
42
43
paus time between preset changes = 1 sec.
44
45
Cycle presets = yes
46
47
Optimize preset cycle = yes
48
49
Cycle Magnification Only = yes
50
51
52
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
53
54
55
*  Grid Targeting/Settings>
56
57
58
59
60
61
*  Select Preset = gr
62
63
64
*  Label =
65
66
67
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
68
69
70
*  Radius = 0.0009 m covers about the whole grid.
71
72
73
*  Mosaic center = stage center
74
75
76
77
78
79
80
81
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
82
83
84
*  Square Targeting/Mosaic Settings>
85
86
87
88
89
90
*  Calibration Parameter = stage model position
91
92
93
*  Scale Image = yes (default)
94
95
96
*  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
97
98
99
100
101
*  Hole Targeting/Settings and Exposure Targeting/Settings>
102
103
104
105
*  Allow for user verification of picked holes = yes
106
107
108
*  Queue up targets= no
109
110
111
*  Declare drift when queue is submitted = yes (default)
112
113
114
*  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
115
116
117
118
119
120
*  Focus/Settings> and Z Focus/Settings>
121
122
123
124
*  Acuire post-focus image = yes
125
126
127
*  Melt time = 0 second (should be customized for each grid)
128
129
130
131
132
133
*  Focus/Focus Sequence>
134
135
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
136
137
138
Select individual step in the list box and set the following parameters
139
140
141
142
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
143
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
144
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
145
|Manual_after|fc|Manual|N/A|Yes|
146
147
148
149
150
For both autofocus steps:
151
*  Tilt = 0.01 radian each direction
152
153
154
*  Image registration = phase correlation
155
156
157
*  Fit limit = 1000
158
159
160
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
161
162
163
*  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
164
165
166
*  Stigmator correction = No (typical) Yes (if well calibrated)
167
168
169
*  Stigmator Defocus Min = 2e-06 meters
170
171
172
*  Stigmator Defocus Max = 4e-06 meters
173
174
175
176
177
178
*  Z Focus/Focus Sequence>
179
180
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
181
182
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
183
184
185
Select individual step in the list box and set the following parameters
186
187
188
189
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
190
|Stage_Wobble|sq|Stage Tilt|No|
191
|Z_to_Eucentric|hl|Beam Tilt|Yes|
192
|Manual_after|fc or hl*|Manual|Yes|
193
194
195
196
197
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
198
199
200
For "Stage_Wobble" Step:
201
*  Tilt = 2 degrees each direction
202
203
204
*  Image registration= phase correlation
205
206
207
*  Fit limit = N/A
208
209
210
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
211
212
213
*  Correction type = Stage Z
214
215
216
*  Check drift greater than 3e-10 meters/s = No
217
218
219
*  Stigmator correction = No
220
221
222
*  Stigmator Defocus Max/Min = N/A
223
224
225
226
227
228
For "Z_to_Eucentric" Step:
229
*  Tilt = 0.01 radian each direction
230
231
232
*  Image registration= phase correlation
233
234
235
*  Fit limit = 1500
236
237
238
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
239
240
241
*  Correction type = Stage Z
242
243
244
*  Reset defocus = Default
245
246
247
*  Check drift greater than 3e-10 meters/s = No
248
249
250
*  Stigmator correction = No
251
252
253
*  Stigmator Defocus Max/Min = N/A
254 2 Amber Herold
255
256
257
______
258
259
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
260
261
______