h1. Initial MSI application preference setup
2
3
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
4
5
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
6
7 5 Eric Hou
Table 16.1 Example MSI node setup
8
9 1 Amber Herold
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
10
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
11
|Grid|gr|modeled stage position|no|no|no|
12
|Square|sq|modeled stage position|yes|no|no|
13
|Hole|hl|modeled stage postion|yes|yes|yes|
14
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
15
|Exposure|en & ef|image shift|no|yes|yes|
16
|Focus|fc|image shift|no|no|yes|
17
18 5 Eric Hou
# *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
19 6 Eric Hou
 
20
Leave other setting options to default. These default options are:
21
 
22 1 Amber Herold
**  Wait 2.5 seconds before acquiring image
23 6 Eric Hou
 
24
**  Save image to database = yes
25
 
26
Remember to press the "+" button in order to assign a preset to a node._
27
 
28
#  Presets Manager/Settings>
29
 
30 1 Amber Herold
paus time between preset changes = 1 sec.
31 6 Eric Hou
 
32 1 Amber Herold
Cycle presets = yes
33 6 Eric Hou
 
34 1 Amber Herold
Optimize preset cycle = yes
35 6 Eric Hou
 
36 1 Amber Herold
Cycle Magnification Only = yes
37 6 Eric Hou
 
38 1 Amber Herold
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
39 6 Eric Hou
 
40
#  Grid Targeting/Settings>
41 1 Amber Herold
42
43
44
45
46
*  Select Preset = gr
47
48
49
*  Label =
50
51
52
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
53
54
55
*  Radius = 0.0009 m covers about the whole grid.
56
57
58
*  Mosaic center = stage center
59
60
61
62
63
64
65
66
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
67
68
69
*  Square Targeting/Mosaic Settings>
70
71
72
73
74
75
*  Calibration Parameter = stage model position
76
77
78
*  Scale Image = yes (default)
79
80
81
*  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
82
83
84
85
86
*  Hole Targeting/Settings and Exposure Targeting/Settings>
87
88
89
90
*  Allow for user verification of picked holes = yes
91
92
93
*  Queue up targets= no
94
95
96
*  Declare drift when queue is submitted = yes (default)
97
98
99
*  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
100
101
102
103
104
105
*  Focus/Settings> and Z Focus/Settings>
106
107
108
109
*  Acuire post-focus image = yes
110
111
112
*  Melt time = 0 second (should be customized for each grid)
113
114
115
116
117
118
*  Focus/Focus Sequence>
119
120
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
121
122
123
Select individual step in the list box and set the following parameters
124
125
126
127
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
128
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
129
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
130
|Manual_after|fc|Manual|N/A|Yes|
131
132
133
134
135
For both autofocus steps:
136
*  Tilt = 0.01 radian each direction
137
138
139
*  Image registration = phase correlation
140
141
142
*  Fit limit = 1000
143
144
145
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
146
147
148
*  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
149
150
151
*  Stigmator correction = No (typical) Yes (if well calibrated)
152
153
154
*  Stigmator Defocus Min = 2e-06 meters
155
156
157
*  Stigmator Defocus Max = 4e-06 meters
158
159
160
161
162
163
*  Z Focus/Focus Sequence>
164
165
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
166
167
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
168
169
170
Select individual step in the list box and set the following parameters
171
172
173
174
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
175
|Stage_Wobble|sq|Stage Tilt|No|
176
|Z_to_Eucentric|hl|Beam Tilt|Yes|
177
|Manual_after|fc or hl*|Manual|Yes|
178
179
180
181
182
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
183
184
185
For "Stage_Wobble" Step:
186
*  Tilt = 2 degrees each direction
187
188
189
*  Image registration= phase correlation
190
191
192
*  Fit limit = N/A
193
194
195
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
196
197
198
*  Correction type = Stage Z
199
200
201
*  Check drift greater than 3e-10 meters/s = No
202
203
204
*  Stigmator correction = No
205
206
207
*  Stigmator Defocus Max/Min = N/A
208
209
210
211
212
213
For "Z_to_Eucentric" Step:
214
*  Tilt = 0.01 radian each direction
215
216
217
*  Image registration= phase correlation
218
219
220
*  Fit limit = 1500
221
222
223
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
224
225
226
*  Correction type = Stage Z
227
228
229
*  Reset defocus = Default
230
231
232
*  Check drift greater than 3e-10 meters/s = No
233
234
235
*  Stigmator correction = No
236
237
238
*  Stigmator Defocus Max/Min = N/A
239 2 Amber Herold
240
241
242
______
243
244
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
245
246
______