h1. Initial MSI application preference setup
2
3
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
4
5
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
6
7 5 Eric Hou
Table 16.1 Example MSI node setup
8
9 1 Amber Herold
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
10
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
11
|Grid|gr|modeled stage position|no|no|no|
12
|Square|sq|modeled stage position|yes|no|no|
13
|Hole|hl|modeled stage postion|yes|yes|yes|
14
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
15
|Exposure|en & ef|image shift|no|yes|yes|
16
|Focus|fc|image shift|no|no|yes|
17
18 5 Eric Hou
# *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
19 6 Eric Hou
 
20
Leave other setting options to default. These default options are:
21
 
22 1 Amber Herold
**  Wait 2.5 seconds before acquiring image
23 6 Eric Hou
 
24
**  Save image to database = yes
25
 
26
Remember to press the "+" button in order to assign a preset to a node._
27
 
28
#  Presets Manager/Settings>
29
 
30 1 Amber Herold
paus time between preset changes = 1 sec.
31 6 Eric Hou
 
32 1 Amber Herold
Cycle presets = yes
33 6 Eric Hou
 
34 1 Amber Herold
Optimize preset cycle = yes
35 6 Eric Hou
 
36 1 Amber Herold
Cycle Magnification Only = yes
37 6 Eric Hou
 
38 1 Amber Herold
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
39 6 Eric Hou
 
40
#  Grid Targeting/Settings>
41 7 Eric Hou
 
42 1 Amber Herold
*  Select Preset = gr
43 7 Eric Hou
 
44 1 Amber Herold
*  Label =
45 7 Eric Hou
 
46 1 Amber Herold
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
47 7 Eric Hou
 
48 1 Amber Herold
*  Radius = 0.0009 m covers about the whole grid.
49 7 Eric Hou
 
50 1 Amber Herold
*  Mosaic center = stage center
51
52
53
54
55
56
57
58
*  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
59
60
61
*  Square Targeting/Mosaic Settings>
62
63
64
65
66
67
*  Calibration Parameter = stage model position
68
69
70
*  Scale Image = yes (default)
71
72
73
*  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
74
75
76
77
78
*  Hole Targeting/Settings and Exposure Targeting/Settings>
79
80
81
82
*  Allow for user verification of picked holes = yes
83
84
85
*  Queue up targets= no
86
87
88
*  Declare drift when queue is submitted = yes (default)
89
90
91
*  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
92
93
94
95
96
97
*  Focus/Settings> and Z Focus/Settings>
98
99
100
101
*  Acuire post-focus image = yes
102
103
104
*  Melt time = 0 second (should be customized for each grid)
105
106
107
108
109
110
*  Focus/Focus Sequence>
111
112
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
113
114
115
Select individual step in the list box and set the following parameters
116
117
118
119
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
120
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
121
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
122
|Manual_after|fc|Manual|N/A|Yes|
123
124
125
126
127
For both autofocus steps:
128
*  Tilt = 0.01 radian each direction
129
130
131
*  Image registration = phase correlation
132
133
134
*  Fit limit = 1000
135
136
137
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
138
139
140
*  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
141
142
143
*  Stigmator correction = No (typical) Yes (if well calibrated)
144
145
146
*  Stigmator Defocus Min = 2e-06 meters
147
148
149
*  Stigmator Defocus Max = 4e-06 meters
150
151
152
153
154
155
*  Z Focus/Focus Sequence>
156
157
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
158
159
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
160
161
162
Select individual step in the list box and set the following parameters
163
164
165
166
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
167
|Stage_Wobble|sq|Stage Tilt|No|
168
|Z_to_Eucentric|hl|Beam Tilt|Yes|
169
|Manual_after|fc or hl*|Manual|Yes|
170
171
172
173
174
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
175
176
177
For "Stage_Wobble" Step:
178
*  Tilt = 2 degrees each direction
179
180
181
*  Image registration= phase correlation
182
183
184
*  Fit limit = N/A
185
186
187
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
188
189
190
*  Correction type = Stage Z
191
192
193
*  Check drift greater than 3e-10 meters/s = No
194
195
196
*  Stigmator correction = No
197
198
199
*  Stigmator Defocus Max/Min = N/A
200
201
202
203
204
205
For "Z_to_Eucentric" Step:
206
*  Tilt = 0.01 radian each direction
207
208
209
*  Image registration= phase correlation
210
211
212
*  Fit limit = 1500
213
214
215
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
216
217
218
*  Correction type = Stage Z
219
220
221
*  Reset defocus = Default
222
223
224
*  Check drift greater than 3e-10 meters/s = No
225
226
227
*  Stigmator correction = No
228
229
230
*  Stigmator Defocus Max/Min = N/A
231 2 Amber Herold
232
233
234
______
235
236
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
237
238