Initial MSI application preference setup » History » Version 8
Eric Hou, 06/24/2010 03:20 PM
1 | 1 | Amber Herold | h1. Initial MSI application preference setup |
---|---|---|---|
2 | |||
3 | Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t |
||
4 | |||
5 | The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users. |
||
6 | |||
7 | 5 | Eric Hou | Table 16.1 Example MSI node setup |
8 | |||
9 | 1 | Amber Herold | |_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:| |
10 | |Grid Targeting|gr|N/A|N/A|N/A|N/A| |
||
11 | |Grid|gr|modeled stage position|no|no|no| |
||
12 | |Square|sq|modeled stage position|yes|no|no| |
||
13 | |Hole|hl|modeled stage postion|yes|yes|yes| |
||
14 | |Z Focus|hl or fc*|modeled stage position|no|no|yes| |
||
15 | |Exposure|en & ef|image shift|no|yes|yes| |
||
16 | |Focus|fc|image shift|no|no|yes| |
||
17 | |||
18 | 5 | Eric Hou | # *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy. |
19 | 6 | Eric Hou | |
20 | Leave other setting options to default. These default options are: |
||
21 | |
||
22 | 1 | Amber Herold | ** Wait 2.5 seconds before acquiring image |
23 | 6 | Eric Hou | |
24 | ** Save image to database = yes |
||
25 | |
||
26 | Remember to press the "+" button in order to assign a preset to a node._ |
||
27 | |
||
28 | # Presets Manager/Settings> |
||
29 | |
||
30 | 1 | Amber Herold | paus time between preset changes = 1 sec. |
31 | 6 | Eric Hou | |
32 | 1 | Amber Herold | Cycle presets = yes |
33 | 6 | Eric Hou | |
34 | 1 | Amber Herold | Optimize preset cycle = yes |
35 | 6 | Eric Hou | |
36 | 1 | Amber Herold | Cycle Magnification Only = yes |
37 | 6 | Eric Hou | |
38 | 1 | Amber Herold | The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef. |
39 | 6 | Eric Hou | |
40 | # Grid Targeting/Settings> |
||
41 | 7 | Eric Hou | |
42 | 1 | Amber Herold | * Select Preset = gr |
43 | 7 | Eric Hou | |
44 | 1 | Amber Herold | * Label = |
45 | 7 | Eric Hou | |
46 | 1 | Amber Herold | *Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred. |
47 | 7 | Eric Hou | |
48 | 1 | Amber Herold | * Radius = 0.0009 m covers about the whole grid. |
49 | 7 | Eric Hou | |
50 | 1 | Amber Herold | * Mosaic center = stage center |
51 | 8 | Eric Hou | |
52 | # Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration |
||
53 | |
||
54 | # Square Targeting/Mosaic Settings> |
||
55 | |
||
56 | 1 | Amber Herold | * Calibration Parameter = stage model position |
57 | 8 | Eric Hou | |
58 | 1 | Amber Herold | * Scale Image = yes (default) |
59 | 8 | Eric Hou | |
60 | 1 | Amber Herold | * Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list. |
61 | 8 | Eric Hou | |
62 | # Hole Targeting/Settings and Exposure Targeting/Settings> |
||
63 | |
||
64 | 1 | Amber Herold | * Allow for user verification of picked holes = yes |
65 | 8 | Eric Hou | |
66 | 1 | Amber Herold | * Queue up targets= no |
67 | 8 | Eric Hou | |
68 | 1 | Amber Herold | * Declare drift when queue is submitted = yes (default) |
69 | 8 | Eric Hou | |
70 | 1 | Amber Herold | * Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm) |
71 | 8 | Eric Hou | |
72 | # Focus/Settings> and Z Focus/Settings> |
||
73 | |
||
74 | 1 | Amber Herold | * Acuire post-focus image = yes |
75 | 8 | Eric Hou | |
76 | 1 | Amber Herold | * Melt time = 0 second (should be customized for each grid) |
77 | 8 | Eric Hou | |
78 | # Focus/Focus Sequence> |
||
79 | |
||
80 | 1 | Amber Herold | Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right. |
81 | 8 | Eric Hou | |
82 | 1 | Amber Herold | Select individual step in the list box and set the following parameters |
83 | 8 | Eric Hou | |
84 | 1 | Amber Herold | |_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:| |
85 | |Z_to_Eucentric|fa|Beam Tilt|Stage Z|No| |
||
86 | |Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes| |
||
87 | 8 | Eric Hou | |Manual_after|fc|Manual|N/A|Yes| |
88 | |
||
89 | 1 | Amber Herold | For both autofocus steps: |
90 | * Tilt = 0.01 radian each direction |
||
91 | 8 | Eric Hou | |
92 | 1 | Amber Herold | * Image registration = phase correlation |
93 | 8 | Eric Hou | |
94 | 1 | Amber Herold | * Fit limit = 1000 |
95 | 8 | Eric Hou | |
96 | 1 | Amber Herold | * Correct for delta Defocus/Z between 0 and 2e-5 meters |
97 | 8 | Eric Hou | |
98 | 1 | Amber Herold | * Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution). |
99 | 8 | Eric Hou | |
100 | 1 | Amber Herold | * Stigmator correction = No (typical) Yes (if well calibrated) |
101 | 8 | Eric Hou | |
102 | 1 | Amber Herold | * Stigmator Defocus Min = 2e-06 meters |
103 | 8 | Eric Hou | |
104 | 1 | Amber Herold | * Stigmator Defocus Max = 4e-06 meters |
105 | 8 | Eric Hou | |
106 | # Z Focus/Focus Sequence> |
||
107 | |
||
108 | 1 | Amber Herold | The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>. |
109 | |||
110 | Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right. |
||
111 | |||
112 | |||
113 | Select individual step in the list box and set the following parameters |
||
114 | |||
115 | |||
116 | |||
117 | |_.Step:|_.preset:|_.focus method:|_.Enabled:| |
||
118 | |Stage_Wobble|sq|Stage Tilt|No| |
||
119 | |Z_to_Eucentric|hl|Beam Tilt|Yes| |
||
120 | |Manual_after|fc or hl*|Manual|Yes| |
||
121 | |||
122 | |||
123 | |||
124 | |||
125 | *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy. |
||
126 | |||
127 | |||
128 | For "Stage_Wobble" Step: |
||
129 | * Tilt = 2 degrees each direction |
||
130 | |||
131 | |||
132 | * Image registration= phase correlation |
||
133 | |||
134 | |||
135 | * Fit limit = N/A |
||
136 | |||
137 | |||
138 | * Correct for delta Defocus/Z between 0 and 0.0002 meters |
||
139 | |||
140 | |||
141 | * Correction type = Stage Z |
||
142 | |||
143 | |||
144 | * Check drift greater than 3e-10 meters/s = No |
||
145 | |||
146 | |||
147 | * Stigmator correction = No |
||
148 | |||
149 | |||
150 | * Stigmator Defocus Max/Min = N/A |
||
151 | |||
152 | |||
153 | |||
154 | |||
155 | |||
156 | For "Z_to_Eucentric" Step: |
||
157 | * Tilt = 0.01 radian each direction |
||
158 | |||
159 | |||
160 | * Image registration= phase correlation |
||
161 | |||
162 | |||
163 | * Fit limit = 1500 |
||
164 | |||
165 | |||
166 | * Correct for delta Defocus/Z between 0 and 0.0002 meters |
||
167 | |||
168 | |||
169 | * Correction type = Stage Z |
||
170 | |||
171 | |||
172 | * Reset defocus = Default |
||
173 | |||
174 | |||
175 | * Check drift greater than 3e-10 meters/s = No |
||
176 | |||
177 | |||
178 | * Stigmator correction = No |
||
179 | |||
180 | |||
181 | * Stigmator Defocus Max/Min = N/A |
||
182 | 2 | Amber Herold | |
183 | |||
184 | |||
185 | ______ |
||
186 | |||
187 | [[MSI-Tomography Preferences|MSI-Tomography Preferences >]] |
||