Project

General

Profile

Initial MSI application preference setup » History » Version 8

Eric Hou, 06/24/2010 03:20 PM

1 1 Amber Herold
h1. Initial MSI application preference setup
2
3
Many of the preferences are matched to the function of the nodes and therefore are more like configuration. Starting from v1.5, the "Administrator" default preferences are loaded during the installation ready for standard MSI operation assuming the use of example preset names and properties. This means that as a new user, you don't have t
4
5
The following is the comprehensive list of what is set as default during the initial loading. In some cases, the reason behind the setting is also included. Each institution can alter these default values according to their particular need for all their new users.
6
7 5 Eric Hou
Table 16.1 Example MSI node setup
8
9 1 Amber Herold
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|
10
|Grid Targeting|gr|N/A|N/A|N/A|N/A|
11
|Grid|gr|modeled stage position|no|no|no|
12
|Square|sq|modeled stage position|yes|no|no|
13
|Hole|hl|modeled stage postion|yes|yes|yes|
14
|Z Focus|hl or fc*|modeled stage position|no|no|yes|
15
|Exposure|en & ef|image shift|no|yes|yes|
16
|Focus|fc|image shift|no|no|yes|
17
18 5 Eric Hou
# *Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on carbon film for better accuracy.
19 6 Eric Hou
 
20
Leave other setting options to default. These default options are:
21
 
22 1 Amber Herold
**  Wait 2.5 seconds before acquiring image
23 6 Eric Hou
 
24
**  Save image to database = yes
25
 
26
Remember to press the "+" button in order to assign a preset to a node._
27
 
28
#  Presets Manager/Settings>
29
 
30 1 Amber Herold
paus time between preset changes = 1 sec.
31 6 Eric Hou
 
32 1 Amber Herold
Cycle presets = yes
33 6 Eric Hou
 
34 1 Amber Herold
Optimize preset cycle = yes
35 6 Eric Hou
 
36 1 Amber Herold
Cycle Magnification Only = yes
37 6 Eric Hou
 
38 1 Amber Herold
The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
39 6 Eric Hou
 
40
#  Grid Targeting/Settings>
41 7 Eric Hou
 
42 1 Amber Herold
*  Select Preset = gr
43 7 Eric Hou
 
44 1 Amber Herold
*  Label =
45 7 Eric Hou
 
46 1 Amber Herold
*Enter a new label for each new atlas that is acquired for this session. The first atlas requires no label. This label will be included in the image file name; therefore, shorter names are preferred.
47 7 Eric Hou
 
48 1 Amber Herold
*  Radius = 0.0009 m covers about the whole grid.
49 7 Eric Hou
 
50 1 Amber Herold
*  Mosaic center = stage center
51 8 Eric Hou
 
52
#  Square Targeting/Settings> Ignore all settings. No need to set these general targeting finding configuration
53
 
54
#  Square Targeting/Mosaic Settings>
55
 
56 1 Amber Herold
*  Calibration Parameter = stage model position
57 8 Eric Hou
 
58 1 Amber Herold
*  Scale Image = yes (default)
59 8 Eric Hou
 
60 1 Amber Herold
*  Scale Image pixel size = 512 (or the size of the grid image that will be used to build the atlas). A larger value can be used in future loading of the tile list.
61 8 Eric Hou
 
62
#  Hole Targeting/Settings and Exposure Targeting/Settings>
63
 
64 1 Amber Herold
*  Allow for user verification of picked holes = yes
65 8 Eric Hou
 
66 1 Amber Herold
*  Queue up targets= no
67 8 Eric Hou
 
68 1 Amber Herold
*  Declare drift when queue is submitted = yes (default)
69 8 Eric Hou
 
70 1 Amber Herold
*  Skip auto picking of holes = yes (change to No after proper setup of the hole-finding algorithm)
71 8 Eric Hou
 
72
#  Focus/Settings> and Z Focus/Settings>
73
 
74 1 Amber Herold
*  Acuire post-focus image = yes
75 8 Eric Hou
 
76 1 Amber Herold
*  Melt time = 0 second (should be customized for each grid)
77 8 Eric Hou
 
78
#  Focus/Focus Sequence>
79
 
80 1 Amber Herold
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
81 8 Eric Hou
 
82 1 Amber Herold
Select individual step in the list box and set the following parameters
83 8 Eric Hou
 
84 1 Amber Herold
|_.Step:|_.preset:|_.focus method:|_.correction type:|_.Enabled:|
85
|Z_to_Eucentric|fa|Beam Tilt|Stage Z|No|
86
|Def_to_Eucentric|fa|Beam Tilt|Defocus|Yes|
87 8 Eric Hou
|Manual_after|fc|Manual|N/A|Yes| 
88
 
89 1 Amber Herold
For both autofocus steps:
90
*  Tilt = 0.01 radian each direction
91 8 Eric Hou
 
92 1 Amber Herold
*  Image registration = phase correlation
93 8 Eric Hou
 
94 1 Amber Herold
*  Fit limit = 1000
95 8 Eric Hou
 
96 1 Amber Herold
*  Correct for delta Defocus/Z between 0 and 2e-5 meters
97 8 Eric Hou
 
98 1 Amber Herold
*  Check drift greater than 3e-10 meters/s (should be customized by the user based on the exposure time and required resolution).
99 8 Eric Hou
 
100 1 Amber Herold
*  Stigmator correction = No (typical) Yes (if well calibrated)
101 8 Eric Hou
 
102 1 Amber Herold
*  Stigmator Defocus Min = 2e-06 meters
103 8 Eric Hou
 
104 1 Amber Herold
*  Stigmator Defocus Max = 4e-06 meters
105 8 Eric Hou
 
106
#  Z Focus/Focus Sequence>
107
 
108 1 Amber Herold
The setup here works well for holey carbon grids. Continuous carbon with negative stain may require a different Z Focus Sequence because its low contrast at intermediate mag. See the Chapter on <link linkend="MSI-R_auto_focus">MSI-Raster application regarding improving autofocus</link>.
109
110
Enter 2 focus step in the sequence list box by typing the name in the entry box and press "+" on its right.
111
112
113
Select individual step in the list box and set the following parameters
114
115
116
117
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
118
|Stage_Wobble|sq|Stage Tilt|No|
119
|Z_to_Eucentric|hl|Beam Tilt|Yes|
120
|Manual_after|fc or hl*|Manual|Yes|
121
122
123
124
125
*Use hl if your z focus target will be at the center of a hole where its fft image don't have obvious Thone rings. Use fc if the target is on the carbon film for better accuracy.
126
127
128
For "Stage_Wobble" Step:
129
*  Tilt = 2 degrees each direction
130
131
132
*  Image registration= phase correlation
133
134
135
*  Fit limit = N/A
136
137
138
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
139
140
141
*  Correction type = Stage Z
142
143
144
*  Check drift greater than 3e-10 meters/s = No
145
146
147
*  Stigmator correction = No
148
149
150
*  Stigmator Defocus Max/Min = N/A
151
152
153
154
155
156
For "Z_to_Eucentric" Step:
157
*  Tilt = 0.01 radian each direction
158
159
160
*  Image registration= phase correlation
161
162
163
*  Fit limit = 1500
164
165
166
*  Correct for delta Defocus/Z between 0 and 0.0002 meters
167
168
169
*  Correction type = Stage Z
170
171
172
*  Reset defocus = Default
173
174
175
*  Check drift greater than 3e-10 meters/s = No
176
177
178
*  Stigmator correction = No
179
180
181
*  Stigmator Defocus Max/Min = N/A
182 2 Amber Herold
183
184
185
______
186
187
[[MSI-Tomography Preferences|MSI-Tomography Preferences >]]
188
189
______