Thanks to David Mastronarde for providing his DM plug-in using socket connection

For your convenience, these are taken from SerialEM download site. Please match it with your Digital Micrograph version. If you have SerialEM installed, this is included.

DM version local copy of SerialEM SEMCCD description
2.2.x SEMCCD-GMS2-64.dll DM for 64-bit GMS 2.0 - 2.2
2.3.0 SEMCCD-GMS2.30-64.dll DM for 64-bit GMS 2.30
2.3.1, 2.3.2, 3.0 SEMCCD-GMS2.31-64.dll DM for 64-bit GMS 2.31 and above
3.31 and above SEMCCD-GMS3.31-64.dll DM for 64-bit GMS 3.31 and above

Note: DM = Digital Micrograph; GMS = Gatan Microscopy Suite
32bit versions also exit, but not for GMS 2.31 and above

Copy it to C:\Program Files\Gatan\Plugins\

Set the environment variable SERIALEMCCD_PORT to an unused port between 50000 and 60000 to avoid conflict with other programs. Avoid ports used by Leginon, especially 55555.