Install SerialEM DigitalMicrograph plugin » History » Revision 18
« Previous |
Revision 18/20
(diff)
| Next »
Anchi Cheng, 05/05/2017 02:26 PM
Thanks to David Mastronarde for providing his DM plug-in using socket connection
Note: If you have SerialEM installed on the same computer, you don't have to install this. The same Plugin can be used by both programs and share the same port¶
For your convenience, these are taken from SerialEM download site: http://bio3d.colorado.edu/ftp/SerialEM/ Please match it with your Digital Micrograph version. If you have SerialEM installed, this is included.
DM version | local copy of SerialEM SEMCCD | description |
2.2.x | SEMCCD-GMS2-64.dll | DM for 64-bit GMS 2.0 - 2.2 |
2.3.0 | SEMCCD-GMS2.30-64.dll | DM for 64-bit GMS 2.30 |
2.3.1, 2.3.2, 3.0 | SEMCCD-GMS2.31-64.dll | DM for 64-bit GMS 2.31 and above |
3.31 and above | SEMCCD-GMS3.31-64.dll | DM for 64-bit GMS 3.31 and above |
Note: DM = Digital Micrograph; GMS = Gatan Microscopy Suite
32bit versions also exit, but not for GMS 2.31 and above
Copy it to C:\ProgramData\Gatan\Plugins\
If you had SEMCCDxxxx.dll in C:\Program File\Gatan\Plugins, you should remove it
Add a new environment variable SERIALEMCCD_PORT if not exist already from SerialEM installation. Set the value to an unused port between 50000 and 60000 to avoid conflict with other programs. Avoid ports used by Leginon, especially not 55555.
Port 50000 or 50001 usually works.
Updated by Anchi Cheng over 7 years ago · 18 revisions