Project

General

Profile

Actions

Instrumentscfg for 30 » History » Revision 1

Revision 1/2 | Next »
Anchi Cheng, 10/31/2013 09:14 PM


Instrumentscfg for 30

TEM

cs is the coefficient of spherical aberration in meters. Find out from TEM manufacture the proper value for your tem.
For example

[tem]
class: fei.Tecnai
cs: 2.0e-3

Camera

height and width are number of pixels based on the orientation presented in Leginon image viewer
  • Some Tietz camera may have an odd physical dimension. Use TCL/EMMenu program to find out the good dimension used in Tietz's own software and entered in this configuration.

For example

[camera]
class: gatan.Gatan
zplane: 50
height: 4096
width: 4096

zplane is a number that gives the order of camera. If a camera with low zplane value is going to acquire an image, Leginon will attempt to retract all cameras at higher zplane values.

HEIGHT and WIDTH are not shown in these examples for simplicity
  • One computer for both tem and digital camera:
    [tem]
    class: fei.Tecnai
    cs: 2.0e-3
    [camera]
    class: gatan.Gatan
    zplane: 50
    
  • Separate computers for the two instruments: configure only the instrument reside on the particular computer
    • On the computer that controls the tem:
      [tem]
      class: fei.Tecnai
      cs: 2.0e-3
      
    • On the computer that controls the digital camera:
      [camera]
      class: gatan.Gatan
      zplane: 50
      
  • Two camera at the same plane:
    • On the computer that controls the digital camera:

      DE12 and DE Survey camera are retract/extend together, with the survey camera at off-axis. Therefore zplane should be set at the same value
      For example,

      [camera1]
      class: de.DE12
      zplane: 50
      
      [camera2]
      class: de.DESurvey
      zplane: 50
      

  • Gatan Orius camera that is installed typically with Gatan K2/Falcon camera is technically on the same plane but is not capable of insertion unless K2 is retracted. On the other hand, K2 camera insertion includes a retraction of Orius at lower level of the function call in DM. Therefore, Orius is considered as a camera at lower zplane.
    [camera1]
    class: dmsem.GatanK2Base
    zplane: 50
    
    [camera2]
    class: dmsem.GatanOrius
    zplane: 49
    

Updated by Anchi Cheng about 11 years ago · 1 revisions