Krios-K2 Presets used for 1.9 A mouse recombinant apoferritin reconstruction data collection at SEMC @ NYSBC

The same as Krios1-K2-NYSBC except

Probe Lens Series Magnification: Preset name: Image Shift (x,y): Dimension: Binning: Camera Dose (e/camera pixel/s): Exposure Time (ms): Specimen Dose Rate (e/A^2): Spot Size: Defocus (m): C2 (um): frame saving: Camera
micro SA 1300** hl Aligned 927x959 4 8 1200 0.01 11 -6e-5 70 no K2
nano SA 37000 en 0,0 3710x3838 1 4.2 e/pixel/s and ~1.12 um on the specimen 6000 58 6 (-1e-6 to -2e-6) 70 200 ms/frame K2

**This magnification is an add-on to standard SA lens series made for NYSBC Krios1. May not be available at other scopes.

  • Note: no defocal pair was acquired in this experiment. Therefore, no ef preset.

Updated by Anchi Cheng over 5 years ago ยท 6 revisions