Here is an example of presets used on our HT7800 TVIPS-XF416

Unbinned image pixel size at 120kx nominal magnification is 1.76 Å

LaB6 filament Operated at HT 100 kV

Magnification: probe mode: Preset name: Image Shift (x,y): Dimension: Binning: Beam coverage: Exposure Time (ms): Spot Size: Defocus (m):
300 low-mag gr Aligned 1024 4 max 100 1 0.0
1500 micro-hc sq Aligned 1024 4 1x CCD size 100 3 0.0
15000 micro-hr hl Aligned 1024 4 1x CCD 100 3 -5e-4
120000 micro-hr fc 0,0 1024 2 ~2x CCD 400 5 -5e-7
120000 micro-hr fa 0,0 1024 4 ~4x CCD 100 5 -2e-6
120000 micro-hr en 0,0 4096 1 ~4x CCD 100 5 -2e-6