MSI-Tomography Preferences » History » Version 2

Amber Herold, 04/27/2010 02:23 PM

1 1 Amber Herold
h1. MSI-Tomography Preferences
2
3
4
Many preferences for nodes in "MSI-Tomography" application are the same as in MSI with queuing option. The example here is based on a working condition at the Jensen Lab.
5
6
The following should be set up by the user "administrator" which will then be used as system-wide default preference for new users. Once the new user starts the application, he/she can customize his/her own preferences. Not all preference has independent function. Therefore if you change one preference, check for the preference(configuration) of other nodes that might depend on it.
7
8
9
10
h2. Nodes that acquires images/Settings> except Tomography node
11
12
13
14
This application is normally used exclusively with the queuing option. Therefore, "wait for a node to process the image" option is always off. Note that the Process target type is default in all nodes except "Tomography Preview".
15
16
17
18
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|_.Process target type:|
19
|Grid Targeting|gr|N/A|N/A|N/A|N/A|N/A|
20
|Grid|gr|modeled stage position|no|no|no|acquisition|
21
|Square Q|sq|modeled stage position|no|no|no|acquisition|
22
|Hole Q|hl|modeled stage postion|no|yes|yes|acquisition|
23
|Z Focus|hl|modeled stage position|no|no|yes|focus|
24
|Tomography Preview|preview|modeled stage position|no|no|yes|preview|
25
|Tomo Focus|fc|modeled stage position|no|no|yes|focus|
26
27
28
29
30
_Remember to press the "+" button in order to assign a preset to a node._
31
32
33
Leave other setting options to default. These default options are:
34
35
36
*  Wait 2.5 seconds before acquiring image
37
38
39
*  Save image to database = yes
40
41
42
*  Mover = Presets Manager
43
44
45
46
47
48
49
50
h2. Presets Manager
51
52
53
54
*  Defaults are used in settings
55
56
57
* The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. The example Tomography application uses gr->sq->hl->fa->fc->tomo->preview.
58
59
60
61
62
63
64
65
h2. Settings that are the same as in other MSI applications in queuing mode
66
67
68
69
* Grid Targeting
70
71
72
* Square Targeting
73
74
75
* Hole Targeting (Queuing=yes, User verification=yes)
76
77
78
* Tomography Targeting (set up as in Hole Targeting; Queuing=yes, User verification=yes)
79
80
81
* Z Focus (See next sections for an example focus sequence settings)
82
83
84
* Focus (See next sections for an example focus sequence settings)
85
86
87
88
89
90
91
92
h2. Z Focus node focus sequence example
93
94
95
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
96
97
98
Select individual step in the list box and set the following parameters
99
100
101
102
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
103
|Stage_Tilt_Auto|hl|Stage Tilt|Yes|
104
|Beam_Tilt_Auto|hl|Beam Tilt|Yes|
105
|Manual_after|fc|Manual|No|
106
107
108
109
110
For Stage_Tilt_Auto Step:
111
112
113
*  Tilt = 5 deg each direction
114
115
116
*  Image registration = phase correlation
117
118
119
*  Fit limit = 5000
120
121
122
*  Correct for delta Defocus/Z between 0 and 5e-5 meters
123
124
125
*  Correction type = Stage Z
126
127
128
*  Reset defocus = Default
129
130
131
*  Check drift = No.
132
133
134
*  Stigmator correction = No
135
136
137
*  Stigmator Defocus Min = N/A
138
139
140
*  Stigmator Defocus Max = N/A
141
142
143
144
145
For Beam_Tilt_Auto Step:
146
147
148
* Same as Stage_Tilt_Auto, except:
149
150
151
*  Tilt = 0.01 radian each direction
152
153
154
155
156
157
158
159
160
161
h2. Tomo Focus node focus sequence example
162
163
164
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
165
166
167
Select individual step in the list box and set the following parameters
168
169
170
171
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
172
|Stage_Tilt_Fine|fa|Stage Tilt|Yes|
173
|Beam_Tilt_Fine|fa|Beam Tilt|Yes|
174
|Manual_after|fc|Manual|No|
175
176
177
178
179
For Stage_Tilt_Fine Step:
180
181
182
*  Tilt = 1 deg each direction
183
184
185
*  Image registration = phase correlation
186
187
188
*  Fit limit = 5000
189
190
191
*  Correct for delta Defocus/Z between 1e-07 and 5e-5 meters
192
193
194
*  Correction type = Stage Z
195
196
197
*  Reset defocus = Default
198
199
200
*  Check drift = No.
201
202
203
*  Stigmator correction = No
204
205
206
*  Stigmator Defocus Min = N/A
207
208
209
*  Stigmator Defocus Max = N/A
210
211
212
213
214
For Beam_Tilt_Fine Step:
215
216
217
* Same as Stage_Tilt_Auto, except:
218
219
220
*  Tilt = 0.01 radian each direction
221
222
223
*  Correct for delta Defocus/Z between 0 and 5e-5 meters
224
225
226
*  Wait for drift to be less than 1.2e-9 m/s
227
228
229
230
231
232
233
234
235
236
h2. Tomography
237
238
239
Tomography has many unique settings:
240
241
242
*  General Acquisition Settings:
243
244
245
* Use "modeled stage position" to move to target.
246
247
248
* Wait 2.5 seconds before acquiring image
249
250
251
* Save image to database
252
253
254
* Float->Integer = No (There is a better settings in the lower panel of the node)
255
256
257
* Correct image = Yes
258
259
260
* Wait for a node to process the image = No
261
262
263
* Publish and wait for rejected targets = Yes (For focusing before acquisition)
264
265
266
* Publish and wait for the reference target = No (unless the beam is not stable, and a beam shift fix is required before most tilt series acquisition)
267
268
269
* Adjust targets for drift = Yes
270
271
272
* Declare drift between targets = Yes ( to ensure proper targets of multiple targets on the same lower mag hl image)
273
274
275
276
277
*  Acquisition Presets Order = tomo
278
279
280
*  Acquisition Mover Settings (See Session on <link linkend="mover">Different ways to move targets</link> in the MSI chapter for details on how to optimize these settings:
281
282
283
* Mover: Navigator
284
285
286
* Navigator Target Tolerance: 6e-08 m (Scope performance dependent. Use a larger value if the stage fails to reach the default value in acceptable, for example, 5, trials).
287
288
289
* Navigator Acceptable Tolerance: 2e-07 m (Targeting accuracy dependent)
290
291
292
* Final Image Shift = Yes
293
294
295
296
297
*  Tilt
298
299
300
* Parameters Min.,Max.,Start
301
302
303
* Step >0: start from "Start" tilt stepping to "Max" then from "Start" tilt stepping to "Min"
304
305
306
* Step <0: start from "Start" tilt stepping to "Min" then from "Start" tilt stepping to "Max"
307
308
309
310
311
*  Exposure
312
313
314
Tomography node divides the total dose by the number of tilt images it needs to acquire to calculate the real exposure time it uses, given the limit set by the user.
315
316
317
* Total dose=200 e-A^2
318
319
320
* Exposure time Min.=0.25 seconds
321
322
323
* Exposure time Max.=2 seconds
324
325
326
327
328
*  Before Collection
329
330
331
* Run buffer cycle=Yes
332
333
334
* Align zero loss peak=No (Yes if energy filter is used)
335
336
337
* Measure dose=Yes
338
339
340
341
342
*  Misc.
343
344
345
* Scale to convert to integer = Yes (Determine the scale factor by your camera sensitivity and the exposure intensity of each tilt image. See the session on Import Notes about Image Intensity Recorded through Tomography Node below)
346
347
348
* Consider images with less than "100" counts of the first image as obstructed. (Determined by camera sensitivity and conversion scale above since the number is post-conversion. In general, 50% of a normal untilted image intensity is reasonable)
349
350
351
* Abort if first half collected less than "90" % of images. (Reduce this number if fewer tilts are acceptable)
352
353
354
* The rest of the settings are various ways to detect the shifts of the images at adjacent tilts. Some are experimental. The most reliable is the Option "Use lpf in peak finding of tilt image correlation".
355
356
357
358
359
*  Model
360
361
362
The goniometer model that predicts the defocus drift as a result of offset goniometer tilt axis relative to the detector imaging axis can be dynamically fitted. The following parameters are used to initialize the model fitting of the goniometer tilt axis. See Session on low magnification model fitting for more details.
363
364
365
* The model can be initialized using existing model in the database either at the same magnification of the tomo preset, at the first available model at lower magnification, or a specified magnification for larger range of flexability, or custom values set in this node.
366
367
368
* Custom Tilt Axis Model is set separately for the two direction of the alpha tilt since many stages behaves differently as the weight of the stage lean on different part of the goniometer.
369
370
371
* During a dynamical model fitting, the fitting modifies the predicted stage z offset(z0) from the data at each tilt. To avoid false fitting that predicts an unreasonable z0, a limit can be added to allow a change of a physically possible value such as 2 um. This limit should be relaxed when the tilt axis model is unknown and is expected to be far from the initial guess.
372
373
374
* Keep the tilt axis parameters fixed = No if the stage tilt axis angle from y axis of the detector and the offset of the tilt axis from the center of the detectors are dynamically fitted during the tilt axis.
375
376
377
378
379
Once the goniometer model gives satisfactory results, Keep the tilt axis parameters fixed is recommended with the model set to be initialized with the model of "only this preset"
380
381
382
Defocus prediction smoothing is default to use shifts of 4 tilt images. A larger number can be set if the goniometer behavior is particularly jiggly.
383
384
385
386
387
388
389
390
h2. Dose Measurement and Align ZLP
391
392
393
394
Use defaults:
395
396
397
* Use "stage position" to move to the reference target
398
399
400
* Wait "3" seconds before performing request
401
402
403
* If reguest performed less than "900" seconds ago, ignore request
404 2 Amber Herold