MSI-Tomography Preferences » History » Version 3

Eric Hou, 06/24/2010 03:49 PM

1 1 Amber Herold
h1. MSI-Tomography Preferences
2
3
Many preferences for nodes in "MSI-Tomography" application are the same as in MSI with queuing option. The example here is based on a working condition at the Jensen Lab.
4
5
The following should be set up by the user "administrator" which will then be used as system-wide default preference for new users. Once the new user starts the application, he/she can customize his/her own preferences. Not all preference has independent function. Therefore if you change one preference, check for the preference(configuration) of other nodes that might depend on it.
6
7
h2. Nodes that acquires images/Settings> except Tomography node
8
9
This application is normally used exclusively with the queuing option. Therefore, "wait for a node to process the image" option is always off. Note that the Process target type is default in all nodes except "Tomography Preview".
10
11 3 Eric Hou
h3. Table 16.4. Example MSI-Tomography node setup (Except Tomography node)
12 1 Amber Herold
13
|_.Node name:|_.preset:|_.move type (to reach the preset) :|_.wait for a node to process the image:|_.publish and wait for rejected targets:|_.adjust target for shift:|_.Process target type:|
14
|Grid Targeting|gr|N/A|N/A|N/A|N/A|N/A|
15
|Grid|gr|modeled stage position|no|no|no|acquisition|
16
|Square Q|sq|modeled stage position|no|no|no|acquisition|
17
|Hole Q|hl|modeled stage postion|no|yes|yes|acquisition|
18
|Z Focus|hl|modeled stage position|no|no|yes|focus|
19
|Tomography Preview|preview|modeled stage position|no|no|yes|preview|
20
|Tomo Focus|fc|modeled stage position|no|no|yes|focus|
21
22
_Remember to press the "+" button in order to assign a preset to a node._
23
24
25
Leave other setting options to default. These default options are:
26
27
*  Wait 2.5 seconds before acquiring image
28
29
*  Save image to database = yes
30
31
*  Mover = Presets Manager
32
33
h2. Presets Manager
34
35
*  Defaults are used in settings
36
37
* The cycle order should match the sequence of data acquisition to minimize hysterisis. If the presets are imported from previous session, the cycle order is imported automatically. The example Tomography application uses gr->sq->hl->fa->fc->tomo->preview.
38
39
h2. Settings that are the same as in other MSI applications in queuing mode
40
41
* Grid Targeting
42
43
* Square Targeting
44
45
* Hole Targeting (Queuing=yes, User verification=yes)
46
47
* Tomography Targeting (set up as in Hole Targeting; Queuing=yes, User verification=yes)
48
49
* Z Focus (See next sections for an example focus sequence settings)
50
51
* Focus (See next sections for an example focus sequence settings)
52
53
h2. Z Focus node focus sequence example
54
55
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
56
57
Select individual step in the list box and set the following parameters
58
59 3 Eric Hou
h3. Table 16.5. Example MSI-Tomography Z Focus Steps
60 1 Amber Herold
61
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
62
|Stage_Tilt_Auto|hl|Stage Tilt|Yes|
63
|Beam_Tilt_Auto|hl|Beam Tilt|Yes|
64
|Manual_after|fc|Manual|No|
65
66
For Stage_Tilt_Auto Step:
67
68
*  Tilt = 5 deg each direction
69
70
*  Image registration = phase correlation
71
72
*  Fit limit = 5000
73
74
*  Correct for delta Defocus/Z between 0 and 5e-5 meters
75
76
*  Correction type = Stage Z
77
78
*  Reset defocus = Default
79
80
*  Check drift = No.
81
82
*  Stigmator correction = No
83
84
*  Stigmator Defocus Min = N/A
85
86
*  Stigmator Defocus Max = N/A
87
88
For Beam_Tilt_Auto Step:
89
90
* Same as Stage_Tilt_Auto, except:
91
92
*  Tilt = 0.01 radian each direction
93
94
h2. Tomo Focus node focus sequence example
95
96
Enter 3 focus steps in the sequence list box by typing the name in the entry box and press "+" on its right.
97
98
Select individual step in the list box and set the following parameters
99
100 3 Eric Hou
h3. Table 16.6. Example MSI-Tomography Tomo Focus Steps
101 1 Amber Herold
102
|_.Step:|_.preset:|_.focus method:|_.Enabled:|
103
|Stage_Tilt_Fine|fa|Stage Tilt|Yes|
104
|Beam_Tilt_Fine|fa|Beam Tilt|Yes|
105
|Manual_after|fc|Manual|No|
106
107
For Stage_Tilt_Fine Step:
108
109
*  Tilt = 1 deg each direction
110
111
*  Image registration = phase correlation
112
113
*  Fit limit = 5000
114
115
*  Correct for delta Defocus/Z between 1e-07 and 5e-5 meters
116
117
*  Correction type = Stage Z
118
119
*  Reset defocus = Default
120
121
*  Check drift = No.
122
123
*  Stigmator correction = No
124
125
*  Stigmator Defocus Min = N/A
126
127
*  Stigmator Defocus Max = N/A
128
129
For Beam_Tilt_Fine Step:
130
131
* Same as Stage_Tilt_Auto, except:
132
133 3 Eric Hou
**  Tilt = 0.01 radian each direction
134 1 Amber Herold
135 3 Eric Hou
**  Correct for delta Defocus/Z between 0 and 5e-5 meters
136 1 Amber Herold
137 3 Eric Hou
**  Wait for drift to be less than 1.2e-9 m/s
138 1 Amber Herold
139
h2. Tomography
140
141
Tomography has many unique settings:
142
143
*  General Acquisition Settings:
144
145
* Use "modeled stage position" to move to target.
146
147
* Wait 2.5 seconds before acquiring image
148
149
* Save image to database
150
151
* Float->Integer = No (There is a better settings in the lower panel of the node)
152
153
* Correct image = Yes
154
155
* Wait for a node to process the image = No
156
157
* Publish and wait for rejected targets = Yes (For focusing before acquisition)
158
159
* Publish and wait for the reference target = No (unless the beam is not stable, and a beam shift fix is required before most tilt series acquisition)
160
161
* Adjust targets for drift = Yes
162
163
* Declare drift between targets = Yes ( to ensure proper targets of multiple targets on the same lower mag hl image)
164
165
166
167
168
*  Acquisition Presets Order = tomo
169
170
171
*  Acquisition Mover Settings (See Session on <link linkend="mover">Different ways to move targets</link> in the MSI chapter for details on how to optimize these settings:
172
173
174
* Mover: Navigator
175
176
177
* Navigator Target Tolerance: 6e-08 m (Scope performance dependent. Use a larger value if the stage fails to reach the default value in acceptable, for example, 5, trials).
178
179
180
* Navigator Acceptable Tolerance: 2e-07 m (Targeting accuracy dependent)
181
182
183
* Final Image Shift = Yes
184
185
186
187
188
*  Tilt
189
190
191
* Parameters Min.,Max.,Start
192
193
194
* Step >0: start from "Start" tilt stepping to "Max" then from "Start" tilt stepping to "Min"
195
196
197
* Step <0: start from "Start" tilt stepping to "Min" then from "Start" tilt stepping to "Max"
198
199
200
201
202
*  Exposure
203
204
205
Tomography node divides the total dose by the number of tilt images it needs to acquire to calculate the real exposure time it uses, given the limit set by the user.
206
207
208
* Total dose=200 e-A^2
209
210
211
* Exposure time Min.=0.25 seconds
212
213
214
* Exposure time Max.=2 seconds
215
216
217
218
219
*  Before Collection
220
221
222
* Run buffer cycle=Yes
223
224
225
* Align zero loss peak=No (Yes if energy filter is used)
226
227
228
* Measure dose=Yes
229
230
231
232
233
*  Misc.
234
235
236
* Scale to convert to integer = Yes (Determine the scale factor by your camera sensitivity and the exposure intensity of each tilt image. See the session on Import Notes about Image Intensity Recorded through Tomography Node below)
237
238
239
* Consider images with less than "100" counts of the first image as obstructed. (Determined by camera sensitivity and conversion scale above since the number is post-conversion. In general, 50% of a normal untilted image intensity is reasonable)
240
241
242
* Abort if first half collected less than "90" % of images. (Reduce this number if fewer tilts are acceptable)
243
244
245
* The rest of the settings are various ways to detect the shifts of the images at adjacent tilts. Some are experimental. The most reliable is the Option "Use lpf in peak finding of tilt image correlation".
246
247
248
249
250
*  Model
251
252
253
The goniometer model that predicts the defocus drift as a result of offset goniometer tilt axis relative to the detector imaging axis can be dynamically fitted. The following parameters are used to initialize the model fitting of the goniometer tilt axis. See Session on low magnification model fitting for more details.
254
255
256
* The model can be initialized using existing model in the database either at the same magnification of the tomo preset, at the first available model at lower magnification, or a specified magnification for larger range of flexability, or custom values set in this node.
257
258
259
* Custom Tilt Axis Model is set separately for the two direction of the alpha tilt since many stages behaves differently as the weight of the stage lean on different part of the goniometer.
260
261
262
* During a dynamical model fitting, the fitting modifies the predicted stage z offset(z0) from the data at each tilt. To avoid false fitting that predicts an unreasonable z0, a limit can be added to allow a change of a physically possible value such as 2 um. This limit should be relaxed when the tilt axis model is unknown and is expected to be far from the initial guess.
263
264
265
* Keep the tilt axis parameters fixed = No if the stage tilt axis angle from y axis of the detector and the offset of the tilt axis from the center of the detectors are dynamically fitted during the tilt axis.
266
267
268
269
270
Once the goniometer model gives satisfactory results, Keep the tilt axis parameters fixed is recommended with the model set to be initialized with the model of "only this preset"
271
272
273
Defocus prediction smoothing is default to use shifts of 4 tilt images. A larger number can be set if the goniometer behavior is particularly jiggly.
274
275
276
277
278
279
280
281
h2. Dose Measurement and Align ZLP
282
283
284
285
Use defaults:
286
287
288
* Use "stage position" to move to the reference target
289
290
291
* Wait "3" seconds before performing request
292
293
294
* If reguest performed less than "900" seconds ago, ignore request
295 2 Amber Herold
296
297