h1. MSI Operation
2
3 6 Amber Herold
The following sections will describe how to operate the MSI application once set-up is complete. MSI operation can vary depending on the quality of the specimen in the microscope and how well the calibrations have been completed. Once squares are manually picked from the atlas, the entire acquisition process is capable of proceeding without human intervention. This sort of hands-off operation does depend on how well holes can be found in the images of squares and how well high magnification exposure and focus targets can be picked in the the image of a hole. If the stage is drifting, the Drift Monitor will not allow acquisition to continue until the drift has settled down. Manual intervention in the acquisition process can make the hole finders operate better. Manual check before and after the autofocus routine runs is also possible in the Focus node. Manually checking after the autofocus routine finishes is a way for the Leginon user to verify that the autofocus routine worked properly.
4 1 Amber Herold
5 6 Amber Herold
If the computer does not have a lot of memory, it is recommended that Leginon should be the only application running on the computer during an experiment because all the imaging and processing take up a lot of memory.
6 1 Amber Herold
7
8 6 Amber Herold
h2.  Start Leginon and MSI (if not already started)
9 1 Amber Herold
10
11 6 Amber Herold
1.  local> start-leginon.py
12 1 Amber Herold
13 6 Amber Herold
2.  Leginon Session> log in and select "Create session" if a new session is needed.
14 1 Amber Herold
15 6 Amber Herold
Create Session> Fill in needed information such as session name and comments. Select instrument and project, if project database is used. Image path is automatically selected.
16 1 Amber Herold
17 6 Amber Herold
3.  Leginon Session> Select the session you have just created or use a previous session to start.
18 1 Amber Herold
19 6 Amber Herold
4.  Leginon/Application/Run> Select the "MSI" application, the scope, and the local launcher. Click Run.
20 1 Amber Herold
21
22
23
h2.  Reload the Atlas (Only if Leginon is being restarted)
24
25 6 Amber Herold
*Important*
26
The Atlas should have already been created by following the [[MSI Quick-start|MSI everyday set-up checklist]].
27 1 Amber Herold
28 11 Anchi Cheng
{{include(reload atlas)}}
29 1 Amber Herold
30
31 8 Anchi Cheng
h2. Select Squares on the atlas
32 1 Amber Herold
33
Pick squares to collect data from by making selections on the atlas.
34
35
36 6 Amber Herold
1.  Leginon/Square Targeting> Refresh the target display by clicking !http://emg.nysbc.org/software/leginon/images/icons/refresh.png!.
37 1 Amber Herold
38 6 Amber Herold
2.  Leginon/Square Targeting> Click the "acquisition" selection button !http://emg.nysbc.org/software/leginon/images/icons/arrow.png!.
39 1 Amber Herold
40 6 Amber Herold
3.  Navigate around the atlas to find suitable squares for acquisition.
41 1 Amber Herold
42 16 Anchi Cheng
[[ Setup Auto Square Finder using example targets]]
43
44 6 Amber Herold
This step is purely empirical at this point (by eye). If using Quantifoil grids with ice, look for squares that have holes paler than the surrounding carbon. Although Leginon has automatic algorithm for square finding, the adjustment of parameters takes more time than what human can accomplish easily. Use the zoom to look at the holes more closely. Note that the zoom will take a long time on such a large image. Therefore...
45 1 Amber Herold
46
NOTE: Be sure to be patient when clicking the zoom button, otherwise Leginon will
47
take a while to catch up to the number of zooms that have been asked for.
48
49
50 6 Amber Herold
4.   Click on squares that will be acquired.
51 1 Amber Herold
52 6 Amber Herold
 *  red cross hair = previously targeted or "done" squares.
53 1 Amber Herold
54 6 Amber Herold
 *  green cross hair = new targets
55 1 Amber Herold
56 6 Amber Herold
 *  orange cross hair = current position (updated by pressing the Leginon/Square Targeting> Refresh button !http://emg.nysbc.org/software/leginon/images/icons/refresh.png!)
57 1 Amber Herold
58
59 6 Amber Herold
5.  Once all the targeted squares have been selected, click "Submit Targets."
60 1 Amber Herold
61 6 Amber Herold
6.  Leginon will indicate where the next step is by showing active green icons next to the nodes that are being used. A question mark icon will appear when Leginon requires user interation. The message log for each node indicates what is happening that particular node.
62 1 Amber Herold
63
64
65 9 Anchi Cheng
h2. Select Holes on the first image of a square
66 1 Amber Herold
67
68
69
Pick holes to collect data from by making selections on the first image of a square.
70
71
72 6 Amber Herold
1. Leginon/Hole Targeting> Check whether holes were selected (acquisition targets = green crosshair, focus target = blue crosshair).
73 1 Amber Herold
74 6 Amber Herold
2. Edit the targets by mouse left click (add) or right click (delete) on the target in the image viewer with the picking tool for the target type active.
75 1 Amber Herold
76 6 Amber Herold
3.  If the hole selection is acceptable, uncheck "Allow for user verification of picked holes" in Leginon/Hole Targeting/Settings>.
77 1 Amber Herold
78 6 Amber Herold
4.  If the hole selection is not acceptable, try adjusting the Hole Targeting node to find holes in the current image of a square. Refer to the [[MSI set-up in more details#Hole-Targeting-Set-up|Hole Targeting Set-up]].
79 1 Amber Herold
80 6 Amber Herold
5.  Click "Submit Targets"
81 1 Amber Herold
82
83
84
h2. Manually check focuser that uses Stage Z as the correction method
85
86
87 6 Amber Herold
Manually checking the focus after eucentric focusing of hl images is recommended for bent grids.
88 1 Amber Herold
89
90 6 Amber Herold
1.  When the Manual Focus window opens automatically (when Leginon reaches the Z Focus node), click "Eucentric focus to instrument" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png. The rings of the CTF in the FFT should disappear if it is looking at the carbon support using fc preset. If hl preset is used to look at a hole during manual focus, look for the loss of contrast in IMAGE.
91 1 Amber Herold
92 6 Amber Herold
2.  Adjust "Stage Z" until the image appears to be focused if not already so. Use a large step such as 5e-6 m.
93 1 Amber Herold
94 6 Amber Herold
3.  If the defocus is consistently off by more than +/- 5 um., then the [[Autofocus Calibration with Beam Tilt Node|Beam Tilt calibrations]] (defocus) may need to be repeated.
95 1 Amber Herold
96 6 Amber Herold
Make sure that the stage is at eucentric height in the calibration and that the [[Microscope Set-up|microscope has been well aligned]].
97 1 Amber Herold
98 6 Amber Herold
4.  Clicking the Stop button !http://emg.nysbc.org/software/leginon/images/icons/stop.png! exits the maunal check.
99 1 Amber Herold
100 6 Amber Herold
5.  Leginon/Z Focus/Focus Sequence> Disable "Manual_after" focus step when it is not required.
101 1 Amber Herold
102
103
104
h2.  Select Exposure and Focus targets on the first image of a hole
105
106
107
108
Pick exposure and focus targets on the first image of a hole.
109
110
111 6 Amber Herold
1.  Leginon/Exposure Targeting> Click the Original, acquisition, and focus selections in the image control panel.
112 1 Amber Herold
113 6 Amber Herold
2.  Check whether exposure (green crosshair) and focus (blue crosshair) targets were selected correctly.
114 1 Amber Herold
115 6 Amber Herold
3.  Edit the targets by mouse left click (add) or right click (delete) on the target in the image viewer with the picking tool for the target type active.
116 1 Amber Herold
117 6 Amber Herold
4.  If the exposure and focus selection is acceptable, uncheck "Allow for user verification of picked holes" in Leginon/Exposure Targeting/Settings>.
118 1 Amber Herold
119 6 Amber Herold
5.  If the exposure and focus selection is not acceptable, try adjusting the Hole Finder to pick the correct targets. Refer to [[MSI set-up in more details#Exposure-Targeting-Set-up|Exposure Targeting Set-up]].
120 1 Amber Herold
121 6 Amber Herold
6.  Click "Submit Targets"
122 1 Amber Herold
123
124
125
h2.  Manually check focuser that uses Defocus as the correction method
126
127
128
129 6 Amber Herold
Manually check the focus after the first high magnification images that are acquired. Then turn off "manual_after" focus step once it has been verified that the autofocus routine is properly correcting the defocus and astigmatism.
130 1 Amber Herold
131 15 Anchi Cheng
{{include(Manual Focusing Tool)}}
132
*  If the rings, after the direct setting to zero defocus in step 1, are still visable or are astigmatic in the FFT, then the [[Autofocus Calibration with Beam Tilt Node|Beam Tilt calibrations]] (defocus and stimator) may need to be repeated*
133
134
135
Make sure that the stage is at eucentric height and that the [[Microscope Set-up|microscope has been well aligned]].
136 1 Amber Herold
137 6 Amber Herold
4.  Leginon/Focus/Focus Sequence> Disable "Manual_after" focus step when it is not required.
138 1 Amber Herold
139
140
141
h2.  View the images online
142
143
144
145 6 Amber Herold
The images Leginon has acquired and published to the database can be viewed through the online viewers, the Leginon Image Viewer, the Leginon 3-Way Viewer, and the Leginon Observer Interface (LOI) which displays the currently acquired images.
146 1 Amber Herold
147 6 Amber Herold
1.  Go to http://yourhost/myamiweb and open either the Leginon Image Viewer, the Leginon 3-Way Viewer, or the LOI.
148 1 Amber Herold
149 6 Amber Herold
2.  Select today's Session from the pull-down list.
150 1 Amber Herold
151 6 Amber Herold
3.  Have fun browsing through your images! You can view the parent images of images selected in the main viewer of the 3-way viewer in the smaller views by specify the appropriate preset.
152 1 Amber Herold
153
154
155
156
h2.  Interruption
157
158
159
160 7 Amber Herold
Targets that are submitted can no longer be modified individually from the node where you selected them. If you need to abort the unprocessed targets in a list, you need to go to the node that receive it and click on the "Abort" button. In other words, if you want to abort all remaining holes in a square, you need to abort at the "Hole" node.
161 1 Amber Herold
162 7 Amber Herold
The same is true with pausing. Pausing is usually used when the cryo stage requires refill. To allow proper tracking of targets and drift, we recommend that you pause at "Hole" node for the purpose if exposure targets are not queued.
163 5 Anchi Cheng
164 6 Amber Herold
More information about interrupting data collection can be found in the [[Pausing and Aborting during data collection|Trouble Shooting Chapter regarding pausing and aborting during data collection]]
165 2 Amber Herold
166
167
______
168
169 13 Anchi Cheng
[[Special Operation setup|< Special Operation setup]] | [[Optimize Autofocusing|Optimizing autofocusing >]]
170 2 Amber Herold
171
______