MSI Operation » History » Version 3
Anchi Cheng, 06/18/2010 03:55 PM
1 | 1 | Amber Herold | h1. MSI Operation |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | h2. Start Leginon & MSI (if not already started) |
||
7 | |||
8 | |||
9 | |||
10 | |||
11 | |||
12 | * local> start-leginon.py |
||
13 | |||
14 | |||
15 | * Leginon Session> log in and select "Create session" if a new session is |
||
16 | needed. |
||
17 | |||
18 | |||
19 | Create Session> Fill in needed information such as session name and comments. |
||
20 | Select instrument and project, if project database is used. Image path is |
||
21 | automatically selected. |
||
22 | |||
23 | |||
24 | * Leginon Session> Select the session you have just created or use a previous |
||
25 | session to start. |
||
26 | |||
27 | |||
28 | * Leginon/Application/Run> Select the "MSI" application, the scope, and the |
||
29 | local launcher. Click Run. |
||
30 | |||
31 | |||
32 | |||
33 | |||
34 | |||
35 | |||
36 | |||
37 | |||
38 | |||
39 | h2. Reload the Atlas (Only if Leginon is being restarted) |
||
40 | |||
41 | |||
42 | |||
43 | |||
44 | <important> |
||
45 | The Atlas should have already been created by following the <link |
||
46 | linkend="MISQuickStart">MSI everyday set-up checklist</link>. |
||
47 | |||
48 | </important> |
||
49 | |||
50 | |||
51 | Reload the atlas image of the grid. This is used if Leginon was stopped and |
||
52 | restarted. |
||
53 | |||
54 | |||
55 | |||
56 | |||
57 | * Leginon/Square Targeting/Mosaic <inlinegraphic |
||
58 | fileref="http://emg.nysbc.org/software/leginon/images/icons/atlasmaker.png" |
||
59 | format="PNG"/>> Increase the size of the mosaic atlas loaded into the image |
||
60 | display for better resolution. For example, Scale image to: 2048 |
||
61 | |||
62 | |||
63 | * Leginon/Square Targeting/Tiles<inlinegraphic |
||
64 | fileref="http://emg.nysbc.org/software/leginon/images/icons/tiles.png" |
||
65 | format="PNG"/>> Select the atlas label in "Load tiles from mosaic." |
||
66 | |||
67 | |||
68 | * Leginon/Square Targeting/Tiles> Click "Load" |
||
69 | |||
70 | |||
71 | |||
72 | |||
73 | |||
74 | |||
75 | |||
76 | |||
77 | |||
78 | h2. Select Squares on the atlas |
||
79 | |||
80 | |||
81 | |||
82 | Pick squares to collect data from by making selections on the atlas. |
||
83 | |||
84 | |||
85 | |||
86 | |||
87 | * Leginon/Square Targeting> Refresh the target display by clicking |
||
88 | <inlinegraphic |
||
89 | fileref="http://emg.nysbc.org/software/leginon/images/icons/refresh.png" |
||
90 | format="PNG"/>. |
||
91 | |||
92 | |||
93 | * Leginon/Square Targeting> Click the "acquisition" selection |
||
94 | button<inlinegraphic |
||
95 | fileref="http://emg.nysbc.org/software/leginon/images/icons/arrow.png" |
||
96 | format="PNG"/>. |
||
97 | |||
98 | |||
99 | * Navigate around the atlas to find suitable squares for acquisition. |
||
100 | |||
101 | This step is purely empirical at this point (by eye). If using Quantifoil grids |
||
102 | with ice, look for squares that have holes paler than the surrounding carbon. Although |
||
103 | Leginon has automatic algorithm for square finding, the adjustment of parameters takes |
||
104 | more time than what human can accomplish easily. Use the zoom to look at the holes |
||
105 | more closely. Note that the zoom will take a long time on such a large image. |
||
106 | Therefore... |
||
107 | |||
108 | |||
109 | NOTE: Be sure to be patient when clicking the zoom button, otherwise Leginon will |
||
110 | take a while to catch up to the number of zooms that have been asked for. |
||
111 | |||
112 | |||
113 | |||
114 | |||
115 | * Click on squares that will be acquired. |
||
116 | |||
117 | |||
118 | |||
119 | |||
120 | * red cross hair = previously targeted or "done" squares. |
||
121 | |||
122 | |||
123 | * green cross hair = new targets |
||
124 | |||
125 | |||
126 | * orange cross hair = current postion (updated by pressing the Leginon/Square |
||
127 | Targeting> Refresh button<inlinegraphic |
||
128 | fileref="http://emg.nysbc.org/software/leginon/images/icons/refresh.png" |
||
129 | format="PNG"/>) |
||
130 | |||
131 | |||
132 | |||
133 | |||
134 | |||
135 | |||
136 | * Once all the targeted squares have been selected, click "Submit Targets." |
||
137 | |||
138 | |||
139 | * Leginon will indicate where the next step is by showing active green icons next |
||
140 | to the nodes that are being used. A question mark icon will appear when Leginon |
||
141 | requires user interation. The message log for each node indicates what is happening |
||
142 | that particular node. |
||
143 | |||
144 | |||
145 | |||
146 | |||
147 | |||
148 | |||
149 | |||
150 | |||
151 | |||
152 | h2. Select Holes on the first image of a square |
||
153 | |||
154 | |||
155 | |||
156 | Pick holes to collect data from by making selections on the first image of a square. |
||
157 | |||
158 | |||
159 | |||
160 | |||
161 | * Leginon/Hole Targeting> Check whether holes were selected (<anchor |
||
162 | id="MSI_target_color"/>acquisition targets = green crosshair, focus target = blue |
||
163 | crosshair). |
||
164 | |||
165 | |||
166 | * Edit the targets by mouse left click (add) or right |
||
167 | click (delete) on the target in the image viewer with the picking tool for the target |
||
168 | type active. |
||
169 | |||
170 | |||
171 | * If the hole selection is acceptable, uncheck "Allow for user verification of |
||
172 | picked holes" in Leginon/Hole Targeting/Settings>. |
||
173 | |||
174 | |||
175 | * If the hole selection is not acceptable, try adjusting the Hole Targeting node to |
||
176 | find holes in the current image of a square. Refer to the <link |
||
177 | linkend="holetargeting_setup">Hole Targeting Set-up</link>. |
||
178 | |||
179 | |||
180 | * Click "Submit Targets" |
||
181 | |||
182 | |||
183 | |||
184 | |||
185 | |||
186 | |||
187 | |||
188 | |||
189 | |||
190 | 3 | Anchi Cheng | h2. Manually check focuser that uses Stage Z as the correction method |
191 | 1 | Amber Herold | |
192 | |||
193 | |||
194 | Manually checking the focus after eucentric focusing of hl images is recommended for |
||
195 | bent grids. |
||
196 | |||
197 | |||
198 | |||
199 | |||
200 | * When the Manual Focus window opens automatically (when Leginon reaches the Z |
||
201 | Focus node), click "Eucentric focus to instrument"<inlinegraphic |
||
202 | fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png" |
||
203 | format="PNG"/>. The rings of the CTF in the FFT should disappear if it is looking at |
||
204 | the carbon support using fc preset. If hl preset is used to look at a hole during |
||
205 | manual focus, look for the loss of contrast in IMAGE. |
||
206 | |||
207 | |||
208 | * Adjust "Stage Z" until the image appears to be focused if not already so. Use a |
||
209 | large step such as 5e-6 m. |
||
210 | |||
211 | |||
212 | * If the defocus is consistently off by more than +/- 5 um., then the <link |
||
213 | linkend="btcal_def">Beam Tilt calibrations</link> (defocus) may need to be |
||
214 | repeated. |
||
215 | |||
216 | |||
217 | Make sure that the stage is at eucentric height in the calibration and that the |
||
218 | <link linkend="microscope_setup">microscope has been well aligned</link>. |
||
219 | |||
220 | |||
221 | * Clicking the Stop button<inlinegraphic |
||
222 | fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png" format="PNG" |
||
223 | /> exits the maunal check. |
||
224 | |||
225 | |||
226 | * Leginon/Z Focus/Focus Sequence> Disable "Manual_after" focus step when it is |
||
227 | not required. |
||
228 | |||
229 | |||
230 | |||
231 | |||
232 | |||
233 | |||
234 | |||
235 | |||
236 | |||
237 | h2. Select Exposure and Focus targets on the first image of a hole |
||
238 | |||
239 | |||
240 | |||
241 | Pick exposure and focus targets on the first image of a hole. |
||
242 | |||
243 | |||
244 | |||
245 | |||
246 | * Leginon/Exposure Targeting> Click the Original, acquisition, and focus |
||
247 | selections in the image control panel. |
||
248 | |||
249 | |||
250 | * Check whether exposure (green crosshair) and focus (blue crosshair) targets were |
||
251 | selected correctly. |
||
252 | |||
253 | |||
254 | * Edit the targets by mouse left click (add) or right click (delete) on the target |
||
255 | in the image viewer with the picking tool for the target type active. |
||
256 | |||
257 | |||
258 | * If the exposure and focus selection is acceptable, uncheck "Allow for user |
||
259 | verification of picked holes" in Leginon/Exposure Targeting/Settings>. |
||
260 | |||
261 | |||
262 | * If the exposure and focus selection is not acceptable, try adjusting the Hole |
||
263 | Finder to pick the correct targets. Refer to <link linkend="exptargeting_setup" |
||
264 | >Exposure Targeting Set-up</link> |
||
265 | |||
266 | |||
267 | * Click "Submit Targets" |
||
268 | |||
269 | |||
270 | |||
271 | |||
272 | |||
273 | |||
274 | |||
275 | |||
276 | |||
277 | h2. Manually check focuser that uses Defocus as the correction method |
||
278 | |||
279 | |||
280 | |||
281 | Manually check the focus after the first high magnification images that are acquired. |
||
282 | Then turn off "manual_after" focus step once it has been verified that the autofocus routine |
||
283 | is properly correcting the defocus and astigmatism. |
||
284 | |||
285 | |||
286 | |||
287 | |||
288 | * When the Manual Focus window opens automatically (when Leginon reaches the Focus |
||
289 | node), click "Set Instrument"<inlinegraphic |
||
290 | fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png" |
||
291 | format="PNG"/> to send what Leginon thinks is zero defocus to the microscope. The |
||
292 | rings of the CTF in the FFT should disappear once the defocus is at zero. |
||
293 | |||
294 | |||
295 | * If the rings are still visable or are astigmatic in the FFT, then the <link |
||
296 | linkend="btcal_def">Beam Tilt calibrations</link> (defocus and stimator) may need to |
||
297 | be repeated. |
||
298 | |||
299 | |||
300 | Make sure that the stage is at eucentric height and that the <link |
||
301 | linkend="microscope_setup">microscope has been well aligned</link>. |
||
302 | |||
303 | |||
304 | * Clicking the Stop button<inlinegraphic |
||
305 | fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png" format="PNG" |
||
306 | /> exits the maunal check. |
||
307 | |||
308 | |||
309 | * Leginon/Focus/Focus Sequence> Disable "Manual_after" focus step when it is not |
||
310 | required. |
||
311 | |||
312 | |||
313 | |||
314 | |||
315 | |||
316 | |||
317 | |||
318 | |||
319 | |||
320 | h2. View the images online |
||
321 | |||
322 | |||
323 | |||
324 | The images Leginon has acquired and published to the database can be viewed through the |
||
325 | online viewers, the Leginon Image Viewer, the Leginon 3-Way Viewer, and the Leginon Observer |
||
326 | Interface (LOI) which displays the currently acquired images. |
||
327 | |||
328 | |||
329 | |||
330 | |||
331 | * Go to http://yourhost/dbem and open either the Leginon Image Viewer, the Leginon |
||
332 | 3-Way Viewer, or the LOI. |
||
333 | |||
334 | |||
335 | * Select today's Session from the pull-down list. |
||
336 | |||
337 | |||
338 | * Have fun browsing through your images! You can view the parent images of images |
||
339 | selected in the main viewer of the 3-way viewer in the smaller views by specify the |
||
340 | appropriate preset. |
||
341 | |||
342 | |||
343 | |||
344 | |||
345 | |||
346 | |||
347 | |||
348 | |||
349 | |||
350 | h2. Interruption |
||
351 | |||
352 | |||
353 | |||
354 | Targets that are submitted can no longer be modified individually from the node where |
||
355 | you selected them. If you need to abort the unprocessed targets in a list, you need to go to |
||
356 | the node that recieve it and click on the "Abort" button. In other words, if you want to |
||
357 | abort all remaining holes in a square, you need to abort at the "Hole" node. |
||
358 | |||
359 | |||
360 | Same is true with pausing. Pausing is usually used when the cryo stage requires refill. |
||
361 | To allow proper tracking of targets and drift, we recommand that you pause at "Hole" node |
||
362 | for the purpose if exposure targets are not queued. |
||
363 | |||
364 | |||
365 | More information about interrupting data collection can be found in the <link |
||
366 | linkend="MSI_pause">Trouble Shooting Chapter regarding pausing and aborting during data |
||
367 | collection</link>. |
||
368 | 2 | Amber Herold | |
369 | |||
370 | ______ |
||
371 | |||
372 | [[Special Operation setup|< Special Operation setup]] | [[Queuing option|Queuing option >]] |
||
373 | |||