h2.  Start Leginon & MSI (if not already started)
7
8
9
10
11
12
*  local> start-leginon.py
13
14
15
*  Leginon Session> log in and select "Create session" if a new session is
16
needed.
17
18
19
Create Session> Fill in needed information such as session name and comments.
20
Select instrument and project, if project database is used. Image path is
21
automatically selected.
22
23
24
*  Leginon Session> Select the session you have just created or use a previous
25
session to start.
26
27
28
*  Leginon/Application/Run> Select the "MSI" application, the scope, and the
29
local launcher. Click Run.
30
31
32
33
34
35
36
37
38
39
h2.  Reload the Atlas (Only if Leginon is being restarted)
40
41
42
43
44
<important>
45
The Atlas should have already been created by following the <link
46
linkend="MISQuickStart">MSI everyday set-up checklist</link>.
47
48
</important>
49
50
51
Reload the atlas image of the grid. This is used if Leginon was stopped and
52
restarted.
53
54
55
56
57
*  Leginon/Square Targeting/Mosaic <inlinegraphic
58
fileref="http://emg.nysbc.org/software/leginon/images/icons/atlasmaker.png"
59
format="PNG"/>> Increase the size of the mosaic atlas loaded into the image
60
display for better resolution. For example, Scale image to: 2048
61
62
63
*  Leginon/Square Targeting/Tiles<inlinegraphic
64
fileref="http://emg.nysbc.org/software/leginon/images/icons/tiles.png"
65
format="PNG"/>> Select the atlas label in "Load tiles from mosaic."
66
67
68
*  Leginon/Square Targeting/Tiles> Click "Load"
69
70
71
72
73
74
75
76
77
78
h2.  Select Squares on the atlas
79
80
81
82
Pick squares to collect data from by making selections on the atlas.
83
84
85
86
87
*  Leginon/Square Targeting> Refresh the target display by clicking
88
<inlinegraphic
89
fileref="http://emg.nysbc.org/software/leginon/images/icons/refresh.png"
90
format="PNG"/>.
91
92
93
*  Leginon/Square Targeting> Click the "acquisition" selection
94
button<inlinegraphic
95
fileref="http://emg.nysbc.org/software/leginon/images/icons/arrow.png"
96
format="PNG"/>.
97
98
99
*  Navigate around the atlas to find suitable squares for acquisition.
100
101
This step is purely empirical at this point (by eye). If using Quantifoil grids
102
with ice, look for squares that have holes paler than the surrounding carbon. Although
103
Leginon has automatic algorithm for square finding, the adjustment of parameters takes
104
more time than what human can accomplish easily. Use the zoom to look at the holes
105
more closely. Note that the zoom will take a long time on such a large image.
106
Therefore...
107
108
109
NOTE: Be sure to be patient when clicking the zoom button, otherwise Leginon will
110
take a while to catch up to the number of zooms that have been asked for.
111
112
113
114
115
*  Click on squares that will be acquired.
116
117
118
119
120
*  red cross hair = previously targeted or "done" squares.
121
122
123
*  green cross hair = new targets
124
125
126
*  orange cross hair = current postion (updated by pressing the Leginon/Square
127
Targeting> Refresh button<inlinegraphic
128
fileref="http://emg.nysbc.org/software/leginon/images/icons/refresh.png"
129
format="PNG"/>)
130
131
132
133
134
135
136
*  Once all the targeted squares have been selected, click "Submit Targets."
137
138
139
*  Leginon will indicate where the next step is by showing active green icons next
140
to the nodes that are being used. A question mark icon will appear when Leginon
141
requires user interation. The message log for each node indicates what is happening
142
that particular node.
143
144
145
146
147
148
149
150
151
152
h2.  Select Holes on the first image of a square
153
154
155
156
Pick holes to collect data from by making selections on the first image of a square.
157
158
159
160
161
* Leginon/Hole Targeting> Check whether holes were selected (<anchor
162
id="MSI_target_color"/>acquisition targets = green crosshair, focus target = blue
163
crosshair).
164
165
166
* Edit the targets by mouse left click (add) or right
167
click (delete) on the target in the image viewer with the picking tool for the target
168
type active.
169
170
171
*  If the hole selection is acceptable, uncheck "Allow for user verification of
172
picked holes" in Leginon/Hole Targeting/Settings>.
173
174
175
*  If the hole selection is not acceptable, try adjusting the Hole Targeting node to
176
find holes in the current image of a square. Refer to the <link
177
linkend="holetargeting_setup">Hole Targeting Set-up</link>.
178
179
180
*  Click "Submit Targets"
181
182
183
184
185
186
187
188
189
190 3 Anchi Cheng
h2. Manually check focuser that uses Stage Z as the correction method
191 1 Amber Herold
192
193
194
Manually checking the focus after eucentric focusing of hl images is recommended for
195
bent grids.
196
197
198
199
200
*  When the Manual Focus window opens automatically (when Leginon reaches the Z
201
Focus node), click "Eucentric focus to instrument"<inlinegraphic
202
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
203
format="PNG"/>. The rings of the CTF in the FFT should disappear if it is looking at
204
the carbon support using fc preset. If hl preset is used to look at a hole during
205
manual focus, look for the loss of contrast in IMAGE.
206
207
208
*  Adjust "Stage Z" until the image appears to be focused if not already so. Use a
209
large step such as 5e-6 m.
210
211
212
*  If the defocus is consistently off by more than +/- 5 um., then the <link
213
linkend="btcal_def">Beam Tilt calibrations</link> (defocus) may need to be
214
repeated.
215
216
217
Make sure that the stage is at eucentric height in the calibration and that the
218
<link linkend="microscope_setup">microscope has been well aligned</link>.
219
220
221
*  Clicking the Stop button<inlinegraphic
222
fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png" format="PNG"
223
/> exits the maunal check.
224
225
226
*  Leginon/Z Focus/Focus Sequence> Disable "Manual_after" focus step when it is
227
not required.
228
229
230
231
232
233
234
235
236
237
h2.  Select Exposure and Focus targets on the first image of a hole
238
239
240
241
Pick exposure and focus targets on the first image of a hole.
242
243
244
245
246
*  Leginon/Exposure Targeting> Click the Original, acquisition, and focus
247
selections in the image control panel.
248
249
250
*  Check whether exposure (green crosshair) and focus (blue crosshair) targets were
251
selected correctly.
252
253
254
*  Edit the targets by mouse left click (add) or right click (delete) on the target
255
in the image viewer with the picking tool for the target type active.
256
257
258
*  If the exposure and focus selection is acceptable, uncheck "Allow for user
259
verification of picked holes" in Leginon/Exposure Targeting/Settings>.
260
261
262
*  If the exposure and focus selection is not acceptable, try adjusting the Hole
263
Finder to pick the correct targets. Refer to <link linkend="exptargeting_setup"
264
>Exposure Targeting Set-up</link>
265
266
267
* Click "Submit Targets"
268
269
270
271
272
273
274
275
276
277
h2.  Manually check focuser that uses Defocus as the correction method
278
279
280
281
Manually check the focus after the first high magnification images that are acquired.
282
Then turn off "manual_after" focus step once it has been verified that the autofocus routine
283
is properly correcting the defocus and astigmatism.
284
285
286
287
288
*  When the Manual Focus window opens automatically (when Leginon reaches the Focus
289
node), click "Set Instrument"<inlinegraphic
290
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
291
format="PNG"/> to send what Leginon thinks is zero defocus to the microscope. The
292
rings of the CTF in the FFT should disappear once the defocus is at zero.
293
294
295
*  If the rings are still visable or are astigmatic in the FFT, then the <link
296
linkend="btcal_def">Beam Tilt calibrations</link> (defocus and stimator) may need to
297
be repeated.
298
299
300
Make sure that the stage is at eucentric height and that the <link
301
linkend="microscope_setup">microscope has been well aligned</link>.
302
303
304
*  Clicking the Stop button<inlinegraphic
305
fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png" format="PNG"
306
/> exits the maunal check.
307
308
309
*  Leginon/Focus/Focus Sequence> Disable "Manual_after" focus step when it is not
310
required.
311
312
313
314
315
316
317
318
319
320
h2.  View the images online
321
322
323
324
The images Leginon has acquired and published to the database can be viewed through the
325
online viewers, the Leginon Image Viewer, the Leginon 3-Way Viewer, and the Leginon Observer
326
Interface (LOI) which displays the currently acquired images.
327
328
329
330
331 4 Anchi Cheng
*  Go to http://yourhost/myamiweb and open either the Leginon Image Viewer, the Leginon
332 1 Amber Herold
3-Way Viewer, or the LOI.
333
334
335
*  Select today's Session from the pull-down list.
336
337
338
*  Have fun browsing through your images! You can view the parent images of images
339
selected in the main viewer of the 3-way viewer in the smaller views by specify the
340
appropriate preset.
341
342
343
344
345
346
347
348
349
350
h2.  Interruption
351
352
353
354
Targets that are submitted can no longer be modified individually from the node where
355
you selected them. If you need to abort the unprocessed targets in a list, you need to go to
356
the node that recieve it and click on the "Abort" button. In other words, if you want to
357
abort all remaining holes in a square, you need to abort at the "Hole" node.
358
359
360
Same is true with pausing. Pausing is usually used when the cryo stage requires refill.
361
To allow proper tracking of targets and drift, we recommand that you pause at "Hole" node
362
for the purpose if exposure targets are not queued.
363
364
365 5 Anchi Cheng
More information about interrupting data collection can be found in the [[Pausing and Aborting during data collection|Trouble Shooting Chapter regarding pausing and aborting during data
366
collection]]
367 2 Amber Herold
368
369
______
370
371
[[Special Operation setup|< Special Operation setup]] | [[Queuing option|Queuing option >]]
372
373
______