Project

General

Profile

MSI set up in more details » History » Version 11

Amber Herold, 06/22/2010 10:40 AM

1 1 Amber Herold
h1. MSI set-up in more details
2
3
4 10 Amber Herold
This section shows some of the more important and complicated steps in the set-up checklist. It assumes that the user knows the nomenclatures. See the full documentation if you want the detailed explanation. It also assumes that general set-up and calibrations exist from previous sessions so that most calibrations can remain unchanged. The steps that are NRAMM specific are marked. MSI now features stage Eucentric height correction at each square.
5 1 Amber Herold
6 10 Amber Herold
NOTE: The Tietz imaging software (EMMenu) must be turned OFF before Leginon is started. In contrast to this, the Gatan imaging software (Digital Micrograph) must be ON before Leginon is started.
7 1 Amber Herold
8
9
10 10 Amber Herold
h2. Starting Up
11 1 Amber Herold
12
13
14 10 Amber Herold
Set up is most stable with a good microscope alignment.
15 1 Amber Herold
16 11 Amber Herold
1.  Initial alignment at the microscope See Chapter on [[Microscope Set-up|Microscope setup]] for details.
17 1 Amber Herold
18 10 Amber Herold
  *  scope> Insert specimen holder
19
  *  scope> Put stage to eucenter and reset defocus in both HM mode AND LM mode
20
  *  scope> Check gun tilt and shift
21
  *  scope> Check rotation center and beam tilt pivot points in the HM mode
22
  *  scope> Save the goniomenter position for the square with intact supporting film where the alignment was done and do the same for a square where the supporting film is broken (where bright/dark images can be acquired).
23 1 Amber Herold
24
25 11 Amber Herold
2. Start python-based leginon main program amd leginon
26 1 Amber Herold
client. See Chapter on [[Microscope Set-up|Start Leginon]].
27
28 10 Amber Herold
  *  scope> Set up [[MSI set-up in more details#Starting-Up|CCD camera]]
29 11 Amber Herold
  *  [[Run Leginon Client on the Microscope Computer|scope> Start Leginon Client]]
30 10 Amber Herold
  *  scope> If the following error message appears, DO NOT touch it, or Leginon Client will need restarting
31 3 Alex Kadokura
32 10 Amber Herold
bq. _"Exception EAccessViolation in module adaExp.exe at...._
33
_Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx"_
34 1 Amber Herold
35 10 Amber Herold
  * [[Start Leginon at your main working station remotely|main> start-leginon.py]]
36
  *  Leginon Setup Wizard> Create new or Select old session, connect to the instrument host, and start.
37 1 Amber Herold
38
39 11 Amber Herold
3.  Start your preferred "MSI" application.
40 1 Amber Herold
41 10 Amber Herold
  a.  Leginon/Application/Run> Select the "MSI" application you would like to use (Edge,T,Raster, or Tomography), the scope, and the main launcher. Click Run.
42
  b.  Leginon/Node Selector> Select "Presets Manager" node.
43
  c. Leginon/Presets Manager/Preset Selector> Import presets from a previous working session !http://emg.nysbc.org/software/leginon/images/icons/import.png!
44
   * Select the TEM and camera to "Find" sessions closest to the condition of your data collection.
45
   * Use shift-key to select a series of presets if needed. Hold-down to ctrl-key allows individual selection/deselection.
46
   * Click "Import" button to import, the presets should appear in the Presets Manager/Preset Selector window.
47
   * Click "OK" to exit "Import Presets" window.
48 1 Amber Herold
49 10 Amber Herold
  Note: Creating new preset and adjusting the preset are covered in Calibration Chapter under [[Creating Presets for the first time|Creating presets the first time]].
50 1 Amber Herold
51
52 10 Amber Herold
  d.  The cycle order should match the sequence of data acquisition to minimize hysteresis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
53 1 Amber Herold
54 10 Amber Herold
  e. Leginon/Presets Manager> Select a preset in the list box and then click "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png! to apply the settings to the scope and synchronize Leginon and the scope settings.
55 1 Amber Herold
56
57 11 Amber Herold
4. Save the alignment and empty position in Leginon
58 1 Amber Herold
59 10 Amber Herold
   a.  scope> Recall the stored intact or empty position or go to such
60 1 Amber Herold
location.
61
62 10 Amber Herold
   b.  Leginon/Navigation/Toolbar/Stage Locations> !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! Enter a New name to store Stage Location such as, for example, "empty" and "align", respectively. Save the xy coordinate only. To recall the position, select the named location under the same tool and click on "send to scope".
63 1 Amber Herold
64
65
h2. Grid preset image shift refinement
66
67
68 11 Amber Herold
1.  Leginon/Presets Manager> Send a preset at an intermediate mag at HM mode as
69 1 Amber Herold
the reference (e.g. hl) "To Scope"<inlinegraphic
70
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
71
format="PNG"/>
72
73
74
(Note: sq preset in LM mode can be used if the image alignment between LM and HM
75
is known to be good).
76
77
78
*  scope> Move the grid to a position on the "align" square, either on a
79
recognizable object or burn a good hole through ice at the center of the CCD as
80
preparation for lower mag preset image shift correction.
81
82
83
*  Leginon/Presets Manager> Click on the tool "Align presets to each other"
84
<inlinegraphic
85
fileref="http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png"
86
format="PNG"/> .
87
88
89
*  Check, adjust, and save the preset(s) to be aligned.
90
91
92
The left panel displays and navigate by "Stage Position" of the reference preset
93
while the right panel displays and navigate by "Image Shift" of the preset to be
94
aligned
95
*  Leginon/Presets Manager/Align Presets> Choose the reference (hl) preset
96
on the left panel.
97
98
99
*  Leginon/Presets Manager/Align Presets> Choose the aligning (gr) preset
100
on the right panel.
101
102
103
*  Leginon/Presets Manager/Align Presets> Select "Similar look across mags"
104
mode at the top left corner of the window. This will use a smaller part of the
105
CCD to acquire the image for the lower mag preset so that it appears to be
106
magnified in the 512x512 image panel.
107
108
109
*  Leginon/Presets Manager/Align Presets> "Start" image acquisition.
110
111
112
*  Leginon/Presets Manager/Align Presets> If the recongnizable object can
113
not be found in both panels, go back to step one to choose a lower mag preset or
114
a better object.
115
116
117
*  Leginon/Presets Manager/Align Presets> If the recongnizable object can
118
be found in both panels but not at the center on the reference left panel,
119
navigate* it to the center using its "Click Tool". Wait for the move to complete
120
and both images refreshed.
121
122
123
*  Leginon/Presets Manager/Align Presets> Navigate* the object on the right
124
panel to the center using its "Click Tool".The new image shift is automatically
125
saved at each move.
126
127
128
*  Repeat 7 if necessary.
129
130
131
132
133
134
135
136
137
138
* "Navigate" means selecting the navigate tool<inlinegraphic
139
fileref="http://emg.nysbc.org/software/leginon/images/icons/arrow.png" format="PNG"/> on
140
the top right of image display and then double clicking the location of the new center that
141
is translated by the given TEM Parameter.
142
143
144
145
146
147
h2. Grid preset stage postion refinement
148
149
150 3 Alex Kadokura
151
Stage position model Scale and Rotation is stable as long as the LM eucentric height
152 1 Amber Herold
defocus is set to zero before each session, although variation from day to day is normal.
153
Check the stage movement calibration of gr preset by navigating it in Navigation node.
154
Within 10% error is satisfactory since reaching a square does not require high accuracy. If
155 3 Alex Kadokura
not, perform stage model position _[[Modeled Stage Position calibration#Magnification-and-image-rotation-adjustment-of-the-model|scale and rotation
156
calibration]]> at gr preset_ using "Calibration" application as
157 1 Amber Herold
following:
158
159
160
_[[Modeled Stage Position calibration|Why goniometer model is
161
better]]._
162
163
164
165
166
*  scope> Remove the objective apperture.
167
168
169
*  Leginon> Run "Calibrations" Application
170
171
172
*  Leginon/Presets Manager> Send "gr" preset to scope.
173
174
175
*  Leginon/GonioModeling/Settings/Measurement> Set axis=x,
176
Points=5,Tolerance=25%, and Interval=1e-05 m. Default model label is the session
177
name.
178
179
180
*  Leginon/GonioModeling/Settings/Modeling> Set "Scale and Rotation Adjustment
181
Only"=Yes. Other settings will be set automatically once measurements are
182
finished.
183
184
185
*  Leginon/GonioModeling> Click on <inlinegraphic
186
fileref="http://emg.nysbc.org/software/leginon/images/icons/cam_ruler.png"
187
format="PNG"/> will start "Measure" the shifts of between points that are moved on x
188
axis of the goniometer.
189
190
191
*  Leginon/GonioModeling> Click on <inlinegraphic
192
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"
193
/> will "Calibrate" the scale and rotation of the model on x axis of the
194
goniometer.
195
196
197
*  Leginon/GonioModeling> repeat 4-7 with the axis set to y.
198
199
200
201
202
203
204
205
206
207
h2. Acquire mosaic atlas of the grid
208
209
210
211
212
213
*  scope> Remove the objective apperture.
214
215
216
*  Leginon/Grid Targeting> Click "Settings" to check the preset used, the size
217
and name of the atlas.
218
219
220
* Leginon/Grid Targeting> Click "Calculate Atlas" <inlinegraphic
221
fileref="http://emg.nysbc.org/software/leginon/images/icons/calculate.png"
222
format="PNG"/>
223
224
225
*  Leginon/Grid Targeting> Click "Submit Targets" <inlinegraphic
226
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"
227
/>to begin acquiring the atlas.
228
229
230
*  Leginon/Square Targeting> Follow the atlas progress in the display. (***There
231
is no sign at the end currently other than that nothing happens for a long time and
232
that the atlas dimension is the same in other direction when it is completed.)
233
234
235
*  scope> Insert the objective aperture.
236
237
238
*  Leginon/Presets Manager/Send an HM preset "To Scope" several times to cycle all
239
magnifications and warm up the objective lens to minimize the image shift hysteresis
240
in switching between LM and HM modes. If the scope has been in LM mode for a long
241
time, wait for at least 30 min before final image shift refinement of needed LM preset
242
during acquisition (i.e. sq)
243
244
245
246
247
248
249
250
251
252
h2. Determine the proper exposure condition for exposure
253
presets
254
255
256
257
Exposure presets are the base presets for data collection. To keep it to low dose
258
condition such as 10 electron/Angstrum, measurement of dose is performed and the exposure
259
time is adjusted. If scope or camera parameters are changed during the session, the dose
260
value is no longer valid and should be remeasured.
261
262
263
264
265
*  Leginon/Navigation/Toolbar/Stage Locations<inlinegraphic
266
fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
267
format="PNG"/>> Send the "empty" position "To scope".
268
269
270
*  Leginon/Presets Manager/Preset Selector> Select a high mag exposure (i.e. "en"
271
or "ef"), confirm that the parameters are what you desire for the experiment, and then
272
send "To Scope"<inlinegraphic
273
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
274
format="PNG"/>.
275
276
277
*  scope> Expand the beam to at least cover the CCD (> medium circle mark on
278
the viewing screen) and to the size you desire.
279
280
281
*  Leginon/Presets Manager/Preset Selector> If you have changed the size or
282
location of the beam or other preset scope parameters at the scope, you must retrieve
283
the preset parameter "From scope"<inlinegraphic
284
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png"
285
format="PNG"/> first before checking and saving the dose value.
286
287
288
*  Leginon/Presets Manager/Preset Selector> "Acquire dose image"<inlinegraphic
289
fileref="http://emg.nysbc.org/software/leginon/images/icons/acquire.png"
290
format="PNG"/> for the selected preset.
291
292
293
*  Leginon/Presets Manager/Dose Image> Check the "Dose" value.
294
295
296
*  Leginon/Presets Manager/Dose Image> If desired dose is known, fill in the
297
value and then click "Match". This will change exposure time to give a close match of
298
the value entered.
299
300
301
*  Leginon/Presets Manager/Dose Image> If not satisfied with the value, click
302
"No", change the exposure time by editing the preset, and try again (No need to send
303
the preset "To scope" again for the repeat)
304
305
306
*  If the scope can not give such illumination to satisfy dose, exposure time and
307
area requirement, alignment is required or change of gun lens or spot size is
308
needed.
309
310
311
*  Leginon/Presets Manager> Copy the preset parameter to the other exposure
312
preset.
313
314
315
* Preset Selector> send the preset that has the dose measurement "To
316
Scope"<inlinegraphic
317
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
318
format="PNG"/>
319
320
321
* Preset Selector> select the preset to recieve the copy
322
323
324
* Preset Selector> retrieve the parameters "From Scope"<inlinegraphic
325
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png"
326
format="PNG"/> for the selected preset.
327
328
329
* Preset Selector/Edit><inlinegraphic
330
fileref="http://emg.nysbc.org/software/leginon/images/icons/edit.png"
331
format="PNG"/> Modify defocus for the preset after the retrieval.
332
333
334
335
336
337
338
339
340
341
342
343
h2. Update bright/dark correction images for exposure
344
presets
345 3 Alex Kadokura
346
347 1 Amber Herold
348
Exposure presets are the base presets for data collection. To make sure the best
349
correction is applied to them, the correction images should be acquired prior to the
350
experiment at session. Since no correlation is normally done on these images, only one [[Bright and Dark reference images#Correction-Channels|channel]] is needed. Normalization image might need to be
351
[[Bright and Dark reference images#Correction-Planmodified]] if a large chunk of region is blocked by a
352
fallen debris.
353
354
355
356
357
*  Leginon/Presets Manager> Select a dose measured high mag exposure (i.e. "en"
358
or "ef"), and then send "To Scope".
359
360
361
*  Leginon/Correction/Settings>
362
363
364
*  Select TEM and camera to match that will be used.
365
366
367
*  Camera Configuration: The exposure preset camera configuration
368
369
370
371
372
*  Leginon/Correction/Settings> Click OK to exit settings.
373
374
375
*  Leginon/Correction/Toolbar> Select "Raw image " from the pull down list of
376
acquisition modes and then click on "Acquire" button next to the selector to view an
377
image that is not corrected.
378
379
380
*  Leginon/Correction/Toolbar> Select "Dark reference" in the acquisition mode
381
selector and then click "Acquire" to acquire the Dark reference image for this
382
particular camera configuration.
383
384
385
*  Leginon/Correction/Toolbar> Select "Bright reference" and repeat the
386
acquisition to obtain the Bright reference.
387
388
389
*  Leginon/Correction/Toolbar> Select "Corrected image" and then "Acquire" to
390
view the corrected image. A corrected image should be free of artifacts and have
391
smaller standard deviation than the raw image, in general.
392
393
394
*  Leginon/Correction/Toolbar> If there are regions consistently over normalizes
395
the acquired image, either select the region and use the "+" tool to "Add Region to
396
Bad Pixel List", or click on<inlinegraphic
397
fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
398
format="PNG"/>to "Add Extreme Points to Bad Pixel List" one of more times to find
399
the outliers. Right click on a shown bad pixel removes it from the list.
400
401
402
403
404 9 Amber Herold
405 1 Amber Herold
406
407
408 3 Alex Kadokura
409 1 Amber Herold
h2. Medium mag preset image and beam shift refinement
410
411
412
413
Preset image shift alignment can be done follow the same procedure as the section "[[MSI set-up in more details#Grid-preset-image-shift-refinement|Grid preset image shift refinement]]" but is more
414
efficient using the automated mode as follows.
415
416
417
418
419
*  Leginon/Presets Manager> Send one of the exposure preset as the reference
420
(e.g. en) "To Scope"<inlinegraphic
421
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
422
format="PNG"/>
423
424
425
*  scope> Move the grid to a position on the "align" square, either on a
426
recognizable object or burn a good hole through ice at the center of the CCD as
427
preparation for lower mag preset image shift correction.
428
429
430
*  Leginon/Presets Manager> Click on the tool "Align presets to each other"
431
<inlinegraphic
432
fileref="http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png"
433
format="PNG"/> . The tool automatically select the reference preset that was
434
selected in the Presets Manager mainpanel and choose a preset at closest magnification
435
as the starting aligning preset.
436
437
438
*  Leginon/Presets Manager/Align Presets> Click "Start" to acquire the first
439
image pair.
440
441
442
*  Check, adjust, and save the preset(s) to be aligned.
443
444
445
The left panel displays and navigate by "Stage Position" of the reference preset
446
while the right panel displays and navigate by "Image Shift" of the preset to be
447
aligned
448
*  Leginon/Presets Manager/Align Presets> Changing Selection of "Similar
449
look across mags" or "Full CCD" mode at the top left corner of the window will
450
cause reacquisition of the image pair in the automated mode.
451
452
453
*  Leginon/Presets Manager/Align Presets> If the recongnizable object can
454
be found in both panels but not at the center on the reference left panel,
455
navigate* it to the center using its "Click Tool". Wait for the move to complete
456
and both images refreshed.
457
458
459
*  Leginon/Presets Manager/Align Presets> Navigate* the object on the right
460
panel to the center using its "Click Tool". The new image shift is automatically
461
saved after each move.
462
463
464
*  Repeat 3 if necessary.
465
466
467
*  Leginon/Presets Manager/Align Presets> Click "Continue" to save the
468
alignment to all presets of the same magnification and continue to the next
469
preset pair.
470
471
472
473
474
475
*  Repeat the process if so desired.
476
477
478
479
480
481
482
483
484
485 8 Anchi Cheng
h2. Go to MSI Operation section
486
487 1 Amber Herold
488
489
The setup that requires you to be next to the microscope is completed. Go to the section
490
on [[MSI Operation|MSI Operation]] You can come back and continue on this section after you
491
have learned how to go through the operation without automated hole finding.
492
493
494
495
496
497
h2. Hole Targeting Set-up
498
499
500 4 Alex Kadokura
501 1 Amber Herold
We can use any image that is already loaded into Hole Targeting to adjust these
502 4 Alex Kadokura
parameters.  The values in this node vary so much from grid to grid that the best way to
503 1 Amber Herold
find these parameters is by going through a step-by-step trial-and-error process. To see the
504 4 Alex Kadokura
final acquisition and focus targets, enable only the Original, acquisition, and focus
505 1 Amber Herold
display selections. Refer to Introduction Chapter on [[MSI set-up in more details
506
507
|how the various tools behave in the control panel]].
508
509
510
The loaded image normally comes from the image acquired through "Square" node when you
511
submit targets on the grid atlas in "Square Targeting" node.  You can also do a practice run
512
using the default image comes with Leginon. Leave the filename entry blank when you click
513
"Load" in the hole finding settings window for the step "Original". The default values in
514
the node gives an example of reasonable target finding with such default image.
515
516
517
518
519 4 Alex Kadokura
Setting up Hole Targeting Depends on which MSI flavor you are running:
520 1 Amber Herold
521
522
523 4 Alex Kadokura
524 1 Amber Herold
* [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the
525
edge image
526
527 4 Alex Kadokura
528 1 Amber Herold
* [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original
529
image
530
531 4 Alex Kadokura
532 1 Amber Herold
* [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given
533
origin
534
535
536
* [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic
537
target finding. Therefore, it does not requires set up.
538
539
540
541
542
543
544
545
546
547
h2.  Exposure Targeting Set-up
548
549
550
551
Like hole targeting, we can use any image that is already loaded into Exposure Targeting
552 4 Alex Kadokura
to adjust these parameters.
553
554 1 Amber Herold
Setting up Exposure Targeting Depends on which MSI flavor you are running.
555
556 4 Alex Kadokura
557 1 Amber Herold
* [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the
558
edge image
559
560 4 Alex Kadokura
561 1 Amber Herold
* [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original
562
image
563
564 4 Alex Kadokura
565
* [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given
566 1 Amber Herold
origin
567
568
569
* [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic
570
target finding. Therefore, it does not requires set up.
571
572
573
574
575
576
577
h2. Autofocus Test
578
579
580
581
There are two focusers in MSI. "Z Focus" node usually moves stage in Z direction to
582
Eucentric height at lower mags. "Focus" node usually just correct the defocus in the way
583
equivalent to turning the focus adjustment knob on the scope at a higher mag. The defocus
584
measurments in both nodes are based on the same principle of beam-tilt induced image shift.
585
A correlation (normally phase) is used to obtain the image shift. Therefore, it is important
586
to test and preform autofocus measurement at area with recognizable feature that does not
587
move significantly during the defocus measurement.
588
589
590
The test use "Manual_after" focus step in the focus sequence for checking. Two
591
procedures can be used to reach a area for good correlation measurements:
592
593
594
595
596
*  Using Simulate Target tool
597
598
599
600
601
*  Leginon/Presets Manager> Send the autofocus preset "To Scope"
602
603
604
*  scope> Move to an area that has recognizable feature on the carbon
605
support at the center.
606
607
608
*  Leginon/(Z) Focus/Focus Sequence<inlinegraphic
609
fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
610
format="PNG"/>> Enable desired autofocus and "Manual_after" focus
611
steps.
612
613
614
*  Leginon/(Z) Focus/Toolbar> Click Simulate Target<inlinegraphic
615
fileref="http://emg.nysbc.org/software/leginon/images/icons/simulatetarget.png"
616
format="PNG"/> and follow the autofocus
617
618
619
620
621
622
623
*  Submiting focus target from Hole Targeting
624
625
626
*To use this procedure, there has to be a grid mosaic, and the Hole Targeting
627
should allow user verification of targets
628
629
630
*  Leginon/(Z) Focus/Focus Sequence<inlinegraphic
631
fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
632
format="PNG"/>> Enable desired autofocus and "Manual_after" focus
633
steps.
634
635
636
*  Leginon/Square Targeting> Select a test square and submit it<inlinegraphic
637
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png"
638
format="PNG"/>
639
640
641
*  Leginon/Hole Targeting> Submit a focus target to test autofocus in Z Focus
642
node and an acquisition target for the next stage. Both targets should be in an
643
area that has recognizable feature with carbon support.
644
645
646
*  Leginon/Z focus> follow the autofocus test in this node.
647
648
649
*  Leginon/Exposure Targeting> Submit a focus target at a location that has
650
recognizable feature with carbon support to test autofocus in Focus node.
651
652
653
*  Leginon/Focus> follow autofocus test in this node.
654
655
656
657
658
659
660
661
662
Checking the performance of autofocus by the following
663
criteria:
664
665
666
667
668
*  During autofocusing
669
670
671
672
673
*  Display the image and correlection peak: The observed movement of images
674
upon beam tilt should be consistent with the correlation peak and the edge of
675
the beam should not be visible in any of the beam-tilt images.
676
677
678
*  Display the correlation image: The correlation peak at its correct position
679
should be significantly higher than background.
680
681
682
683
684
685
686
*  During manual focusing
687
688
Manual Focusing can be activated either by enabling "Manual_before/Manual_after"
689
steps or directly when leginon is idle using (Z) Focus/Toolbar/MF. To activate manual
690
focus directly, it is necessary to send to scope the preset of the choice (normally
691
fc) first, though.
692
693
694
*  MF> The FFT should show visible and isotropic Thon ring according to the
695
fc preset defocus.
696
697
698
*  MF/Toolbar> Make sure the move type is "Defocus" and the set value in the
699
box is 0 m then click "Set instrument" botton<inlinegraphic
700
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
701
format="PNG"/>. This will set the defocus to the current 0.
702
703
704
*  MF> The perfect autofocus result is that the FFT indicates that the scope
705
is at Gaussian focus.
706
707
708
*  For relative movement, press "+" or "-". The step size is determined in
709
MF/Toolbar/Settings>
710
711
712
*  To reset defocus to zero at the new focus, press the so-named
713
button<inlinegraphic
714
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentsetnew.png"
715
format="PNG"/>.
716 2 Amber Herold
717
718
*  Clicking the Stop button<inlinegraphic
719
fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png"
720
format="PNG"/> exits the maunal check.
721
722
723
724 1 Amber Herold
______
725
726
[[MSI Quick-start|< MSI Quick-start]] | [[Special Operation setup|Special Operation setup >]]
727
728
______