MSI set up in more details » History » Version 13

Amber Herold, 06/22/2010 10:44 AM

1 1 Amber Herold
h1. MSI set-up in more details
2
3
4 10 Amber Herold
This section shows some of the more important and complicated steps in the set-up checklist. It assumes that the user knows the nomenclatures. See the full documentation if you want the detailed explanation. It also assumes that general set-up and calibrations exist from previous sessions so that most calibrations can remain unchanged. The steps that are NRAMM specific are marked. MSI now features stage Eucentric height correction at each square.
5 1 Amber Herold
6 10 Amber Herold
NOTE: The Tietz imaging software (EMMenu) must be turned OFF before Leginon is started. In contrast to this, the Gatan imaging software (Digital Micrograph) must be ON before Leginon is started.
7 1 Amber Herold
8
9
10 10 Amber Herold
h2. Starting Up
11 1 Amber Herold
12
13
14 10 Amber Herold
Set up is most stable with a good microscope alignment.
15 1 Amber Herold
16 11 Amber Herold
1.  Initial alignment at the microscope See Chapter on [[Microscope Set-up|Microscope setup]] for details.
17 1 Amber Herold
18 10 Amber Herold
  *  scope> Insert specimen holder
19
  *  scope> Put stage to eucenter and reset defocus in both HM mode AND LM mode
20
  *  scope> Check gun tilt and shift
21
  *  scope> Check rotation center and beam tilt pivot points in the HM mode
22
  *  scope> Save the goniomenter position for the square with intact supporting film where the alignment was done and do the same for a square where the supporting film is broken (where bright/dark images can be acquired).
23 1 Amber Herold
24
25 11 Amber Herold
2. Start python-based leginon main program amd leginon
26 1 Amber Herold
client. See Chapter on [[Microscope Set-up|Start Leginon]].
27
28 10 Amber Herold
  *  scope> Set up [[MSI set-up in more details#Starting-Up|CCD camera]]
29 11 Amber Herold
  *  [[Run Leginon Client on the Microscope Computer|scope> Start Leginon Client]]
30 10 Amber Herold
  *  scope> If the following error message appears, DO NOT touch it, or Leginon Client will need restarting
31 3 Alex Kadokura
32 10 Amber Herold
bq. _"Exception EAccessViolation in module adaExp.exe at...._
33
_Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx"_
34 1 Amber Herold
35 10 Amber Herold
  * [[Start Leginon at your main working station remotely|main> start-leginon.py]]
36
  *  Leginon Setup Wizard> Create new or Select old session, connect to the instrument host, and start.
37 1 Amber Herold
38
39 11 Amber Herold
3.  Start your preferred "MSI" application.
40 1 Amber Herold
41 13 Amber Herold
 *  Leginon/Application/Run> Select the "MSI" application you would like to use (Edge,T,Raster, or Tomography), the scope, and the main launcher. Click Run.
42
 *  Leginon/Node Selector> Select "Presets Manager" node.
43
 * Leginon/Presets Manager/Preset Selector> Import presets from a previous working session !http://emg.nysbc.org/software/leginon/images/icons/import.png!
44
 ** Select the TEM and camera to "Find" sessions closest to the condition of your data collection.
45
 ** Use shift-key to select a series of presets if needed. Hold-down to ctrl-key allows individual selection/deselection.
46
 ** Click "Import" button to import, the presets should appear in the Presets Manager/Preset Selector window.
47
 ** Click "OK" to exit "Import Presets" window.
48
 Note: Creating new preset and adjusting the preset are covered in Calibration Chapter under [[Creating Presets for the first time|Creating presets the first time]].
49
 *  The cycle order should match the sequence of data acquisition to minimize hysteresis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
50
 * Leginon/Presets Manager> Select a preset in the list box and then click "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png! to apply the settings to the scope and synchronize Leginon and the scope settings.
51 1 Amber Herold
52
53 11 Amber Herold
4. Save the alignment and empty position in Leginon
54 1 Amber Herold
55 12 Amber Herold
  a.  scope> Recall the stored intact or empty position or go to such
56 1 Amber Herold
location.
57
58 12 Amber Herold
  b.  Leginon/Navigation/Toolbar/Stage Locations> !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! Enter a New name to store Stage Location such as, for example, "empty" and "align", respectively. Save the xy coordinate only. To recall the position, select the named location under the same tool and click on "send to scope".
59 1 Amber Herold
60
61
h2. Grid preset image shift refinement
62
63
64 11 Amber Herold
1.  Leginon/Presets Manager> Send a preset at an intermediate mag at HM mode as
65 1 Amber Herold
the reference (e.g. hl) "To Scope"<inlinegraphic
66
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
67
format="PNG"/>
68
69
70
(Note: sq preset in LM mode can be used if the image alignment between LM and HM
71
is known to be good).
72
73
74
*  scope> Move the grid to a position on the "align" square, either on a
75
recognizable object or burn a good hole through ice at the center of the CCD as
76
preparation for lower mag preset image shift correction.
77
78
79
*  Leginon/Presets Manager> Click on the tool "Align presets to each other"
80
<inlinegraphic
81
fileref="http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png"
82
format="PNG"/> .
83
84
85
*  Check, adjust, and save the preset(s) to be aligned.
86
87
88
The left panel displays and navigate by "Stage Position" of the reference preset
89
while the right panel displays and navigate by "Image Shift" of the preset to be
90
aligned
91
*  Leginon/Presets Manager/Align Presets> Choose the reference (hl) preset
92
on the left panel.
93
94
95
*  Leginon/Presets Manager/Align Presets> Choose the aligning (gr) preset
96
on the right panel.
97
98
99
*  Leginon/Presets Manager/Align Presets> Select "Similar look across mags"
100
mode at the top left corner of the window. This will use a smaller part of the
101
CCD to acquire the image for the lower mag preset so that it appears to be
102
magnified in the 512x512 image panel.
103
104
105
*  Leginon/Presets Manager/Align Presets> "Start" image acquisition.
106
107
108
*  Leginon/Presets Manager/Align Presets> If the recongnizable object can
109
not be found in both panels, go back to step one to choose a lower mag preset or
110
a better object.
111
112
113
*  Leginon/Presets Manager/Align Presets> If the recongnizable object can
114
be found in both panels but not at the center on the reference left panel,
115
navigate* it to the center using its "Click Tool". Wait for the move to complete
116
and both images refreshed.
117
118
119
*  Leginon/Presets Manager/Align Presets> Navigate* the object on the right
120
panel to the center using its "Click Tool".The new image shift is automatically
121
saved at each move.
122
123
124
*  Repeat 7 if necessary.
125
126
127
128
129
130
131
132
133
134
* "Navigate" means selecting the navigate tool<inlinegraphic
135
fileref="http://emg.nysbc.org/software/leginon/images/icons/arrow.png" format="PNG"/> on
136
the top right of image display and then double clicking the location of the new center that
137
is translated by the given TEM Parameter.
138
139
140
141
142
143
h2. Grid preset stage postion refinement
144
145
146 3 Alex Kadokura
147
Stage position model Scale and Rotation is stable as long as the LM eucentric height
148 1 Amber Herold
defocus is set to zero before each session, although variation from day to day is normal.
149
Check the stage movement calibration of gr preset by navigating it in Navigation node.
150
Within 10% error is satisfactory since reaching a square does not require high accuracy. If
151 3 Alex Kadokura
not, perform stage model position _[[Modeled Stage Position calibration#Magnification-and-image-rotation-adjustment-of-the-model|scale and rotation
152
calibration]]> at gr preset_ using "Calibration" application as
153 1 Amber Herold
following:
154
155
156
_[[Modeled Stage Position calibration|Why goniometer model is
157
better]]._
158
159
160
161
162
*  scope> Remove the objective apperture.
163
164
165
*  Leginon> Run "Calibrations" Application
166
167
168
*  Leginon/Presets Manager> Send "gr" preset to scope.
169
170
171
*  Leginon/GonioModeling/Settings/Measurement> Set axis=x,
172
Points=5,Tolerance=25%, and Interval=1e-05 m. Default model label is the session
173
name.
174
175
176
*  Leginon/GonioModeling/Settings/Modeling> Set "Scale and Rotation Adjustment
177
Only"=Yes. Other settings will be set automatically once measurements are
178
finished.
179
180
181
*  Leginon/GonioModeling> Click on <inlinegraphic
182
fileref="http://emg.nysbc.org/software/leginon/images/icons/cam_ruler.png"
183
format="PNG"/> will start "Measure" the shifts of between points that are moved on x
184
axis of the goniometer.
185
186
187
*  Leginon/GonioModeling> Click on <inlinegraphic
188
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"
189
/> will "Calibrate" the scale and rotation of the model on x axis of the
190
goniometer.
191
192
193
*  Leginon/GonioModeling> repeat 4-7 with the axis set to y.
194
195
196
197
198
199
200
201
202
203
h2. Acquire mosaic atlas of the grid
204
205
206
207
208
209
*  scope> Remove the objective apperture.
210
211
212
*  Leginon/Grid Targeting> Click "Settings" to check the preset used, the size
213
and name of the atlas.
214
215
216
* Leginon/Grid Targeting> Click "Calculate Atlas" <inlinegraphic
217
fileref="http://emg.nysbc.org/software/leginon/images/icons/calculate.png"
218
format="PNG"/>
219
220
221
*  Leginon/Grid Targeting> Click "Submit Targets" <inlinegraphic
222
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"
223
/>to begin acquiring the atlas.
224
225
226
*  Leginon/Square Targeting> Follow the atlas progress in the display. (***There
227
is no sign at the end currently other than that nothing happens for a long time and
228
that the atlas dimension is the same in other direction when it is completed.)
229
230
231
*  scope> Insert the objective aperture.
232
233
234
*  Leginon/Presets Manager/Send an HM preset "To Scope" several times to cycle all
235
magnifications and warm up the objective lens to minimize the image shift hysteresis
236
in switching between LM and HM modes. If the scope has been in LM mode for a long
237
time, wait for at least 30 min before final image shift refinement of needed LM preset
238
during acquisition (i.e. sq)
239
240
241
242
243
244
245
246
247
248
h2. Determine the proper exposure condition for exposure
249
presets
250
251
252
253
Exposure presets are the base presets for data collection. To keep it to low dose
254
condition such as 10 electron/Angstrum, measurement of dose is performed and the exposure
255
time is adjusted. If scope or camera parameters are changed during the session, the dose
256
value is no longer valid and should be remeasured.
257
258
259
260
261
*  Leginon/Navigation/Toolbar/Stage Locations<inlinegraphic
262
fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
263
format="PNG"/>> Send the "empty" position "To scope".
264
265
266
*  Leginon/Presets Manager/Preset Selector> Select a high mag exposure (i.e. "en"
267
or "ef"), confirm that the parameters are what you desire for the experiment, and then
268
send "To Scope"<inlinegraphic
269
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
270
format="PNG"/>.
271
272
273
*  scope> Expand the beam to at least cover the CCD (> medium circle mark on
274
the viewing screen) and to the size you desire.
275
276
277
*  Leginon/Presets Manager/Preset Selector> If you have changed the size or
278
location of the beam or other preset scope parameters at the scope, you must retrieve
279
the preset parameter "From scope"<inlinegraphic
280
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png"
281
format="PNG"/> first before checking and saving the dose value.
282
283
284
*  Leginon/Presets Manager/Preset Selector> "Acquire dose image"<inlinegraphic
285
fileref="http://emg.nysbc.org/software/leginon/images/icons/acquire.png"
286
format="PNG"/> for the selected preset.
287
288
289
*  Leginon/Presets Manager/Dose Image> Check the "Dose" value.
290
291
292
*  Leginon/Presets Manager/Dose Image> If desired dose is known, fill in the
293
value and then click "Match". This will change exposure time to give a close match of
294
the value entered.
295
296
297
*  Leginon/Presets Manager/Dose Image> If not satisfied with the value, click
298
"No", change the exposure time by editing the preset, and try again (No need to send
299
the preset "To scope" again for the repeat)
300
301
302
*  If the scope can not give such illumination to satisfy dose, exposure time and
303
area requirement, alignment is required or change of gun lens or spot size is
304
needed.
305
306
307
*  Leginon/Presets Manager> Copy the preset parameter to the other exposure
308
preset.
309
310
311
* Preset Selector> send the preset that has the dose measurement "To
312
Scope"<inlinegraphic
313
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
314
format="PNG"/>
315
316
317
* Preset Selector> select the preset to recieve the copy
318
319
320
* Preset Selector> retrieve the parameters "From Scope"<inlinegraphic
321
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png"
322
format="PNG"/> for the selected preset.
323
324
325
* Preset Selector/Edit><inlinegraphic
326
fileref="http://emg.nysbc.org/software/leginon/images/icons/edit.png"
327
format="PNG"/> Modify defocus for the preset after the retrieval.
328
329
330
331
332
333
334
335
336
337
338
339
h2. Update bright/dark correction images for exposure
340
presets
341 3 Alex Kadokura
342
343 1 Amber Herold
344
Exposure presets are the base presets for data collection. To make sure the best
345
correction is applied to them, the correction images should be acquired prior to the
346
experiment at session. Since no correlation is normally done on these images, only one [[Bright and Dark reference images#Correction-Channels|channel]] is needed. Normalization image might need to be
347
[[Bright and Dark reference images#Correction-Planmodified]] if a large chunk of region is blocked by a
348
fallen debris.
349
350
351
352
353
*  Leginon/Presets Manager> Select a dose measured high mag exposure (i.e. "en"
354
or "ef"), and then send "To Scope".
355
356
357
*  Leginon/Correction/Settings>
358
359
360
*  Select TEM and camera to match that will be used.
361
362
363
*  Camera Configuration: The exposure preset camera configuration
364
365
366
367
368
*  Leginon/Correction/Settings> Click OK to exit settings.
369
370
371
*  Leginon/Correction/Toolbar> Select "Raw image " from the pull down list of
372
acquisition modes and then click on "Acquire" button next to the selector to view an
373
image that is not corrected.
374
375
376
*  Leginon/Correction/Toolbar> Select "Dark reference" in the acquisition mode
377
selector and then click "Acquire" to acquire the Dark reference image for this
378
particular camera configuration.
379
380
381
*  Leginon/Correction/Toolbar> Select "Bright reference" and repeat the
382
acquisition to obtain the Bright reference.
383
384
385
*  Leginon/Correction/Toolbar> Select "Corrected image" and then "Acquire" to
386
view the corrected image. A corrected image should be free of artifacts and have
387
smaller standard deviation than the raw image, in general.
388
389
390
*  Leginon/Correction/Toolbar> If there are regions consistently over normalizes
391
the acquired image, either select the region and use the "+" tool to "Add Region to
392
Bad Pixel List", or click on<inlinegraphic
393
fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
394
format="PNG"/>to "Add Extreme Points to Bad Pixel List" one of more times to find
395
the outliers. Right click on a shown bad pixel removes it from the list.
396
397
398
399
400 9 Amber Herold
401 1 Amber Herold
402
403
404 3 Alex Kadokura
405 1 Amber Herold
h2. Medium mag preset image and beam shift refinement
406
407
408
409
Preset image shift alignment can be done follow the same procedure as the section "[[MSI set-up in more details#Grid-preset-image-shift-refinement|Grid preset image shift refinement]]" but is more
410
efficient using the automated mode as follows.
411
412
413
414
415
*  Leginon/Presets Manager> Send one of the exposure preset as the reference
416
(e.g. en) "To Scope"<inlinegraphic
417
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
418
format="PNG"/>
419
420
421
*  scope> Move the grid to a position on the "align" square, either on a
422
recognizable object or burn a good hole through ice at the center of the CCD as
423
preparation for lower mag preset image shift correction.
424
425
426
*  Leginon/Presets Manager> Click on the tool "Align presets to each other"
427
<inlinegraphic
428
fileref="http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png"
429
format="PNG"/> . The tool automatically select the reference preset that was
430
selected in the Presets Manager mainpanel and choose a preset at closest magnification
431
as the starting aligning preset.
432
433
434
*  Leginon/Presets Manager/Align Presets> Click "Start" to acquire the first
435
image pair.
436
437
438
*  Check, adjust, and save the preset(s) to be aligned.
439
440
441
The left panel displays and navigate by "Stage Position" of the reference preset
442
while the right panel displays and navigate by "Image Shift" of the preset to be
443
aligned
444
*  Leginon/Presets Manager/Align Presets> Changing Selection of "Similar
445
look across mags" or "Full CCD" mode at the top left corner of the window will
446
cause reacquisition of the image pair in the automated mode.
447
448
449
*  Leginon/Presets Manager/Align Presets> If the recongnizable object can
450
be found in both panels but not at the center on the reference left panel,
451
navigate* it to the center using its "Click Tool". Wait for the move to complete
452
and both images refreshed.
453
454
455
*  Leginon/Presets Manager/Align Presets> Navigate* the object on the right
456
panel to the center using its "Click Tool". The new image shift is automatically
457
saved after each move.
458
459
460
*  Repeat 3 if necessary.
461
462
463
*  Leginon/Presets Manager/Align Presets> Click "Continue" to save the
464
alignment to all presets of the same magnification and continue to the next
465
preset pair.
466
467
468
469
470
471
*  Repeat the process if so desired.
472
473
474
475
476
477
478
479
480
481 8 Anchi Cheng
h2. Go to MSI Operation section
482
483 1 Amber Herold
484
485
The setup that requires you to be next to the microscope is completed. Go to the section
486
on [[MSI Operation|MSI Operation]] You can come back and continue on this section after you
487
have learned how to go through the operation without automated hole finding.
488
489
490
491
492
493
h2. Hole Targeting Set-up
494
495
496 4 Alex Kadokura
497 1 Amber Herold
We can use any image that is already loaded into Hole Targeting to adjust these
498 4 Alex Kadokura
parameters.  The values in this node vary so much from grid to grid that the best way to
499 1 Amber Herold
find these parameters is by going through a step-by-step trial-and-error process. To see the
500 4 Alex Kadokura
final acquisition and focus targets, enable only the Original, acquisition, and focus
501 1 Amber Herold
display selections. Refer to Introduction Chapter on [[MSI set-up in more details
502
503
|how the various tools behave in the control panel]].
504
505
506
The loaded image normally comes from the image acquired through "Square" node when you
507
submit targets on the grid atlas in "Square Targeting" node.  You can also do a practice run
508
using the default image comes with Leginon. Leave the filename entry blank when you click
509
"Load" in the hole finding settings window for the step "Original". The default values in
510
the node gives an example of reasonable target finding with such default image.
511
512
513
514
515 4 Alex Kadokura
Setting up Hole Targeting Depends on which MSI flavor you are running:
516 1 Amber Herold
517
518
519 4 Alex Kadokura
520 1 Amber Herold
* [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the
521
edge image
522
523 4 Alex Kadokura
524 1 Amber Herold
* [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original
525
image
526
527 4 Alex Kadokura
528 1 Amber Herold
* [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given
529
origin
530
531
532
* [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic
533
target finding. Therefore, it does not requires set up.
534
535
536
537
538
539
540
541
542
543
h2.  Exposure Targeting Set-up
544
545
546
547
Like hole targeting, we can use any image that is already loaded into Exposure Targeting
548 4 Alex Kadokura
to adjust these parameters.
549
550 1 Amber Herold
Setting up Exposure Targeting Depends on which MSI flavor you are running.
551
552 4 Alex Kadokura
553 1 Amber Herold
* [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the
554
edge image
555
556 4 Alex Kadokura
557 1 Amber Herold
* [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original
558
image
559
560 4 Alex Kadokura
561
* [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given
562 1 Amber Herold
origin
563
564
565
* [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic
566
target finding. Therefore, it does not requires set up.
567
568
569
570
571
572
573
h2. Autofocus Test
574
575
576
577
There are two focusers in MSI. "Z Focus" node usually moves stage in Z direction to
578
Eucentric height at lower mags. "Focus" node usually just correct the defocus in the way
579
equivalent to turning the focus adjustment knob on the scope at a higher mag. The defocus
580
measurments in both nodes are based on the same principle of beam-tilt induced image shift.
581
A correlation (normally phase) is used to obtain the image shift. Therefore, it is important
582
to test and preform autofocus measurement at area with recognizable feature that does not
583
move significantly during the defocus measurement.
584
585
586
The test use "Manual_after" focus step in the focus sequence for checking. Two
587
procedures can be used to reach a area for good correlation measurements:
588
589
590
591
592
*  Using Simulate Target tool
593
594
595
596
597
*  Leginon/Presets Manager> Send the autofocus preset "To Scope"
598
599
600
*  scope> Move to an area that has recognizable feature on the carbon
601
support at the center.
602
603
604
*  Leginon/(Z) Focus/Focus Sequence<inlinegraphic
605
fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
606
format="PNG"/>> Enable desired autofocus and "Manual_after" focus
607
steps.
608
609
610
*  Leginon/(Z) Focus/Toolbar> Click Simulate Target<inlinegraphic
611
fileref="http://emg.nysbc.org/software/leginon/images/icons/simulatetarget.png"
612
format="PNG"/> and follow the autofocus
613
614
615
616
617
618
619
*  Submiting focus target from Hole Targeting
620
621
622
*To use this procedure, there has to be a grid mosaic, and the Hole Targeting
623
should allow user verification of targets
624
625
626
*  Leginon/(Z) Focus/Focus Sequence<inlinegraphic
627
fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
628
format="PNG"/>> Enable desired autofocus and "Manual_after" focus
629
steps.
630
631
632
*  Leginon/Square Targeting> Select a test square and submit it<inlinegraphic
633
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png"
634
format="PNG"/>
635
636
637
*  Leginon/Hole Targeting> Submit a focus target to test autofocus in Z Focus
638
node and an acquisition target for the next stage. Both targets should be in an
639
area that has recognizable feature with carbon support.
640
641
642
*  Leginon/Z focus> follow the autofocus test in this node.
643
644
645
*  Leginon/Exposure Targeting> Submit a focus target at a location that has
646
recognizable feature with carbon support to test autofocus in Focus node.
647
648
649
*  Leginon/Focus> follow autofocus test in this node.
650
651
652
653
654
655
656
657
658
Checking the performance of autofocus by the following
659
criteria:
660
661
662
663
664
*  During autofocusing
665
666
667
668
669
*  Display the image and correlection peak: The observed movement of images
670
upon beam tilt should be consistent with the correlation peak and the edge of
671
the beam should not be visible in any of the beam-tilt images.
672
673
674
*  Display the correlation image: The correlation peak at its correct position
675
should be significantly higher than background.
676
677
678
679
680
681
682
*  During manual focusing
683
684
Manual Focusing can be activated either by enabling "Manual_before/Manual_after"
685
steps or directly when leginon is idle using (Z) Focus/Toolbar/MF. To activate manual
686
focus directly, it is necessary to send to scope the preset of the choice (normally
687
fc) first, though.
688
689
690
*  MF> The FFT should show visible and isotropic Thon ring according to the
691
fc preset defocus.
692
693
694
*  MF/Toolbar> Make sure the move type is "Defocus" and the set value in the
695
box is 0 m then click "Set instrument" botton<inlinegraphic
696
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
697
format="PNG"/>. This will set the defocus to the current 0.
698
699
700
*  MF> The perfect autofocus result is that the FFT indicates that the scope
701
is at Gaussian focus.
702
703
704
*  For relative movement, press "+" or "-". The step size is determined in
705
MF/Toolbar/Settings>
706
707
708
*  To reset defocus to zero at the new focus, press the so-named
709
button<inlinegraphic
710
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentsetnew.png"
711
format="PNG"/>.
712 2 Amber Herold
713
714
*  Clicking the Stop button<inlinegraphic
715
fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png"
716
format="PNG"/> exits the maunal check.
717
718
719
720 1 Amber Herold
______
721
722
[[MSI Quick-start|< MSI Quick-start]] | [[Special Operation setup|Special Operation setup >]]