MSI set up in more details » History » Version 16

Amber Herold, 06/22/2010 11:34 AM

1 1 Amber Herold
h1. MSI set-up in more details
2
3
4 10 Amber Herold
This section shows some of the more important and complicated steps in the set-up checklist. It assumes that the user knows the nomenclatures. See the full documentation if you want the detailed explanation. It also assumes that general set-up and calibrations exist from previous sessions so that most calibrations can remain unchanged. The steps that are NRAMM specific are marked. MSI now features stage Eucentric height correction at each square.
5 1 Amber Herold
6 10 Amber Herold
NOTE: The Tietz imaging software (EMMenu) must be turned OFF before Leginon is started. In contrast to this, the Gatan imaging software (Digital Micrograph) must be ON before Leginon is started.
7 1 Amber Herold
8
9
10 10 Amber Herold
h2. Starting Up
11 1 Amber Herold
12
13
14 10 Amber Herold
Set up is most stable with a good microscope alignment.
15 1 Amber Herold
16 11 Amber Herold
1.  Initial alignment at the microscope See Chapter on [[Microscope Set-up|Microscope setup]] for details.
17 1 Amber Herold
18 10 Amber Herold
  *  scope> Insert specimen holder
19
  *  scope> Put stage to eucenter and reset defocus in both HM mode AND LM mode
20
  *  scope> Check gun tilt and shift
21
  *  scope> Check rotation center and beam tilt pivot points in the HM mode
22
  *  scope> Save the goniomenter position for the square with intact supporting film where the alignment was done and do the same for a square where the supporting film is broken (where bright/dark images can be acquired).
23 1 Amber Herold
24
25 11 Amber Herold
2. Start python-based leginon main program amd leginon
26 1 Amber Herold
client. See Chapter on [[Microscope Set-up|Start Leginon]].
27
28 10 Amber Herold
  *  scope> Set up [[MSI set-up in more details#Starting-Up|CCD camera]]
29 11 Amber Herold
  *  [[Run Leginon Client on the Microscope Computer|scope> Start Leginon Client]]
30 10 Amber Herold
  *  scope> If the following error message appears, DO NOT touch it, or Leginon Client will need restarting
31 3 Alex Kadokura
32 10 Amber Herold
bq. _"Exception EAccessViolation in module adaExp.exe at...._
33
_Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx"_
34 1 Amber Herold
35 10 Amber Herold
  * [[Start Leginon at your main working station remotely|main> start-leginon.py]]
36
  *  Leginon Setup Wizard> Create new or Select old session, connect to the instrument host, and start.
37 1 Amber Herold
38
39 11 Amber Herold
3.  Start your preferred "MSI" application.
40 1 Amber Herold
41 13 Amber Herold
 *  Leginon/Application/Run> Select the "MSI" application you would like to use (Edge,T,Raster, or Tomography), the scope, and the main launcher. Click Run.
42
 *  Leginon/Node Selector> Select "Presets Manager" node.
43
 * Leginon/Presets Manager/Preset Selector> Import presets from a previous working session !http://emg.nysbc.org/software/leginon/images/icons/import.png!
44
 ** Select the TEM and camera to "Find" sessions closest to the condition of your data collection.
45
 ** Use shift-key to select a series of presets if needed. Hold-down to ctrl-key allows individual selection/deselection.
46
 ** Click "Import" button to import, the presets should appear in the Presets Manager/Preset Selector window.
47
 ** Click "OK" to exit "Import Presets" window.
48
 Note: Creating new preset and adjusting the preset are covered in Calibration Chapter under [[Creating Presets for the first time|Creating presets the first time]].
49
 *  The cycle order should match the sequence of data acquisition to minimize hysteresis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
50
 * Leginon/Presets Manager> Select a preset in the list box and then click "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png! to apply the settings to the scope and synchronize Leginon and the scope settings.
51 1 Amber Herold
52
53 11 Amber Herold
4. Save the alignment and empty position in Leginon
54 1 Amber Herold
55 14 Amber Herold
*  scope> Recall the stored intact or empty position or go to such
56 1 Amber Herold
location.
57
58 14 Amber Herold
*  Leginon/Navigation/Toolbar/Stage Locations> !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! Enter a New name to store Stage Location such as, for example, "empty" and "align", respectively. Save the xy coordinate only. To recall the position, select the named location under the same tool and click on "send to scope".
59 1 Amber Herold
60
61
h2. Grid preset image shift refinement
62
63 11 Amber Herold
64 16 Amber Herold
1.  Leginon/Presets Manager> Send a preset at an intermediate mag at HM mode as the reference (e.g. hl) "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!
65 1 Amber Herold
66 15 Amber Herold
Note: sq preset in LM mode can be used if the image alignment between LM and HM
67 14 Amber Herold
is known to be good.
68 1 Amber Herold
69
70 14 Amber Herold
2.  scope> Move the grid to a position on the "align" square, either on a recognizable object or burn a good hole through ice at the center of the CCD as preparation for lower mag preset image shift correction.
71 1 Amber Herold
72 14 Amber Herold
73 16 Amber Herold
3.  Leginon/Presets Manager> Click on the tool "Align presets to each other" !http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png!.
74 1 Amber Herold
75
76 14 Amber Herold
4.  Check, adjust, and save the preset(s) to be aligned.
77 1 Amber Herold
78
79 14 Amber Herold
 The left panel displays and navigate by "Stage Position" of the reference preset while the right panel displays and navigate by "Image Shift" of the preset to be aligned
80
 *  Leginon/Presets Manager/Align Presets> Choose the reference (hl) preset on the left panel.
81 1 Amber Herold
82 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Choose the aligning (gr) preset on the right panel.
83 1 Amber Herold
84 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Select "Similar look across mags" mode at the top left corner of the window. This will use a smaller part of the CCD to acquire the image for the lower mag preset so that it appears to be magnified in the 512x512 image panel.
85 1 Amber Herold
86 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> "Start" image acquisition.
87 1 Amber Herold
88 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> If the recognizable object can not be found in both panels, go back to step one to choose a lower mag preset or a better object.
89 1 Amber Herold
90 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> If the recognizable object can be found in both panels but not at the center on the reference left panel, navigate* it to the center using its "Click Tool". Wait for the move to complete and both images refreshed.
91 1 Amber Herold
92 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Navigate* the object on the right panel to the center using its "Click Tool".The new image shift is automatically saved at each move.
93 1 Amber Herold
94 14 Amber Herold
 *  Repeat 7 if necessary.
95 1 Amber Herold
96
97 14 Amber Herold
Note: "Navigate" means selecting the navigate tool !http://emg.nysbc.org/software/leginon/images/icons/arrow.png! on the top right of image display and then double clicking the location of the new center that is translated by the given TEM Parameter.
98 1 Amber Herold
99
100
101 14 Amber Herold
h2. Grid preset stage position refinement
102 1 Amber Herold
103 3 Alex Kadokura
104 1 Amber Herold
105 14 Amber Herold
Stage position model Scale and Rotation is stable as long as the LM eucentric height defocus is set to zero before each session, although variation from day to day is normal. Check the stage movement calibration of gr preset by navigating it in Navigation node. Within 10% error is satisfactory since reaching a square does not require high accuracy. If not, perform stage model position _[[Modeled Stage Position calibration#Magnification-and-image-rotation-adjustment-of-the-model|scale and rotation calibration]]> at gr preset_ using "Calibration" application as following:
106 1 Amber Herold
107 14 Amber Herold
_[[Modeled Stage Position calibration|Why goniometer model is better]]._
108 1 Amber Herold
109
110
111 14 Amber Herold
1.  scope> Remove the objective aperture.
112 1 Amber Herold
113 14 Amber Herold
2.  Leginon> Run "Calibrations" Application
114 1 Amber Herold
115 14 Amber Herold
3.  Leginon/Presets Manager> Send "gr" preset to scope.
116 1 Amber Herold
117 14 Amber Herold
4.  Leginon/GonioModeling/Settings/Measurement> Set axis=x, Points=5,Tolerance=25%, and Interval=1e-05 m. Default model label is the session name.
118 1 Amber Herold
119 14 Amber Herold
5.  Leginon/GonioModeling/Settings/Modeling> Set "Scale and Rotation Adjustment Only"=Yes. Other settings will be set automatically once measurements are finished.
120 1 Amber Herold
121 14 Amber Herold
6.  Leginon/GonioModeling> Click on !http://emg.nysbc.org/software/leginon/images/icons/cam_ruler.png! will start "Measure" the shifts of between points that are moved on x axis of the goniometer.
122 1 Amber Herold
123 14 Amber Herold
7.  Leginon/GonioModeling> Click on !http://emg.nysbc.org/software/leginon/images/icons/play.png! will "Calibrate" the scale and rotation of the model on x axis of the goniometer.
124 1 Amber Herold
125 14 Amber Herold
8.  Leginon/GonioModeling> repeat 4-7 with the axis set to y.
126 1 Amber Herold
127
128
129
130
h2. Acquire mosaic atlas of the grid
131
132
133 14 Amber Herold
1.  scope> Remove the objective apperture.
134 1 Amber Herold
135 14 Amber Herold
2.  Leginon/Grid Targeting> Click "Settings" to check the preset used, the size and name of the atlas.
136 1 Amber Herold
137 14 Amber Herold
3. Leginon/Grid Targeting> Click "Calculate Atlas" !http://emg.nysbc.org/software/leginon/images/icons/calculate.png!
138 1 Amber Herold
139 14 Amber Herold
4.  Leginon/Grid Targeting> Click "Submit Targets" !http://emg.nysbc.org/software/leginon/images/icons/play.png! to begin acquiring the atlas.
140 1 Amber Herold
141 14 Amber Herold
5.  Leginon/Square Targeting> Follow the atlas progress in the display. (***There is no sign at the end currently other than that nothing happens for a long time and that the atlas dimension is the same in other direction when it is completed.)
142 1 Amber Herold
143 14 Amber Herold
6.  scope> Insert the objective aperture.
144 1 Amber Herold
145 14 Amber Herold
7.  Leginon/Presets Manager/Send an HM preset "To Scope" several times to cycle all magnifications and warm up the objective lens to minimize the image shift hysteresis in switching between LM and HM modes. If the scope has been in LM mode for a long time, wait for at least 30 min before final image shift refinement of needed LM preset during acquisition (i.e. sq)
146 1 Amber Herold
147
148
149
150 14 Amber Herold
h2. Determine the proper exposure condition for exposure presets
151 1 Amber Herold
152
153
154
Exposure presets are the base presets for data collection. To keep it to low dose
155
condition such as 10 electron/Angstrum, measurement of dose is performed and the exposure
156
time is adjusted. If scope or camera parameters are changed during the session, the dose
157
value is no longer valid and should be remeasured.
158
159
160
161 14 Amber Herold
1.  Leginon/Navigation/Toolbar/Stage Locations !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! Send the "empty" position "To scope".
162 1 Amber Herold
163 14 Amber Herold
2.  Leginon/Presets Manager/Preset Selector> Select a high mag exposure (i.e. "en" or "ef"), confirm that the parameters are what you desire for the experiment, and then send "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!.
164 3 Alex Kadokura
165 14 Amber Herold
3.  scope> Expand the beam to at least cover the CCD (> medium circle mark on the viewing screen) and to the size you desire.
166 1 Amber Herold
167 14 Amber Herold
4.  Leginon/Presets Manager/Preset Selector> If you have changed the size or location of the beam or other preset scope parameters at the scope, you must retrieve
168
the preset parameter "From scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png! first before checking and saving the dose value.
169 1 Amber Herold
170 14 Amber Herold
5.  Leginon/Presets Manager/Preset Selector> "Acquire dose image" !http://emg.nysbc.org/software/leginon/images/icons/acquire.png! for the selected preset.
171 1 Amber Herold
172 14 Amber Herold
6.  Leginon/Presets Manager/Dose Image> Check the "Dose" value.
173 1 Amber Herold
174 14 Amber Herold
7.  Leginon/Presets Manager/Dose Image> If desired dose is known, fill in the value and then click "Match". This will change exposure time to give a close match of
175 1 Amber Herold
the value entered.
176 3 Alex Kadokura
177 14 Amber Herold
8.  Leginon/Presets Manager/Dose Image> If not satisfied with the value, click "No", change the exposure time by editing the preset, and try again (No need to send
178 1 Amber Herold
the preset "To scope" again for the repeat)
179
180 14 Amber Herold
9.  If the scope can not give such illumination to satisfy dose, exposure time and area requirement, alignment is required or change of gun lens or spot size is needed.
181 1 Amber Herold
182 14 Amber Herold
10.  Leginon/Presets Manager> Copy the preset parameter to the other exposure preset.
183 1 Amber Herold
184 14 Amber Herold
 * Preset Selector> send the preset that has the dose measurement "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!
185
 * Preset Selector> select the preset to receive the copy
186
 * Preset Selector> retrieve the parameters "From Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png! for the selected preset.
187
 * Preset Selector/Edit> !http://emg.nysbc.org/software/leginon/images/icons/edit.png! Modify defocus for the preset after the retrieval.
188 1 Amber Herold
189
190
191 14 Amber Herold
h2. Update bright/dark correction images for exposure presets
192 1 Amber Herold
193
194
195 14 Amber Herold
Exposure presets are the base presets for data collection. To make sure the best correction is applied to them, the correction images should be acquired prior to the experiment at session. Since no correlation is normally done on these images, only one [[Bright and Dark reference images#Correction-Channels|channel]] is needed. Normalization image might need to be [[Bright and Dark reference images#Correction-Planmodified]] if a large chunk of region is blocked by a fallen debris.
196 1 Amber Herold
197
198 14 Amber Herold
1.  Leginon/Presets Manager> Select a dose measured high mag exposure (i.e. "en" or "ef"), and then send "To Scope".
199 1 Amber Herold
200 14 Amber Herold
2.  Leginon/Correction/Settings>
201 1 Amber Herold
202 14 Amber Herold
 *  Select TEM and camera to match that will be used.
203
 
204
 *  Camera Configuration: The exposure preset camera configuration
205 1 Amber Herold
206 14 Amber Herold
3.  Leginon/Correction/Settings> Click OK to exit settings.
207 1 Amber Herold
208 14 Amber Herold
4.  Leginon/Correction/Toolbar> Select "Raw image " from the pull down list of acquisition modes and then click on "Acquire" button next to the selector to view an image that is not corrected.
209 1 Amber Herold
210 14 Amber Herold
5.  Leginon/Correction/Toolbar> Select "Dark reference" in the acquisition mode selector and then click "Acquire" to acquire the Dark reference image for this particular camera configuration.
211 1 Amber Herold
212 14 Amber Herold
6.  Leginon/Correction/Toolbar> Select "Bright reference" and repeat the acquisition to obtain the Bright reference.
213 1 Amber Herold
214 14 Amber Herold
7.  Leginon/Correction/Toolbar> Select "Corrected image" and then "Acquire" to view the corrected image. A corrected image should be free of artifacts and have
215 1 Amber Herold
smaller standard deviation than the raw image, in general.
216
217 14 Amber Herold
8.  Leginon/Correction/Toolbar> If there are regions consistently over normalizes the acquired image, either select the region and use the "+" tool to "Add Region to Bad Pixel List", or click on !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! to "Add Extreme Points to Bad Pixel List" one of more times to find the outliers. Right click on a shown bad pixel removes it from the list.
218 1 Amber Herold
219
220
221
h2. Medium mag preset image and beam shift refinement
222
223
224 14 Amber Herold
Preset image shift alignment can be done follow the same procedure as the section "[[MSI set-up in more details#Grid-preset-image-shift-refinement|Grid preset image shift refinement]]" but is more efficient using the automated mode as follows.
225 1 Amber Herold
226
227 14 Amber Herold
1.  Leginon/Presets Manager> Send one of the exposure preset as the reference (e.g. en) "To Scope" !http://emg.nysbc.org/software/leginon/images/icon /instrumentset.png!
228 1 Amber Herold
229 14 Amber Herold
2.  scope> Move the grid to a position on the "align" square, either on a recognizable object or burn a good hole through ice at the center of the CCD as preparation for lower mag preset image shift correction.
230 1 Amber Herold
231 14 Amber Herold
3.  Leginon/Presets Manager> Click on the tool "Align presets to each other" !http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png!. The tool automatically select the reference preset that was selected in the Presets Manager mainpanel and choose a preset at closest magnification as the starting aligning preset.
232 1 Amber Herold
233 14 Amber Herold
4.  Leginon/Presets Manager/Align Presets> Click "Start" to acquire the first image pair.
234 1 Amber Herold
235 14 Amber Herold
5.  Check, adjust, and save the preset(s) to be aligned.
236 1 Amber Herold
237 14 Amber Herold
 The left panel displays and navigate by "Stage Position" of the reference preset while the right panel displays and navigate by "Image Shift" of the preset to be aligned
238 1 Amber Herold
239 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Changing Selection of "Similar look across mags" or "Full CCD" mode at the top left corner of the window will cause reacquisition of the image pair in the automated mode.
240 1 Amber Herold
241 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> If the recognizable object can be found in both panels but not at the center on the reference left panel, navigate* it to the center using its "Click Tool". Wait for the move to complete and both images refreshed.
242 1 Amber Herold
243 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Navigate* the object on the right panel to the center using its "Click Tool". The new image shift is automatically saved after each move.
244 1 Amber Herold
245 14 Amber Herold
 *  Repeat 3 if necessary.
246 1 Amber Herold
247 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Click "Continue" to save the alignment to all presets of the same magnification and continue to the next preset pair.
248 1 Amber Herold
249 14 Amber Herold
6.  Repeat the process if desired.
250 1 Amber Herold
251
252
253 8 Anchi Cheng
h2. Go to MSI Operation section
254
255 1 Amber Herold
256
257
The setup that requires you to be next to the microscope is completed. Go to the section
258
on [[MSI Operation|MSI Operation]] You can come back and continue on this section after you
259
have learned how to go through the operation without automated hole finding.
260
261
262
263
264
265
h2. Hole Targeting Set-up
266
267
268 4 Alex Kadokura
269 1 Amber Herold
We can use any image that is already loaded into Hole Targeting to adjust these
270 4 Alex Kadokura
parameters.  The values in this node vary so much from grid to grid that the best way to
271 1 Amber Herold
find these parameters is by going through a step-by-step trial-and-error process. To see the
272 4 Alex Kadokura
final acquisition and focus targets, enable only the Original, acquisition, and focus
273 1 Amber Herold
display selections. Refer to Introduction Chapter on [[MSI set-up in more details
274
275
|how the various tools behave in the control panel]].
276
277
278
The loaded image normally comes from the image acquired through "Square" node when you
279
submit targets on the grid atlas in "Square Targeting" node.  You can also do a practice run
280
using the default image comes with Leginon. Leave the filename entry blank when you click
281
"Load" in the hole finding settings window for the step "Original". The default values in
282
the node gives an example of reasonable target finding with such default image.
283
284
285
286
287 4 Alex Kadokura
Setting up Hole Targeting Depends on which MSI flavor you are running:
288 1 Amber Herold
289
290
291 4 Alex Kadokura
292 1 Amber Herold
* [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the
293
edge image
294
295 4 Alex Kadokura
296 1 Amber Herold
* [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original
297
image
298
299 4 Alex Kadokura
300 1 Amber Herold
* [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given
301
origin
302
303
304
* [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic
305
target finding. Therefore, it does not requires set up.
306
307
308
309
310
311
312
313
314
315
h2.  Exposure Targeting Set-up
316
317
318
319
Like hole targeting, we can use any image that is already loaded into Exposure Targeting
320 4 Alex Kadokura
to adjust these parameters.
321
322 1 Amber Herold
Setting up Exposure Targeting Depends on which MSI flavor you are running.
323
324 4 Alex Kadokura
325 1 Amber Herold
* [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the
326
edge image
327
328 4 Alex Kadokura
329 1 Amber Herold
* [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original
330
image
331
332 4 Alex Kadokura
333
* [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given
334 1 Amber Herold
origin
335
336
337
* [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic
338
target finding. Therefore, it does not requires set up.
339
340
341
342
343
344
345
h2. Autofocus Test
346
347
348
349
There are two focusers in MSI. "Z Focus" node usually moves stage in Z direction to
350
Eucentric height at lower mags. "Focus" node usually just correct the defocus in the way
351
equivalent to turning the focus adjustment knob on the scope at a higher mag. The defocus
352
measurments in both nodes are based on the same principle of beam-tilt induced image shift.
353
A correlation (normally phase) is used to obtain the image shift. Therefore, it is important
354
to test and preform autofocus measurement at area with recognizable feature that does not
355
move significantly during the defocus measurement.
356
357
358
The test use "Manual_after" focus step in the focus sequence for checking. Two
359
procedures can be used to reach a area for good correlation measurements:
360
361
362
363
364
*  Using Simulate Target tool
365
366
367
368
369
*  Leginon/Presets Manager> Send the autofocus preset "To Scope"
370
371
372
*  scope> Move to an area that has recognizable feature on the carbon
373
support at the center.
374
375
376
*  Leginon/(Z) Focus/Focus Sequence<inlinegraphic
377
fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
378
format="PNG"/>> Enable desired autofocus and "Manual_after" focus
379
steps.
380
381
382
*  Leginon/(Z) Focus/Toolbar> Click Simulate Target<inlinegraphic
383
fileref="http://emg.nysbc.org/software/leginon/images/icons/simulatetarget.png"
384
format="PNG"/> and follow the autofocus
385
386
387
388
389
390
391
*  Submiting focus target from Hole Targeting
392
393
394
*To use this procedure, there has to be a grid mosaic, and the Hole Targeting
395
should allow user verification of targets
396
397
398
*  Leginon/(Z) Focus/Focus Sequence<inlinegraphic
399
fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
400
format="PNG"/>> Enable desired autofocus and "Manual_after" focus
401
steps.
402
403
404
*  Leginon/Square Targeting> Select a test square and submit it<inlinegraphic
405
fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png"
406
format="PNG"/>
407
408
409
*  Leginon/Hole Targeting> Submit a focus target to test autofocus in Z Focus
410
node and an acquisition target for the next stage. Both targets should be in an
411
area that has recognizable feature with carbon support.
412
413
414
*  Leginon/Z focus> follow the autofocus test in this node.
415
416
417
*  Leginon/Exposure Targeting> Submit a focus target at a location that has
418
recognizable feature with carbon support to test autofocus in Focus node.
419
420
421
*  Leginon/Focus> follow autofocus test in this node.
422
423
424
425
426
427
428
429
430
Checking the performance of autofocus by the following
431
criteria:
432
433
434
435
436
*  During autofocusing
437
438
439
440
441
*  Display the image and correlection peak: The observed movement of images
442
upon beam tilt should be consistent with the correlation peak and the edge of
443
the beam should not be visible in any of the beam-tilt images.
444
445
446
*  Display the correlation image: The correlation peak at its correct position
447
should be significantly higher than background.
448
449
450
451
452
453
454
*  During manual focusing
455
456
Manual Focusing can be activated either by enabling "Manual_before/Manual_after"
457
steps or directly when leginon is idle using (Z) Focus/Toolbar/MF. To activate manual
458
focus directly, it is necessary to send to scope the preset of the choice (normally
459
fc) first, though.
460
461
462
*  MF> The FFT should show visible and isotropic Thon ring according to the
463
fc preset defocus.
464
465
466
*  MF/Toolbar> Make sure the move type is "Defocus" and the set value in the
467
box is 0 m then click "Set instrument" botton<inlinegraphic
468
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
469
format="PNG"/>. This will set the defocus to the current 0.
470
471
472
*  MF> The perfect autofocus result is that the FFT indicates that the scope
473
is at Gaussian focus.
474
475
476
*  For relative movement, press "+" or "-". The step size is determined in
477
MF/Toolbar/Settings>
478
479
480
*  To reset defocus to zero at the new focus, press the so-named
481
button<inlinegraphic
482
fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentsetnew.png"
483
format="PNG"/>.
484 2 Amber Herold
485
486
*  Clicking the Stop button<inlinegraphic
487
fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png"
488
format="PNG"/> exits the maunal check.
489
490
491